Intraoperative Cone-Beam Computed Tomography Assessment of Spinal Pedicle Screws Placement Precision Is in Full Agreement with Postoperative Computed Tomography Assessment

被引:1
作者
Ohashi, Hiroki [1 ]
Kawamura, Daichi [2 ]
Hatano, Keisuke [1 ]
Ohashi, So [2 ]
Tochigi, Satoru [2 ]
Isoshima, Akira [3 ]
Nagashima, Hiroyasu [4 ]
Otani, Katharina [5 ]
Karagiozov, Kostadin [1 ]
Tani, Satoshi [1 ]
Murayama, Yuichi [1 ]
机构
[1] Jikei Univ, Dept Neurosurg, Sch Med, Minato Ku, Tokyo, Japan
[2] Jikei Univ, Dept Neurosurg, Sch Med, Kashiwa Hosp, Kashiwa, Chile
[3] Omori Red Cross Hosp, Dept Neurosurg, Ota Ku, Tokyo, Japan
[4] Jikei Univ, Dept Neurosurg, Sch Med, Katsushika Med Ctr, Katsushika Ku, Tokyo, Japan
[5] Siemens Healthcare KK, AT Innovat Dept, Gate City Osaki West Tower,Shinagawa Ku, Tokyo, Japan
关键词
Pedicle screws; Robotic C-arm system; Spinal surgery; Spine; IMAGING TECHNIQUE; ACCURACY; RELIABILITY; FIXATION; GUIDANCE;
D O I
10.1016/j.wneu.2023.03.062
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
■ OBJECTIVE: To assess agreement between pedicle screw placement evaluated on postoperative computed tomography (CT) and on intraoperative cone-beam CT (CBCT) and compare procedure characteristics when using first-generation and second-generation robotic C-arm sys-tems in the hybrid operating room. ■ METHODS: All patients who received pedicle screws for spinal fusion at our institution between June 2009 and September 2019 and underwent intraoperative CBCT and postoperative CT were included. The CBCT and CT images were reviewed by 2 surgeons to assess the screw place-ment using the Gertzbein-Robbins and the Heary classifi-cations. Intermethod agreement of screw placement classifications as well as interrater agreement were assessed using Brennan-Prediger and Gwet agreement coefficients. Procedure characteristics using first -generation and second-generation generation robotic C-arm systems were compared. ■ RESULTS: Fifty-seven patients were treated with 315 pedicle screws at thoracic, lumbar, and sacral levels. No screw had to be repositioned. On CBCT, accurate place-ment was found for 309 screws (98.1%) using the Gertzbein-Robbins classification and 289 (91.7%) using the Heary classification and on CT, these were 307 (97.4%) and 293 (93.0%), respectively. Intermethod between CBCT and CT and interrater agreements between the 2 raters were almost perfect (>0.90) for all assessment. There were no significant differences in mean radiation dose (P = 0.83) and fluoroscopy time (P = 0.82), but length of surgery using the second-generation system was estimated at 107.7 mi -n utes (95% confidence interval, 31.9e183.5 minutes; P = 0.006) shorter. ■ CONCLUSIONS: Intraoperative CBCT provides accurate assessment of pedicle screw placement and enables intraoperative repositioning of misplaced screws.
引用
收藏
页码:E254 / E263
页数:10
相关论文
共 21 条
  • [1] Percutaneous Pedicle Screw Stabilization: Surgical Technique, Fracture Reduction, and Review of Current Spine Trauma Applications
    Alander, Dirk H.
    Cui, Shari
    [J]. JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2018, 26 (07) : 231 - 240
  • [2] Accuracy of upper thoracic pedicle screw placement using three-dimensional image guidance
    Bledsoe, Jonathan M.
    Fenton, Doug
    Fogelson, Jeremy L.
    Nottmeier, Eric W.
    [J]. SPINE JOURNAL, 2009, 9 (10) : 817 - 821
  • [3] Pedicle screw insertion accuracy in terms of breach and reposition using a new intraoperative cone beam computed tomography imaging technique and evaluation of the factors associated with these parameters of accuracy: a series of 695 screws
    Cordemans, Virginie
    Kaminski, Ludovic
    Banse, Xavier
    Francq, Bernard G.
    Detrembleur, Christine
    Cartiaux, Olivier
    [J]. EUROPEAN SPINE JOURNAL, 2017, 26 (11) : 2917 - 2926
  • [4] Accuracy of a new intraoperative cone beam CT imaging technique (Artis zeego II) compared to postoperative CT scan for assessment of pedicle screws placement and breaches detection
    Cordemans, Virginie
    Kaminski, Ludovic
    Banse, Xavier
    Francq, Bernard G.
    Cartiaux, Olivier
    [J]. EUROPEAN SPINE JOURNAL, 2017, 26 (11) : 2906 - 2916
  • [5] Gazzeri Roberto, 2016, Surg Technol Int, V28, P303
  • [6] Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniques
    Gelalis, Ioannis D.
    Paschos, Nikolaos K.
    Pakos, Emilios E.
    Politis, Angelos N.
    Arnaoutoglou, Christina M.
    Karageorgos, Athanasios C.
    Ploumis, Avraam
    Xenakis, Theodoros A.
    [J]. EUROPEAN SPINE JOURNAL, 2012, 21 (02) : 247 - 255
  • [7] ACCURACY OF PEDICULAR SCREW PLACEMENT INVIVO
    GERTZBEIN, SD
    ROBBINS, SE
    [J]. SPINE, 1990, 15 (01) : 11 - 14
  • [8] Gwet K. L., 2014, HDB INTERRATER RELIA
  • [9] Thoracic pedicle screws: postoperative computerized tomography scanning assessment
    Heary, RF
    Bono, CM
    Black, M
    [J]. JOURNAL OF NEUROSURGERY, 2004, 100 (04) : 325 - 331
  • [10] Radiation exposure to the patients in thoracic and lumbar spine fusion using a new intraoperative cone-beam computed tomography imaging technique: a preliminary study
    Kaminski, Ludovic
    Cordemans, V.
    Cartiaux, O.
    Van Cauter, M.
    [J]. EUROPEAN SPINE JOURNAL, 2017, 26 (11) : 2811 - 2817