A Doherty Power Amplifier for Ultrasound Instrumentation

被引:7
作者
Choi, Hojong [1 ]
机构
[1] Gachon Univ, Dept Elect Engn, 1342 Seongnam Daero, Seongnam 13120, Gyeonggi Do, South Korea
基金
新加坡国家研究基金会;
关键词
Doherty power amplifier; ultrasound instrumentation; ultrasound transducer; TRANSDUCER; SIGNALS; SYSTEM;
D O I
10.3390/s23052406
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The ultrasound instrumentation uses linear power amplifiers with low power efficiency, generating unwanted heat and resulting in the deterioration of the echo signal quality of measured targets. Therefore, this study aims to develop a power amplifier scheme to increase power efficiency while maintaining appropriate echo signal quality. In communication systems, the Doherty power amplifier has shown relatively good power efficiency while producing high signal distortion. The same design scheme cannot be directly applied to ultrasound instrumentation. Therefore, the Doherty power amplifier needs to be re-designed. To verify the feasibility of the instrumentation, a Doherty power amplifier was designed to obtain high power efficiency. The measured gain, output 1-dB compression point, and power-added efficiency of the designed Doherty power amplifier were 33.71 dB, 35.71 dB(m), and 57.24% at 25 MHz, respectively. In addition, the performance of the developed amplifier was measured and tested using the ultrasound transducer through the pulse-echo responses. The output power with 25 MHz, 5-cycle, and 43.06 dB(m) generated from the Doherty power amplifier was sent through the expander to the focused ultrasound transducer with 25 MHz and 0.5 '' diameter. The detected signal was sent via a limiter. Afterwards, the signal was amplified by a 36.8 dB gain preamplifier, and then displayed in the oscilloscope. The measured peak-to-peak amplitude in the pulse-echo response with an ultrasound transducer was 0.9698 V. The data showed a comparable echo signal amplitude. Therefore, the designed Doherty power amplifier can improve the power efficiency used for medical ultrasound instrumentation.
引用
收藏
页数:15
相关论文
共 73 条
[21]   Class-DE Ultrasound Transducer Driver for HIFU Therapy [J].
Christoffersen, Carlos ;
Wong, Wai ;
Pichardo, Samuel ;
Togtema, Greg ;
Curiel, Laura .
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2016, 10 (02) :375-382
[22]  
Colantonio P., 2009, High Efficiency RF and Microwave Solid State Power Amplifiers
[23]  
Cripps S.C., 2002, Advanced techniques in RF power amplifier design
[24]  
Daniels J.M., 2016, Practical Point-of-Care Medical Ultrasound
[25]  
Davidse J., 1991, ANALOG ELECT CIRCUIT
[26]  
Davidsen R.E., 2017, U.S. Patent, Patent No. [9730677B2, 9730677]
[27]   A new high efficiency power amplifier for modulated waves [J].
Doherty, WH .
PROCEEDINGS OF THE INSTITUTE OF RADIO ENGINEERS, 1936, 24 (09) :1163-1182
[28]  
Golio M., 2018, RF and microwave circuits, measurements, and modeling
[29]  
Grebennikov A., 2005, RF and Microwave Power Amplifier Design
[30]  
Grebennikov A., 2011, SWITCHMODE RF POWER