Validation of a digital method for patient-specific verification of VMAT treatment using a 2D ionisation detector array

被引:1
作者
U'wais, Fatima Adel [1 ]
Radzi, Yasmin [2 ]
Rizan, Naslinda Noor [3 ]
Zin, Hafiz M. [1 ]
机构
[1] Univ Sains Malaysia, Adv Med & Dent Inst, Kepala Batas 13200, Pulau Pinang, Malaysia
[2] Univ Sains Malaysia, Sch Phys, George Town 11800, Malaysia
[3] Gleneagles Hosp, George Town, Malaysia
关键词
VMAT; Patient-specific QA; Log file; QUALITY-ASSURANCE; LOG FILES; MLC; THERAPY; ACCURACY; QA;
D O I
10.1016/j.radphyschem.2022.110536
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
VMAT treatment involves the delivery of complex beam parameters including the dynamic multi-leaf collimators (MLC) from a linac. The use of a detector array has become a standard method to verify the treatment accuracy before the treatment planned is delivered to the patient with some limitations. This study aims to evaluate the accuracy of a high-resolution, digital method using linac log data for VMAT treatment verification. The log data was obtained from a Versa HD (Elekta ltd., Crawley, UK) linac during VMAT treatment. All the dynamic treatment parameters including MLC position, and MLC speed, were recorded at the rate of 4 Hz in the log data. The treatment delivery was also measured simultaneously using an Octavius 1500 2D ionisation chamber array (PTW, Freiburg, Germany) during the log data measurements. The parameters were extracted and analysed using algorithms written in MATLAB (MathWorks, Natick, MA). The results demonstrated that the log data can track MLC leaves with an accuracy of 1 mm at speeds ranging from 3.04 to 3.40 cm/s. Evaluation of the fluence generated for the VMAT delivery using log data was shown to agree well with the planned dose distribution measured in the 2D detector array, with an average gamma pass rate of 97.5% at 3%/3 mm. Log data obtained a higher gamma pass rate of 98.5% compared to the 2D detector array of 97.5%. The digital linac log data provides the basis for an essential high-resolution real-time verification tool that may be used for routine VMAT verification.
引用
收藏
页数:7
相关论文
共 47 条
[1]  
AAPM, 2009, DIFFERENCES, V21, P308
[2]   Estimation of patients organ doses and staff exposure during bone scan examination [J].
Ali, W. ;
Sulieman, A. ;
Tamam, Nissren ;
Boshara, N. ;
Aldhebaib, A. ;
Alkhorayef, M. ;
Khandaker, Mayeen Uddin ;
Bradley, D. A. .
RADIATION PHYSICS AND CHEMISTRY, 2021, 188
[3]   A comprehensive tool to analyse dynamic log files from an Elekta-Synergy accelerator [J].
Arumugam, Sankar ;
Xing, Aitang ;
Pagulayan, Claire ;
Holloway, Lois .
XVII INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY (ICCR 2013), 2014, 489
[4]   Insensitivity of machine log files to MLC leaf backlash and effect of MLC backlash on clinical dynamic MLC motion: An experimental investigation [J].
Barnes, Michael ;
Pomare, Dennis ;
Doebrich, Marcus ;
Standen, Therese S. ;
Wolf, Joshua ;
Greer, Peter ;
Simpson, John .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (09)
[5]  
Calvo-ortega J.F., 2014, IMRT QA, V15, P5
[6]  
Ceylan C., 2021, EFFECT MULTILEAF COL, DOI [10.7759/cureus.12684, DOI 10.7759/CUREUS.12684]
[7]   Effect of Multileaf Collimator Leaf Position Error Determined by Picket Fence Test on Gamma Index Value in Patient-Specific Quality Assurance of Volumetric-Modulated Arc Therapy Plans [J].
Ceylan, Cemile ;
Inal, Serpil Yondem ;
Senol, Elif ;
Yilmaz, Berrin ;
Sahin, Sevim .
CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (01)
[8]   Investigation of error detection capabilities of phantom, EPID and MLC log file based IMRT QA methods [J].
Defoor, Dewayne L. ;
Stathakis, Sotirios ;
Roring, Joseph E. ;
Kirby, Neil A. ;
Mavroidis, Panayiotis ;
Obeidat, Mohammad ;
Papanikolaou, Nikos .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (04) :172-179
[9]   Anatomy-based, patient-specific VMAT QA using EPID or MLC log files [J].
Defoor, Dewayne L. ;
Vazquez-Quino, Luis A. ;
Mavroidis, Panayiotis ;
Papanikolaou, Nikos ;
Stathakis, Sotirios .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03) :206-215
[10]   Commissioning of volumetric modulated arc therapy (VMAT) in a dual-vendor environment [J].
Dobler, Barbara ;
Groeger, Christian ;
Treutwein, Marius ;
Alvarez-Moret, Judith ;
Goetzfried, Thomas ;
Weidner, Karin ;
Haertl, Petra ;
Koelbl, Oliver .
RADIOTHERAPY AND ONCOLOGY, 2011, 99 (01) :86-89