Comparison of quantitative [11C]PE2I brain PET studies between an integrated PET/MR and a stand-alone PET system

被引:0
作者
Sousa, Joao M. [1 ,8 ]
Appel, Lieuwe [1 ,2 ]
Engstrom, Mathias [3 ]
Nyholm, Dag [4 ,5 ]
Ahlstrom, Hakan [1 ,2 ,6 ]
Lubberink, Mark [1 ,7 ]
机构
[1] Uppsala Univ, Dept Surg Sci Nucl Med & PET, Uppsala, Sweden
[2] Uppsala Univ Hosp, Med Imaging Ctr, Uppsala, Sweden
[3] Collect Minds Radiol AB, Taby, Sweden
[4] Uppsala Univ Hosp, Dept Neurol, Uppsala, Sweden
[5] Uppsala Univ, Dept Med Sci, Neurol, Uppsala, Sweden
[6] Antaros Med AB, BioVenture Hub, Molndal, Sweden
[7] Uppsala Univ Hosp, Med Phys, Uppsala, Sweden
[8] Uppsala Univ, Uppsala Univ Hosp, Dept Surg Sci, PET Ctr, S-75185 Uppsala, Sweden
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2024年 / 117卷
基金
瑞典研究理事会;
关键词
Positron emission tomography; Dopamine transporter; Cerebral blood flow; PET quantification; ATTENUATION CORRECTION; PERFORMANCE EVALUATION; MAGNETIC-RESONANCE; ACOUSTIC NOISE; ATLAS; BINDING; HEAD;
D O I
10.1016/j.ejmp.2023.103185
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PET/MR systems demanded great efforts for accurate attenuation correction (AC) but differences in technology, geometry and hardware attenuation may also affect quantitative results. Dedicated PET systems using transmission-based AC are regarded as the gold standard for quantitative brain PET. The study aim was to investigate the agreement between quantitative PET outcomes from a PET/MR scanner against a stand-alone PET system.Nine patients with Parkinsonism underwent two 80-min dynamic PET scans with the dopamine transporter ligand [11C]PE2I. Images were reconstructed with resolution-matched settings using 68Ge-transmission (standalone PET), and zero-echo-time MR (PET/MR) scans for AC. Non-displaceable binding potential (BPND) and relative delivery (R1) were evaluated using volumes of interest and voxel-wise analysis.Correlations between systems were high (r >= 0.85) for both quantitative outcome parameters in all brain regions. Striatal BPND was significantly lower on PET/MR than on stand-alone PET (-7%). R1 was significantly overestimated in posterior cortical regions (9%) and underestimated in striatal (-9%) and limbic areas (-6%). The voxel-wise evaluation revealed that the MR-safe headphones caused a negative bias in both parametric BPND and R1 images. Additionally, a significant positive bias of R1 was found in the auditory cortex, most likely due to the acoustic background noise during MR imaging. The relative bias of the quantitative [11C]PE2I PET data acquired from a SIGNA PET/MR system was in the same order as the expected test-retest reproducibility of [11C]PE2I BPND and R1, compared to a stand-alone ECAT PET scanner. MR headphones and background noise are potential sources of error in functional PET/MR studies.
引用
收藏
页数:9
相关论文
共 52 条
  • [1] Performance evaluation of the whole-body PET scanner ECAT EXACT HR+ following the IEC standard
    Adam, LE
    Zaers, J
    Ostertag, H
    Trojan, H
    Bellemann, ME
    Brix, G
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) : 1172 - 1179
  • [2] Auditory intensity processing: Effect of MRI background noise
    Angenstein, Nicole
    Stadler, Joerg
    Brechmann, Andre
    [J]. HEARING RESEARCH, 2016, 333 : 87 - 92
  • [3] Use of 11C-PE2I PET in Differential Diagnosis of Parkinsonian Disorders
    Appel, Lieuwe
    Jonasson, My
    Danfors, Torsten
    Nyholm, Dag
    Askmark, Hakan
    Lubberink, Mark
    Soerensen, Jens
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (02) : 234 - 242
  • [4] Transmission scanning in emission tomography
    Bailey, DL
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (07) : 774 - 787
  • [5] PET/MR in Dementia and Other Neurodegenerative Diseases
    Barthel, Henryk
    Schroeter, Matthias L.
    Hoffmann, Karl-Titus
    Sabri, Osama
    [J]. SEMINARS IN NUCLEAR MEDICINE, 2015, 45 (03) : 224 - 233
  • [6] Brix G, 1997, J NUCL MED, V38, P1614
  • [7] Impact of MR-safe headphones on PET attenuation in combined PET/MRI scans
    Buether, Florian
    Vrachimis, Alexis
    Becker, Anne
    Stegger, Lars
    [J]. EJNMMI RESEARCH, 2016, 6
  • [8] NEMA NU 2-2007 performance characteristics of GE Signa integrated PET/MR for different PET isotopes
    Caribe, Paulo R. R. V.
    Koole, M.
    D'Asseler, Yves
    Deller, Timothy W.
    Van Laere, K.
    Vandenberghe, S.
    [J]. EJNMMI PHYSICS, 2019, 6 (1)
  • [9] CARSON RE, 1988, J NUCL MED, V29, P1558
  • [10] Current commercial techniques for MRI-guided attenuation correction are insufficient and will limit the wider acceptance of PET/MRI technology in the clinic
    Catana, Ciprian
    Quick, Harald H.
    Zaidi, Habib
    [J]. MEDICAL PHYSICS, 2018, 45 (09) : 4007 - 4010