Impacts of Skin Color and Hypoxemia on Noninvasive Assessment of Peripheral Blood Oxygen Saturation: A Scoping Review

被引:10
|
作者
Aoki, Kawaiola C. [1 ]
Barrant, Maya [1 ]
Gai, Mam Jarra [1 ]
Handal, Marina [1 ]
Xu, Vivian [1 ]
Mayrovitz, Harvey N. [2 ]
机构
[1] Nova Southeastern Univ, Dr Kiran C Patel Coll Osteopath Med, Med Sch, Ft Lauderdale, FL 33314 USA
[2] Nova Southeastern Univ, Dr Kiran C Patel Coll Allopath Med, Med Educ & Simulat, Cardiopulm Physiol, Ft Lauderdale, FL 33328 USA
关键词
racial disparity; racial bias; skin color; occult hypoxia; electronic monitoring devices; false hypoxemia; skin pigmentation; pulse oximetry; PULSE OXIMETRY; ACCURACY; PIGMENTATION; DISPARITIES; HEMOGLOBIN; AGREEMENT; MELANIN;
D O I
10.7759/cureus.46078
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Standard pulse oximeters estimate arterial blood saturation (SaO(2)) non-invasively by emitting and detecting light of a specific wavelength through a cutaneous vascular bed, such as a digit or the ear lobe. The quantity measured at these peripheral sites is designated as oxygen saturation (SpO(2)). Most reliable pulse oximeters are calibrated from measurements of healthy volunteers using some form of oxygen desaturation method. As the degree of inducible hypoxemia is limited, the calibration below achievable desaturation levels is usually extrapolated, leading to potential measurement error at low SaO(2) values, especially in highly pigmented skin. Such skin color-related errors (SCRE) are the topic of this scoping review. Specifically, this study aimed to identify the combined impact of skin color and reduced SaO2 on the non-invasive assessment of SpO(2) and report the consequences of potential inaccuracies. Three databases were searched (Cumulated Index to Nursing and Allied Health Literature (CINAHL), PubMed, and Web of Science) for peer-reviewed prospective and retrospective studies published in English between 2000 and 2022 involving human patients with hypoxemia that included a measure of skin color (Fitzpatrick scale or race/ethnicity). Ten studies met the criteria and were included in the final review. Eight of these studies reported statistically significant higher pulse oximeter readings in darker-skinned patients with hypoxia compared to their arterial blood gas measurements. Occult hypoxia was more prevalent in Black and Hispanic patients than in White patients. Minority patients overall (Black, Asian, and American Indian) were more likely to have a SaO(2) < 88% that was not detected by pulse oximetry (occult hypoxemia) during hospitalization. With greater levels of hypoxemia, the differences between SpO(2) and SaO(2) were greater. If SaO(2) was < 90%, then SpO(2) was overestimated in all ethnicities but worse in minorities. In conclusion, the bias found in pulse oximeter readings in the skin of color broadly impacts patients with hypoxemia. The failure of SpO(2) measuring devices to detect occult hypoxemia can delay the delivery of lifesaving treatment to critically ill patients requiring respiratory rehabilitation and supplemental oxygen therapy. This may lead to adverse health outcomes, increased in-hospital mortality, and complications such as organ dysfunction. An improvement in pulse oximeter detection mechanisms that would include all skin pigmentations is therefore much desired to optimize individual healthcare status and minimize disparities in treatment.
引用
收藏
页数:7
相关论文
共 24 条
  • [1] Exploring the bias: how skin color influences oxygen saturation readings via Monte Carlo simulations
    Swamy, Suvvi K. Narayana
    Liu, Chong
    Correia, Ricardo
    Hayes-Gill, Barrie R.
    Morgan, Stephen P.
    JOURNAL OF BIOMEDICAL OPTICS, 2024, 29
  • [2] Objective Assessment and Quantification of Skin Color and Melanin in Neonates and Infants: A State-Of-The-Art Review
    Sharma, Megha
    Pickhardt, Abigail Jo
    Madison, Lauren
    Tong, Lauren
    Evans, Megan S.
    PEDIATRIC DERMATOLOGY, 2025,
  • [3] NONINVASIVE METHODS OF MEASURING OXYGEN SATURATION IN VENOUS BLOOD
    Yesman, S. S.
    Mamilov, S. A.
    Asimov, M. M.
    Gisbreht, A. I.
    JOURNAL OF APPLIED SPECTROSCOPY, 2011, 78 (03) : 406 - 413
  • [4] Noninvasive methods of measuring oxygen saturation in venous blood
    S. S. Yesman
    S. A. Mamilov
    M. M. Asimov
    A. I. Gisbreht
    Journal of Applied Spectroscopy, 2011, 78 : 406 - 413
  • [5] Accuracy of pulse oximeters in measuring oxygen saturation in patients with poor peripheral perfusion: a systematic review
    Poorzargar, Khashayar
    Pham, Chi
    Ariaratnam, Jennita
    Lee, Kang
    Parotto, Matteo
    Englesakis, Marina
    Chung, Frances
    Nagappa, Mahesh
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2022, 36 (04) : 961 - 973
  • [6] Oxygen saturation targets for adults with acute hypoxemia in low and lower-middle income countries: a scoping review with analysis of contextual factors
    Herbst, Austin
    Goel, Swati
    Beane, Abi
    Brotherton, B. Jason
    Dula, Dingase
    Ely, E. Wesley
    Gordon, Stephen B.
    Haniffa, Rashan
    Hedt-Gauthier, Bethany
    Limbani, Felix
    Lipnick, Michael S.
    Lyon, Samuel
    Njoki, Carolyne
    Oduor, Peter
    Otieno, George
    Pisani, Luigi
    Rylance, Jamie
    Shrime, Mark G.
    Uwamahoro, Doris Lorette
    Vanderburg, Sky
    Waweru-Siika, Wangari
    Twagirumugabe, Theogene
    Riviello, Elisabeth
    FRONTIERS IN MEDICINE, 2023, 10
  • [7] Measurement of blood-oxygen saturation using a photoacoustic technique in the rabbit hypoxemia model
    Sei, Kiguna
    Fujita, Masanori
    Hirasawa, Takeshi
    Okawa, Shinpei
    Kushibiki, Toshihiro
    Sasa, Hidenori
    Furuya, Kenichi
    Ishihara, Miya
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2019, 33 (02) : 269 - 279
  • [8] The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: a systematic review and meta-analysis
    Chunhu Shi
    Mark Goodall
    Jo Dumville
    James Hill
    Gill Norman
    Oliver Hamer
    Andrew Clegg
    Caroline Leigh Watkins
    George Georgiou
    Alexander Hodkinson
    Catherine Elizabeth Lightbody
    Paul Dark
    Nicky Cullum
    BMC Medicine, 20
  • [9] AI-Based Noninvasive Blood Glucose Monitoring: Scoping Review
    Chan, Pin Zhong
    Jin, Eric
    Jansson, Miia
    Chew, Han Shi Jocelyn
    JOURNAL OF MEDICAL INTERNET RESEARCH, 2024, 26
  • [10] The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: a systematic review and meta-analysis
    Shi, Chunhu
    Goodall, Mark
    Dumville, Jo
    Hill, James
    Norman, Gill
    Hamer, Oliver
    Clegg, Andrew
    Watkins, Caroline Leigh
    Georgiou, George
    Hodkinson, Alexander
    Lightbody, Catherine Elizabeth
    Dark, Paul
    Cullum, Nicky
    BMC MEDICINE, 2022, 20 (01)