Validation of a digital method for patient-specific verification of VMAT treatment using a 2D ionisation detector array

被引:1
|
作者
U'wais, Fatima Adel [1 ]
Radzi, Yasmin [2 ]
Rizan, Naslinda Noor [3 ]
Zin, Hafiz M. [1 ]
机构
[1] Univ Sains Malaysia, Adv Med & Dent Inst, Kepala Batas 13200, Pulau Pinang, Malaysia
[2] Univ Sains Malaysia, Sch Phys, George Town 11800, Malaysia
[3] Gleneagles Hosp, George Town, Malaysia
关键词
VMAT; Patient-specific QA; Log file; QUALITY-ASSURANCE; LOG FILES; MLC; THERAPY; ACCURACY; QA;
D O I
10.1016/j.radphyschem.2022.110536
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
VMAT treatment involves the delivery of complex beam parameters including the dynamic multi-leaf collimators (MLC) from a linac. The use of a detector array has become a standard method to verify the treatment accuracy before the treatment planned is delivered to the patient with some limitations. This study aims to evaluate the accuracy of a high-resolution, digital method using linac log data for VMAT treatment verification. The log data was obtained from a Versa HD (Elekta ltd., Crawley, UK) linac during VMAT treatment. All the dynamic treatment parameters including MLC position, and MLC speed, were recorded at the rate of 4 Hz in the log data. The treatment delivery was also measured simultaneously using an Octavius 1500 2D ionisation chamber array (PTW, Freiburg, Germany) during the log data measurements. The parameters were extracted and analysed using algorithms written in MATLAB (MathWorks, Natick, MA). The results demonstrated that the log data can track MLC leaves with an accuracy of 1 mm at speeds ranging from 3.04 to 3.40 cm/s. Evaluation of the fluence generated for the VMAT delivery using log data was shown to agree well with the planned dose distribution measured in the 2D detector array, with an average gamma pass rate of 97.5% at 3%/3 mm. Log data obtained a higher gamma pass rate of 98.5% compared to the 2D detector array of 97.5%. The digital linac log data provides the basis for an essential high-resolution real-time verification tool that may be used for routine VMAT verification.
引用
收藏
页数:7
相关论文
共 43 条
  • [41] Design of a QA method to characterize submillimeter-sized PBS beam properties using a 2D ionization chamber array
    Lin, Yuting
    Bentefour, Hassan
    Flanz, Jacob
    Kooy, Hanne
    Clasie, Benjamin
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (10):
  • [42] Validation of a method for in vivo 3D dose reconstruction for IMRT and VMAT treatments using on-treatment EPID images and a model-based forward-calculation algorithm
    Van Uytven, Eric
    Van Beek, Timothy
    McCowan, Peter M.
    Chytyk-Praznik, Krista
    Greer, Peter B.
    McCurdy, Boyd M. C.
    MEDICAL PHYSICS, 2015, 42 (12) : 6945 - 6954
  • [43] Precision of computed tomography and cartilage-reproducing image reconstruction method in generating digital model for potential use in 3D printing of patient-specific radial head prosthesis: a human cadaver study
    Luenam, Suriya
    Bantuchai, Theeraset
    Kosiyatrakul, Arkaphat
    Chanpoo, Malee
    Phakdeewisetkul, Kantapat
    Puncreobutr, Chedtha
    3D PRINTING IN MEDICINE, 2021, 7 (01)