FINITE ELEMENT ANALYSIS OF CUSTOMIZED KNEE IMPLANTS BY VARYING LOADS DURING FLEXION-EXTENSION MOVEMENT

被引:0
|
作者
Alshewaier, Shady A. A. [1 ]
Alshehri, Majed Ali Ghurmallah [2 ]
Begum, S. Sabarunisha [3 ]
Sikkandar, Mohamed Yacin [2 ]
Almukil, Ali Abdullah A. [2 ]
Alhazmi, Aishah Mohammad [4 ]
机构
[1] Majmaah Univ, Coll Appl Med Sci, Dept Phys Therapy, Al Majmaah 11952, Saudi Arabia
[2] Majmaah Univ, Coll Appl Med Sci, Dept Med Equipment Technol, Al Majmaah 11952, Saudi Arabia
[3] PSR Engn Coll, Dept Biotechnol, Sivakasi 626140, Tamil Nadu, India
[4] Ad Diriyah Hosp, Minist Hlth, Diriyah, Saudi Arabia
关键词
Total knee replacement; knee osteoarthritis; off-the-shelf knee; patient-specific; biomechanical; ARTHROPLASTY; BIOMECHANICS;
D O I
10.1142/S021951942340016X
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A customized knee implant numerical modeling using finite element analysis (FEA) during flexion extension has been investigated in this paper with varying loads with an objective of studying its kinematics. Computed tomography (CT) images of 15 osteoarthritis subjects' images were used in this work. Various morphological characteristics were extracted from clinical images using a commercial CAE software and biomechanical properties were studied on applying standard loads on customized implant and off-the-shelf (OTS) implants. Patient-specific knee implants have been designed according to the morphological characteristics and bone dimension of patient with compressive loads (1500, 1700 and 3000 N) during normal gait and were compared with the normal knee. Results showed that the stresses are distributed equally to the spacer and the tibial plate, unlike the standard femoral component where the stresses get concentrated on the cut edges. In compressive load, active stress and strain (similar to 5-20 MPa) are lesser (similar to 5%) and in flexion extension also lesser with a scaling factor of 0.785 and 1.0. The designed implant was found to produce similar biomechanical properties when compared to normal knee joint and thus it can be considered as a valuable implant and could replace the standard OTS knee implants.
引用
收藏
页数:15
相关论文
共 26 条
  • [1] Contact pressure distribution in postmortem human knee during dynamic flexion-extension movement
    Quintelier, J.
    De Baets, P.
    Almqvist, F.
    4TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, 2009, 22 (1-3): : 1798 - 1802
  • [2] Experimental measurement of flexion-extension movement in normal and osteoarthritic human knee
    Tarnita, Daniela
    Catana, M.
    Tarnita, D. N.
    ROMANIAN JOURNAL OF MORPHOLOGY AND EMBRYOLOGY, 2013, 54 (02) : 309 - 313
  • [3] The flexion-extension response of a novel lumbar intervertebral disc prosthesis: A finite element study
    Garcia Vacas, Francisco
    Ezquerro Juanco, Francisco
    Perez de la Blanca, Ana
    Prado Novoa, Maria
    Postigo Pozo, Sergio
    MECHANISM AND MACHINE THEORY, 2014, 73 : 273 - 281
  • [4] No differences in knee kinematics between active and passive flexion-extension movement: an intra-operative kinematic analysis performed during total knee arthroplasty
    Alberto Grassi
    Nicola Pizza
    Nicola Francesco Lopomo
    Maurilio Marcacci
    Michele Capozzi
    Giulio Maria Marcheggiani Muccioli
    Francesca Colle
    Stefano Zaffagnini
    Journal of Experimental Orthopaedics, 7
  • [5] No differences in knee kinematics between active and passive flexion-extension movement: an intra-operative kinematic analysis performed during total knee arthroplasty
    Grassi, Alberto
    Pizza, Nicola
    Lopomo, Nicola Francesco
    Marcacci, Maurilio
    Capozzi, Michele
    Muccioli, Giulio Maria Marcheggiani
    Colle, Francesca
    Zaffagnini, Stefano
    JOURNAL OF EXPERIMENTAL ORTHOPAEDICS, 2020, 7 (01)
  • [6] ACL Forces and Knee Kinematics Produced by Axial Tibial Compression During a Passive Flexion-Extension Cycle
    Markolf, Keith L.
    Jackson, Steven R.
    Foster, Brock
    McAllister, David R.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2014, 32 (01) : 89 - 95
  • [7] Effect of degenerative factors on cervical spinal cord during flexion and extension: a dynamic finite element analysis
    Xu, Meng-lei
    Zeng, Hui-zi
    Zheng, Liang-dong
    Jin, Chen
    Zhu, Shi-jie
    Yang, Yi-ting
    Cao, Yu-ting
    Zhu, Rui
    Cheng, Li-ming
    BIOMECHANICS AND MODELING IN MECHANOBIOLOGY, 2022, 21 (06) : 1743 - 1759
  • [8] The Study of Contact Pressure of a Customized Knee Spacer Using Finite Element Analysis
    Choh, Alex C. T.
    Ng, Jaryl C. K.
    Yew, Andy K. S.
    Chia Shi-Lu
    Chong, Desmond Y. R.
    7TH WACBE WORLD CONGRESS ON BIOENGINEERING 2015, 2015, 52 : 92 - 95
  • [9] Investigation of customized total knee implant with articular cartilage under loading conditions using finite element analysis
    Alshewaier, Shady A.
    Sikkandar, Mohamed Yacin
    Almakrami, Ali Ahmed Ali
    Begum, S. Sabarunisha
    Alassaf, Ahmad
    Almohimeed, Ibrahim
    Sundaram, S. Meenatchi
    Poojary, Dheeraj
    Sudharsan, Natteri M.
    Ng, Eddie Y. K.
    PLOS ONE, 2025, 20 (02):
  • [10] Wear estimation of hip implants with varying chamfer geometry at the trunnion junction: a finite element analysis
    Gutmann, Celine
    Shaikh, Numa
    Shenoy, B. Satish
    Bhat, N. Shaymasunder
    Keni, Laxmikant G.
    Chethan, K. N.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2023, 9 (03)