Randomised crossover study on pulse oximeter readings from different sensors in very preterm infants

被引:4
|
作者
Maiwald, Christian Achim [1 ,2 ]
Schwarz, Christoph E. [2 ,3 ]
Boeckmann, Katrin [2 ]
Springer, Laila [2 ]
Poets, Christian F. [2 ,4 ]
Franz, Axel [1 ,2 ]
机构
[1] Univ Hosp Tubingen, Ctr Pediat Clin Studies CPCS, Dept Pediat, Tubingen, Germany
[2] Tubingen Univ Hosp, Dept Neonatol, Tubingen, Germany
[3] Heidelberg Univ, Dept Neonatol, Heidelberg, Germany
[4] Univ Hosp Tubingen, Dept Neonatol, Dept Pediat, D-72076 Tubingen, Germany
关键词
Intensive Care Units; Neonatal; Neonatology; EMERGING PNEUMONIA; SATURATION; ACCURACY; SEPSIS;
D O I
10.1136/archdischild-2023-325961
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objective In extremely preterm infants, different target ranges for pulse oximeter saturation (SpO(2)) may affect mortality and morbidity. Thus, the impact of technical changes potentially affecting measurements should be assessed. We studied SpO(2) readings from different sensors for systematic deviations.Design Single-centre, randomised, triple crossover study.Setting Tertiary neonatal intensive care unit.Patients: 24 infants, born at <32 weeks' gestation, with current weight <1500 g and without right-to-left shunt via a patent ductus arteriosus.Interventions Simultaneous readings from three SpO(2) sensors (Red Diamond (RD), Photoplethysmography (PPG), Low Noise Cabled Sensors (LNCS)) were logged at 0.5 Hz over 6 hour/infant and compared with LNCS as control using analysis of variance. Sensor position was randomly allocated and rotated every 2 hours. Seven different batches each were used.Outcomes Primary outcome was the difference in SpO(2) readings. Secondary outcomes were differences between sensors in the proportion of time within the SpO(2)-target range (90-95 (100)%).Results Mean gestational age at birth (+/- SD) was 27(4/7) (+/- 2(3/7)) weeks, postnatal age 20 (+/- 20) days. 134 hours of recording were analysed. Mean SpO(2) (+/- SD) was 94.0% (+/- 3.8; LNCS) versus 92.2% (+/- 4.0; RD; p<0.0001) and 94.5% (+/- 3.9; PPG; p<0.0001), respectively. Mean SpO(2) difference (95% CI) was -1.8% (-1.9 to -1.8; RD) and 0.5% (0.4 to 0.5; PPG). Proportion of time in target was significantly lower with RD sensors (84.8% vs 91.7%; p=0.0001) and similar with PPG sensors (91.1% vs 91.7%; p=0.63).Conclusion There were systematic differences in SpO(2) readings between RD sensors versus LNCS. These findings may impact mortality and morbidity of preterm infants, particularly when aiming for higher SpO(2)-target ranges (eg, 90-95%).
引用
收藏
页码:391 / 396
页数:6
相关论文
共 50 条
  • [1] Response to: 'Randomised crossover study on pulse oximeter readings from different sensors in very preterm infants' by Sharma et al
    Maiwald, Christian Achim
    Schwarz, Christoph E.
    Bockmann, Katrin
    Springer, Laila
    Poets, Christian F.
    Franz, Axel R.
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2024,
  • [2] Response to 'Randomised crossover study on pulse oximeter readings from different sensors in very preterm infants' by Maiwald et al
    Sharma, Vikrant
    Barker, Steven J.
    Sola, Augusto
    Cantillon, Daniel
    Sorci, Rebecca
    Wilson, William C.
    ARCHIVES OF DISEASE IN CHILDHOOD, 2025, 110 (01) : 110 - 111
  • [3] Response to 'Randomised crossover study on pulse oximeter readings from different sensors in very preterm infants' by Maiwald et al
    Sharma, Vikrant
    Barker, Steven J.
    Sola, Augusto
    Cantillon, Daniel
    Sorci, Rebecca
    Wilson, William C.
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2024,
  • [4] Racial discrepancy in pulse oximeter accuracy in preterm infants
    Zachary Vesoulis
    Anna Tims
    Hafsa Lodhi
    Natasha Lalos
    Halana Whitehead
    Journal of Perinatology, 2022, 42 : 79 - 85
  • [5] Racial discrepancy in pulse oximeter accuracy in preterm infants
    Vesoulis, Zachary
    Tims, Anna
    Lodhi, Hafsa
    Lalos, Natasha
    Whitehead, Halana
    JOURNAL OF PERINATOLOGY, 2022, 42 (01) : 79 - 85
  • [6] Effects of motion artifacts on pulse oximeter readings from different facial regions
    Nagre, A
    Mendelson, Y
    2005 IEEE 31ST ANNUAL NORTHEAST BIOENGINEERING CONFERENCE, 2005, : 220 - 222
  • [7] Influence of pulmonary factors on pulse oximeter saturation in preterm infants
    Jones, J. G.
    Lockwood, G. G.
    Fung, N.
    Lasenby, J.
    Ross-Russell, R. I.
    Quine, D.
    Stenson, B. J.
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2016, 101 (04): : F319 - F322
  • [8] Randomised crossover study comparing speed of heart rate display in newborns using ECG plus pulse oximeter versus pulse oximeter alone
    Murphy, Madeleine C.
    De Angelis, Laura
    Fitzgerald, Eoin
    McCarthy, Lisa K.
    ODonnell, Colm Patrick Finbarr
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2017, 102 (05): : F464 - F465
  • [9] Longevity of Masimo and Nellcor Pulse Oximeter Sensors in the Care of Infants
    Thomas Erler
    Stefan Avenarius
    Esther Wischniewski
    Katerina Schmidt
    Hans-Georg Kläber
    Journal of Perinatology, 2003, 23 (2) : 133 - 135
  • [10] Randomised crossover triad of different postural interventions on bradycardia and intermittent hypoxia in preterm infants
    Reher, C.
    Kuny, K. D.
    Pantalltschka, T.
    Urschitz, M. S.
    Poets, C. F.
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2008, 93 (04): : F289 - F291