Removal performance of pre- and post-dilution online hemodiafiltration using identical hemodiafilters in the same patients

被引:1
|
作者
Okada, Kazuyoshi [1 ]
Michiwaki, Hiroyuki [2 ]
Mori, Hiroaki [2 ]
Tashiro, Manabu [1 ]
Inoue, Tomoko [1 ]
Shima, Hisato [1 ]
Ohshima, Koji [3 ]
Minakuchi, Jun [1 ]
Kawashima, Shu [1 ]
机构
[1] Kawashima Hosp, Dept Nephrol, 6-1 Kitasakoichibancho, Tokushima 7700011, Japan
[2] Kawashima Hosp, Dept Clin Engn, Tokushima, Tokushima, Japan
[3] Asahi Kasei Med Co Ltd, Chiyoda Ku, 1-1-2 Yurakucho, Tokyo 1000006, Japan
关键词
Pre-dilution; Post-dilution; Online hemodiafiltration; Albumin leakage; Alpha-1-microglobulin; ALL-CAUSE MORTALITY; FLUX HEMODIALYSIS; ALPHA(1)-MICROGLOBULIN; ALBUMIN;
D O I
10.1007/s10047-022-01379-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Online hemodiafiltration (OHDF) for renal replacement therapy has two modes: pre- (pre-OHDF) and post-dilution OHDF (post-OHDF). To elucidate the precise differences between the two modes, a clinical study was performed using the same polysulfone hemodiafilters in the same patients. Eight patients were treated with ABH((TM))-22PA for 6 weeks: 3 weeks of pre-OHDF (with substitution volumes of 24, 36, and 48 L) and 3 weeks of post-OHDF (6, 8, and 10 L). The reduction ratios of urea, uric acid (UA), creatinine (CRE), inorganic phosphorus (iP), beta-2-microglobulin (beta(2)-MG), and alpha-1-microglobulin (alpha(1)-MG) were evaluated. The removal amounts of beta(2)-MG, alpha(1)-MG, and albumin were also evaluated by analyzing the spent dialysis fluids. The types and numbers of adverse events (AEs) and device malfunctions were recorded. The reduction ratios of urea, UA, CRE, iP, and beta(2)-MG were comparable among all conditions, while that of alpha(1)-MG tended to be slightly higher in post-OHDF than in pre-OHDF. The removal amounts of alpha(1)-MG and albumin in pre-OHDF and post-OHDF were significantly greater with the maximum substitution volume than with the minimum volume. However, the selective removal indices, which were obtained by dividing the amount of alpha(1)-MG removed by the albumin level, tended to be slightly higher in pre- than in post-OHDF. No device-related AEs or device malfunctions occurred in either mode. No significant differences in inflammatory responses, evaluated by high-sensitivity C-reactive protein and interleukin-6, were observed. This study provides removal performance and safety data regarding the application of ABH-22PA for pre- and post-OHDF.
引用
收藏
页码:309 / 315
页数:7
相关论文
共 13 条
  • [1] Removal performance of pre- and post-dilution online hemodiafiltration using identical hemodiafilters in the same patients
    Kazuyoshi Okada
    Hiroyuki Michiwaki
    Hiroaki Mori
    Manabu Tashiro
    Tomoko Inoue
    Hisato Shima
    Koji Ohshima
    Jun Minakuchi
    Shu Kawashima
    Journal of Artificial Organs, 2023, 26 : 309 - 315
  • [2] Comparison of the pre-dilution and post-dilution methods for online hemodiafiltration
    Kawai, Yusuke
    Maeda, Kazuya
    Moriishi, Misaki
    Kawanishi, Hideki
    Masaki, Takao
    JOURNAL OF ARTIFICIAL ORGANS, 2024, 27 (01) : 48 - 56
  • [3] Comparison of the pre-dilution and post-dilution methods for online hemodiafiltration
    Yusuke Kawai
    Kazuya Maeda
    Misaki Moriishi
    Hideki Kawanishi
    Takao Masaki
    Journal of Artificial Organs, 2024, 27 : 48 - 56
  • [4] Evaluation of low-volume post-dilution online hemodiafiltration with Japanese high-performance hemodiafilters
    Sakurai, Kenji
    Hosoya, Hiromi
    Kurihara, Yoshitaka
    Yamauchi, Fumi
    Suzuki, Ayumi
    Kurosawa, Kaori
    Saito, Takeshi
    JOURNAL OF ARTIFICIAL ORGANS, 2020, 23 (03) : 234 - 239
  • [5] Comparison of citrate dialysate in pre- and post-dilution online hemodiafiltration: effect on clot formation and adequacy of dialysis in hemodialysis patients
    Nakornchai, Pasu
    Jitraree, Arisara
    Homjan, Monpinya Charttong
    Laykhram, Thanachit
    Trakarnvanich, Thananda
    RENAL FAILURE, 2024, 46 (01)
  • [6] Why choose high volume online post-dilution hemodiafiltration?
    Basile, Carlo
    Davenport, Andrew
    Blankestijn, Peter J.
    JOURNAL OF NEPHROLOGY, 2017, 30 (02) : 181 - 186
  • [7] Selection of dilutional method for on-line HDF, pre- or post-dilution
    Masakane, I
    BLOOD PURIFICATION, 2004, 22 : 49 - 54
  • [8] High volume online post-dilution hemodiafiltration: how relevant is it in chronic kidney disease?
    Martins Castro, Manuel Carlos
    JORNAL BRASILEIRO DE NEFROLOGIA, 2022, 44 (02): : 238 - 243
  • [9] Evaluation of pre- and post-dilution continuous veno-venous hemofiltration on leukocyte and platelet function in patients with sepsis
    Chihara, Shinya
    Masuda, Yoshiki
    Tatsumi, Hiroomi
    Yamakage, Michiaki
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2019, 42 (01) : 9 - 16
  • [10] Who is the winner, pre-, post-, or mixed-dilution hemodiafiltration?
    Park, Hayne Cho
    Lee, Young-Ki
    KIDNEY RESEARCH AND CLINICAL PRACTICE, 2021, 40 (03) : 332 - 334