A Doherty Power Amplifier for Ultrasound Instrumentation

被引:7
作者
Choi, Hojong [1 ]
机构
[1] Gachon Univ, Dept Elect Engn, 1342 Seongnam Daero, Seongnam 13120, Gyeonggi Do, South Korea
基金
新加坡国家研究基金会;
关键词
Doherty power amplifier; ultrasound instrumentation; ultrasound transducer; TRANSDUCER; SIGNALS; SYSTEM;
D O I
10.3390/s23052406
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The ultrasound instrumentation uses linear power amplifiers with low power efficiency, generating unwanted heat and resulting in the deterioration of the echo signal quality of measured targets. Therefore, this study aims to develop a power amplifier scheme to increase power efficiency while maintaining appropriate echo signal quality. In communication systems, the Doherty power amplifier has shown relatively good power efficiency while producing high signal distortion. The same design scheme cannot be directly applied to ultrasound instrumentation. Therefore, the Doherty power amplifier needs to be re-designed. To verify the feasibility of the instrumentation, a Doherty power amplifier was designed to obtain high power efficiency. The measured gain, output 1-dB compression point, and power-added efficiency of the designed Doherty power amplifier were 33.71 dB, 35.71 dB(m), and 57.24% at 25 MHz, respectively. In addition, the performance of the developed amplifier was measured and tested using the ultrasound transducer through the pulse-echo responses. The output power with 25 MHz, 5-cycle, and 43.06 dB(m) generated from the Doherty power amplifier was sent through the expander to the focused ultrasound transducer with 25 MHz and 0.5 '' diameter. The detected signal was sent via a limiter. Afterwards, the signal was amplified by a 36.8 dB gain preamplifier, and then displayed in the oscilloscope. The measured peak-to-peak amplitude in the pulse-echo response with an ultrasound transducer was 0.9698 V. The data showed a comparable echo signal amplitude. Therefore, the designed Doherty power amplifier can improve the power efficiency used for medical ultrasound instrumentation.
引用
收藏
页数:15
相关论文
共 73 条
[1]   Class D amplifier for a power piezoelectric load [J].
Agbossou, K ;
Dion, JL ;
Carignan, S ;
Abdelkrim, M ;
Cheriti, A .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2000, 47 (04) :1036-1041
[2]  
Albulet Mihai., 2001, RF POWER AMPLIFIERS
[3]  
Allen P., 2002, CMOS ANALOG CIRCUIT, VSecond
[4]   Analysis and Design of a High Voltage Integrated Class-B Amplifier for Ultra-Sound Transducers [J].
Bianchi, Dario ;
Quaglia, Fabio ;
Mazzanti, Andrea ;
Svelto, Francesco .
IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS I-REGULAR PAPERS, 2014, 61 (07) :1942-1951
[6]   Suppression Technique of HeLa Cell Proliferation Using Ultrasonic Power Amplifiers Integrated with a Series-Diode Linearizer [J].
Choe, Se-woon ;
Choi, Hojong .
SENSORS, 2018, 18 (12)
[7]   Pre-Matching Circuit for High-Frequency Ultrasound Transducers [J].
Choi, Hojong .
SENSORS, 2022, 22 (22)
[8]   Novel dual-resistor-diode limiter circuit structures for high-voltage reliable ultrasound receiver systems [J].
Choi, Hojong .
TECHNOLOGY AND HEALTH CARE, 2022, 30 :S513-S520
[9]   Stacked Transistor Bias Circuit of Class-B Amplifier for Portable Ultrasound Systems [J].
Choi, Hojong .
SENSORS, 2019, 19 (23)
[10]   Development of negative-group-delay circuit for high-frequency ultrasonic transducer applications [J].
Choi, Hojong .
SENSORS AND ACTUATORS A-PHYSICAL, 2019, 299