Patellar Dislocation Masking a Tibial Post Fracture: An Unusual Complication of Posterior Stabilized Total Knee Arthroplasty

被引:0
|
作者
Seth, Aditya [1 ]
Buddhapuram, Pranav K. [1 ]
Kadwad, Arvind [1 ]
Boddeda, Sandeep [1 ]
Vecham, Ratnakar [1 ]
Annapareddy, Adarsh [1 ]
Reddy, A. V. Gurava [1 ]
机构
[1] KIMS Sunshine Hosp, Dept Orthoped, Hyderabad, India
关键词
posterior stabilized knee; cam and post; polyethylene post breakage; arthroplasty complication; chronic patella dislocation; arthroplasty; tibial post fracture; COMPONENT DESIGN; WEAR; PROSTHESIS; FAILURE; DAMAGE;
D O I
10.7759/cureus.48950
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Posterior stabilized total knee arthroplasty (TKA) has established itself as a highly effective design for total knee arthroplasty, renowned for its longevity and success. However, a subset of cases, approximately 6-12%, faces early failure, necessitating revision procedures. This case report presents a unique and previously undocumented complication involving a tibial post fracture following hyperflexion of the knee, masked by chronic patellar dislocation. This case highlights the importance of considering polyethylene wear-related failure in cases of instability without an apparent history of trauma. The surgical intervention involved polyethylene insert exchange, patellar debulking, lateral retinacular release, and quadriceps tendon doublebreasting.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] No difference in gait between posterior cruciate retention and the posterior stabilized design after total knee arthroplasty
    van den Boom, Lennard G. H.
    Halbertsma, Jan P. K.
    van Raaij, Jos J. A. M.
    Brouwer, Reinoud W.
    Bulstra, Sjoerd K.
    van den Akker-Scheek, Inge
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2014, 22 (12) : 3135 - 3141
  • [42] No difference in gait between posterior cruciate retention and the posterior stabilized design after total knee arthroplasty
    Lennard G. H. van den Boom
    Jan P. K. Halbertsma
    Jos J. A. M. van Raaij
    Reinoud W. Brouwer
    Sjoerd K. Bulstra
    Inge van den Akker-Scheek
    Knee Surgery, Sports Traumatology, Arthroscopy, 2014, 22 : 3135 - 3141
  • [43] Simultaneous Post and Cone Fractures of a Polyethylene Insert in a Mobile-Bearing Posterior-Stabilized Total Knee Arthroplasty
    Ueyama, Hideki
    Fukunaga, Kenji
    Takemura, Susumu
    Koyanagi, Junichiro
    Yamamura, Mitsuyoshi
    Nakagawa, Shigeru
    ARTHROPLASTY TODAY, 2024, 26
  • [44] Same-Level Fracture of the Tibial Metal Tray and Polyethylene Insert After Total Knee Arthroplasty
    Kang, Jong Yeal
    Lee, Yong Seuk
    ORTHOPEDICS, 2016, 39 (04) : E787 - E789
  • [45] Impact of Soft Tissue Imbalance on Knee Flexion Angle After Posterior Stabilized Total Knee Arthroplasty
    Tsukada, Sachiyuki
    Fujii, Tomoko
    Wakui, Motohiro
    JOURNAL OF ARTHROPLASTY, 2017, 32 (08) : 2399 - 2403
  • [46] Stability with a Constrained Posterior Stabilized Primary Total Knee Arthroplasty Does Not Compromise Durability
    Yohe, Nicholas
    Vanderbrook, Douglas J.
    Sherman, Alain E.
    Papas, Paraskevi Vivian
    Scuderi, Giles R.
    JOURNAL OF KNEE SURGERY, 2023, 36 (08) : 801 - 805
  • [47] Intra-operative fracture in posterior-stabilised total knee arthroplasty
    Pun, Alex H. F.
    Pun, Wai-Ki
    Storey, Peter
    JOURNAL OF ORTHOPAEDIC SURGERY, 2015, 23 (02) : 205 - 208
  • [48] Total knee arthroplasty in patients with a proximal tibial stress fracture associated with bilateral severe knee osteoarthritis
    Manojlovic, Slavko
    Dodik, Rajko
    Manojlovic, Mirko
    Jovicic, Zeljko
    Dragicevic-Cvjetkovic, Dragana
    SRPSKI ARHIV ZA CELOKUPNO LEKARSTVO, 2024, 152 (1-2) : 39 - 43
  • [49] Post-Cam Design and Contact Stress on Tibial Posts in Posterior-Stabilized Total Knee Prostheses: Comparison Between a Rounded and a Squared Design
    Watanabe, Toshifumi
    Koga, Hideyuki
    Horie, Masafumi
    Katagiri, Hiroki
    Sekiya, Ichiro
    Muneta, Takeshi
    JOURNAL OF ARTHROPLASTY, 2017, 32 (12) : 3757 - 3762
  • [50] High rate of tibial post-fracture in highly cross-linked polyethylene compared to conventional polyethylene in posterior-stabilized total knee arthroplasty at a minimum 12-year follow-up
    Baek, Ji-Hoon
    Lee, Su Chan
    Choi, Kyungwon
    Nam, Chang Hyun
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2022, 30 (12) : 4072 - 4077