A paired emitter-detector diode-based photometer for the determination of sodium hypochlorite adulteration in milk

被引:7
作者
Bastan, Narges [1 ]
Ahmadi, Mazaher [1 ]
Madrakian, Tayyebeh [1 ]
Afkhami, Abbas [1 ]
Khalili, Sina [1 ]
Majidi, Mohsen [1 ]
Moradi, Mohammadreza [2 ]
机构
[1] Bu Ali Sina Univ, Fac Chem, Hamadan 6517838695, Iran
[2] Hamadan Univ Med Sci, Res Ctr Mol Med, Hamadan, Iran
关键词
ABSORPTION DETECTION; TRANSITION-METALS; CHLORINE; SENSITIVITY; PERSISTENCE; OXIDATION; DIETARY;
D O I
10.1038/s41598-023-33527-y
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
This paper reports on developing a low cost but efficient paired emitter-detector diode (PEDD)-based photometer. The photometer consists of a white light-emitting diode (LED) as the emitter diode, an RGB LED as the detector diode, and a multimeter for recoding the signal. The developed PEDD-based photometer was utilized for the determination of liquid bleach adulteration in cow milk samples. N,N-Diethyl-p-phenylenediamine sulfate aqueous solution of pH 6 was used as a probe to monitor the presence of residual active chlorine in milk. The results showed that the developed method could be used to determine sodium hypochlorite in the concentration range of 0.5 to 20.0 ppm Cl-2 with 0.14 and 0.46 ppm Cl-2 limit of detection and limit of quantification, respectively. The intraday and interday precisions of the method at two concentration levels of 5.5 and 13.7 ppm Cl-2 were 1.04% and 0.52%, and 1.81% and 1.02%, respectively. The recoveries of 114.2% and 106.9% were obtained for 5.5 and 13.7 ppm Cl-2 concentrations levels, respectively. Real sample analyzes results showed that "maybe" liquid bleach adulteration in milk is the case for local distributors of raw milk.
引用
收藏
页数:9
相关论文
共 53 条
[1]  
Amouzegar Z., 2022, MICRO NANOTECHNOLOGY, P17
[2]  
[Anonymous], 2014, An evaluation of the effectiveness of farmland protection policy in China
[3]   Separation of transition metals on a poly-iminodiacetic acid grafted polymeric resin column with post-column reaction detection utilising a paired emitter-detector diode system [J].
Barron, Leon ;
O'Toole, Martina ;
Diamond, Dermot ;
Nesterenko, Pavel N. ;
Paull, Brett .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1213 (01) :31-36
[4]   Analytical devices based on light-emitting diodes - a review of the state-of-the-art [J].
Bui, Duy Anh ;
Hauser, Peter C. .
ANALYTICA CHIMICA ACTA, 2015, 853 :46-58
[5]  
Burggraaff O., MICRO NANOTECHNOLOGY, VXII, P84
[6]   The SLIM spectrometer [J].
Cantrell, KM ;
Ingle, JD .
ANALYTICAL CHEMISTRY, 2003, 75 (01) :27-35
[7]   Oxidant/antioxidant balance in animal nutrition and health: the role of protein oxidation [J].
Celi, Pietro ;
Gabai, Gianfranco .
FRONTIERS IN VETERINARY SCIENCE, 2015, 2
[8]  
COOPER WJ, 1982, J AM WATER WORKS ASS, V74, P362
[9]   Preservatives and neutralizing substances in milk: analytical sensitivity of official specific and nonspecific tests, microbial inhibition effect, and residue persistence in milk [J].
Correa da Silva, Livia Cavaletti ;
Tamanini, Ronaldo ;
Pereira, Juliana Ramos ;
Rios, Edson Antonio ;
Ribeiro Junior, Jose Carlos ;
Beloti, Vanerli .
CIENCIA RURAL, 2015, 45 (09) :1613-1618
[10]  
Crocombe R.A., 2008, SPECTROSCOPY-US, V23, P38