Favorable mid-term performance of fully biodegradable implantable device for ventricular septal defect closure

被引:2
作者
Song, Shubo [1 ]
Shao, Zehua [1 ]
Fan, Taibing [1 ,3 ]
Li, Bin [1 ]
Liang, Weijie [1 ]
Dong, Haoju [1 ]
Wu, Kaiyuan [1 ]
Hu, Manman [1 ]
Han, Yu [1 ]
Cui, Cunying [2 ]
机构
[1] Zhengzhou Univ, Dept Childrens Heart Ctr, Peoples Hosp, Zhengzhou, Peoples R China
[2] Zhengzhou Univ Cent China, Fuwai Hosp, Dept Ultrasound, Zhengzhou, Peoples R China
[3] Zhengzhou Univ, Childrens Heart Ctr, Peoples Hosp, Dept, Weiwu Rd, Zhengzhou 450000, Peoples R China
关键词
ventricular septal defect; biodegradable de-vice; echocardiography; transthoracic intervention; mini-mally invasive surgery; TRANSCATHETER CLOSURE; EXPERIENCE;
D O I
10.1016/j.xjtc.2022.11.010
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: To assess the mid-term safety and efficacy of transthoracic perimem-branous ventricular septal defect (Pm-VSD) closure using a new biodegradable device. Implantation entailed right subaxillary minithoracotomy under transesopha-geal echocardiography guidance.Methods: Between October 2019 and January 2020, 13 patients (males, 5; mean age, 3.6 +/- 2.5 years) with Pm-VSDs underwent transthoracic device closures at Zhengzhou University Central China Fuwai Hospital as described previously. Deliv-ery pathways were established by manipulating a hollow probe from right atrium through tricuspid valve to right ventricle and then through VSDs to left ventricle, whereupon installation took place.Results: All occluder implantations were successfully executed. Mean defect size was 4.1 +/- 1.0 mm, and mean device waist size was 5.2 +/- 1.1 mm. One patient (7.7%) with 1.5-mm residual shunt showed complete closure at discharge. There was 1 instance of postoperative incomplete right bundle branch block, which con-verted to complete right bundle branch block at month 1. During patient follow-up (mean, 24.6 +/- 0.8 months), no device dislocations, new residual shunts, new valvular regurgitation, or detectable atrioventricular block ensued.Conclusions: Closure of Pm-VSDs using a novel, fully biodegradable occluder in the manner described has proven safe and effective at mid-term follow-up. Long-term safety and efficacy of this device must be further corroborated in a large patient cohort going forward. (JTCVS Techniques 2023;17:133-7)
引用
收藏
页码:133 / 137
页数:5
相关论文
共 14 条
[1]   Short- and Long-term Outcome After Interventional VSD Closure: A Single-Center Experience in Pediatric and Adult Patients [J].
Bergmann, M. ;
Germann, C. P. ;
Nordmeyer, J. ;
Peters, B. ;
Berger, F. ;
Schubert, S. .
PEDIATRIC CARDIOLOGY, 2021, 42 (01) :78-88
[2]   A novel biodegradable occluder for the closure of ventricular septal defects: immediate and medium-term results in a canine model [J].
Bu, Haisong ;
Yang, Yifeng ;
Hu, Shijun ;
Wu, Qin ;
Gong, Xueyang ;
Zhao, Tianli .
INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2019, 29 (05) :783-792
[3]   First-in-Human Experience With a Novel Fully Bioabsorbable Occluder for Ventricular Septal Defect [J].
Chen, Long ;
Hu, Shengshou ;
Luo, Zhiling ;
Butera, Gianfranco ;
Cao, Qiling ;
Zhang, Fengwen ;
Lu, Minjie ;
Wang, Shouzheng ;
Wang, Weiwei ;
Pan, Xiangbin .
JACC-CARDIOVASCULAR INTERVENTIONS, 2020, 13 (09) :1139-1141
[4]   First Experience With the Biostar-Device for Various Applications in Pediatric Patients With Congenital Heart Disease [J].
Hoehn, Rene ;
Hesse, Carola ;
Ince, Hueseyin ;
Peuster, Matthias .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 75 (01) :72-77
[5]   Development and preclinical evaluation of a biodegradable ventricular septal defect occluder [J].
Huang, Xin-miao ;
Zhu, Yu-feng ;
Cao, Jiang ;
Hu, Jian-qiang ;
Bai, Yuan ;
Jiang, Hai-bing ;
Li, Zhao-feng ;
Chen, Ying ;
Wang, Wei ;
Qin, Yong-wen .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 81 (02) :324-330
[6]   Postimplant complications with transcatheter closure of congenital perimembranous ventricular septal defects [J].
Jin, Youpeng ;
Han, Bo ;
Zhang, Jianjun ;
Zhuang, Jianxin ;
Yan, Jinchun ;
Wang, Yulin .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 81 (04) :666-673
[7]   Transcatheter Closure of Perimembranous Ventricular Septal Defect Using the Lifetech Konar-Multi Functional Occluder: Early to Midterm Results of the Indonesian Multicenter Study [J].
Kuswiyanto, Rahmat Budi ;
Gunawijaya, Eka ;
Djer, Mulyadi M. ;
Noormanto ;
Rahman, Mahrus A. ;
Murni, Indah K. ;
Sukardi, Rubiana ;
Utamayasa, Alit ;
Ardiansyah, Rizky ;
Nova, Ria ;
Liliyanti, Sri ;
Rahayuningsih, Sri E. ;
Anggriawan, Shirley L. ;
Rahayuningsih, Tri Yanti ;
Koentartiwi, Dyahris ;
Soewarniaty, Renny ;
Yantie, Venny Kartika ;
Nugroho, Sasmito ;
Hidayat, Taufiq ;
Ontoseno, Teddy ;
Tobing, Tina C. ;
Ali, Muhamad ;
Bashari, Muhammad Hasan ;
Yosy, Denny S. ;
Arafuri, Nadya ;
Hilmanto, Dany ;
Yanuarso, Piprim B. ;
Advani, Najib ;
Sastroasmoro, Sudigdo ;
Putra, Sukman Tulus .
GLOBAL HEART, 2022, 17 (01)
[8]   TRANSCATHETER CLOSURE OF VENTRICULAR SEPTAL-DEFECTS [J].
LOCK, JE ;
BLOCK, PC ;
MCKAY, RG ;
BAIM, DS ;
KEANE, JF .
CIRCULATION, 1988, 78 (02) :361-368
[9]   Transcatheter Closure of Perimembranous Ventricular Septal Defects Using Different Generations of Amplatzer Devices: Multicenter Experience [J].
Mijangos-Vazquez, Roberto ;
El-Sisi, Amal ;
Sandoval Jones, Juan P. ;
Garcia-Montes, Jose A. ;
Hernandez-Reyes, Rogelio ;
Sobhy, Rodina ;
Abdelmassih, Antoine ;
Soliman, Mohammed M. ;
Ali, Safaa ;
Molina-Sanchez, Tatiana ;
Zabal, Carlos .
JOURNAL OF INTERVENTIONAL CARDIOLOGY, 2020, 2020
[10]   Biodegradable atrial septal defect occluders: A current review [J].
Shi, Daokun ;
Kang, Yahong ;
Zhang, Guoyi ;
Gao, Chenguang ;
Lu, Wei ;
Zou, Hua ;
Jiang, Hongyan .
ACTA BIOMATERIALIA, 2019, 96 :68-80