Quantifying lumen diameter in coronary artery stents with high-resolution photon counting detector CT and convolutional neural network denoising

被引:9
作者
Koons, Emily K. K. [1 ,2 ]
Thorne, Jamison E. E. [1 ]
Huber, Nathan R. R. [1 ]
Chang, Shaojie [1 ]
Rajendran, Kishore [1 ]
McCollough, Cynthia H. H. [1 ]
Leng, Shuai [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
[2] Mayo Clin, Dept Biomed Engn & Physiol, Rochester, MN 55905 USA
基金
美国国家卫生研究院;
关键词
computed tomography; convolutional neural network; coronary artery stents; denoising; photon counting detector; RESTENOSIS;
D O I
10.1002/mp.16415
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundSmall coronary arteries containing stents pose a challenge in CT imaging due to metal-induced blooming artifact. High spatial resolution imaging capability is as the presence of highly attenuating materials limits noninvasive assessment of luminal patency. PurposeThe purpose of this study was to quantify the effective lumen diameter within coronary stents using a clinical photon-counting-detector (PCD) CT in concert with a convolutional neural network (CNN) denoising algorithm, compared to an energy-integrating-detector (EID) CT system. MethodsSeven coronary stents of different materials and inner diameters between 3.43 and 4.72 mm were placed in plastic tubes of diameters 3.96-4.87 mm containing 20 mg/mL of iodine solution, mimicking stented contrast-enhanced coronary arteries. Tubes were placed parallel with or perpendicular to the scanner's z-axis in an anthropomorphic phantom emulating an average-sized patient and scanned with a clinical EID-CT and PCD-CT. EID scans were performed using our standard coronary computed tomography angiography (cCTA) protocol (120 kV, 180 quality reference mAs). PCD scans were performed using the ultra-high-resolution (UHR) mode (120 x 0.2 mm collimation) at 120 kV with tube current adjusted so that CTDIvol was matched to that of EID scans. EID images were reconstructed per our routine clinical protocol (Br40, 0.6 mm thickness), and with the sharpest available kernel (Br69). PCD images were reconstructed at a thickness of 0.6 mm and a dedicated sharp kernel (Br89) which is only possible with the PCD UHR mode. To address increased image noise introduced by the Br89 kernel, an image-based CNN denoising algorithm was applied to the PCD images of stents scanned parallel to the scanner's z-axis. Stents were segmented based on full-width half maximum thresholding and morphological operations, from which effective lumen diameter was calculated and compared to reference sizes measured with a caliper. ResultsSubstantial blooming artifacts were observed on EID Br40 images, resulting in larger stent struts and reduced lumen diameter (effective diameter underestimated by 41% and 47% for parallel and perpendicular orientations, respectively). Blooming artifacts were observed on EID Br69 images with 19% and 31% underestimation of lumen diameter compared to the caliper for parallel and perpendicular scans, respectively. Overall image quality was substantially improved on PCD, with higher spatial resolution and reduced blooming artifacts, resulting in the clearer delineation of stent struts. Effective lumen diameters were underestimated by 9% and 19% relative to the reference for parallel and perpendicular scans, respectively. CNN reduced image noise by about 50% on PCD images without impacting lumen quantification (<0.3% difference). ConclusionThe PCD UHR mode improved in-stent lumen quantification for all seven stents as compared to EID images due to decreased blooming artifacts. Implementation of CNN denoising algorithms to PCD data substantially improved image quality.
引用
收藏
页码:4173 / 4181
页数:9
相关论文
共 50 条
  • [41] Quantum Iterative Reconstruction for Low-Dose Ultra-High-Resolution Photon-Counting Detector CT of the Lung
    Sartoretti, Thomas
    Racine, Damien
    Mergen, Victor
    Jungblut, Lisa
    Monnin, Pascal
    Flohr, Thomas G.
    Martini, Katharina
    Frauenfelder, Thomas
    Alkadhi, Hatem
    Euler, Andre
    DIAGNOSTICS, 2022, 12 (02)
  • [42] Potential of employing a quantum iterative reconstruction algorithm for ultra-high-resolution photon-counting detector CT of the hip
    Huflage, H.
    Grunz, J. -P
    Kunz, A. S.
    Patzer, T. S.
    Sauer, S. T.
    Christner, S. A.
    Petritsch, B.
    Erguen, S.
    Bley, T. A.
    Luetkens, K. S.
    RADIOGRAPHY, 2023, 29 (01) : 44 - 49
  • [43] Lung Nodule Volume Quantification and Shape Differentiation with an Ultra-High Resolution Technique on a Photon Counting Detector CT System
    Zhou, W.
    Montoya, J.
    Gutjahr, R.
    Ferrero, A.
    Halaweish, A.
    Kappler, S.
    McCollough, C.
    Leng, S.
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [44] Adaptive Feedback Convolutional-Neural-Network-Based High-Resolution Reflection-Waveform Inversion
    Wu, Yulang
    McMechan, George A.
    Wang, Yanfei
    JOURNAL OF GEOPHYSICAL RESEARCH-SOLID EARTH, 2022, 127 (06)
  • [45] Optimization of the Reconstruction Settings for Low-Dose Ultra-High-Resolution Photon-Counting Detector CT of the Lungs
    Graafen, Dirk
    Halfmann, Moritz C.
    Emrich, Tilman
    Yang, Yang
    Kreuter, Michael
    Dueber, Christoph
    Kloeckner, Roman
    Mueller, Lukas
    Jorg, Tobias
    DIAGNOSTICS, 2023, 13 (23)
  • [46] High-resolution Remote Sensing Vehicle Automatic Detection Based on Feature Fusion Convolutional Neural Network
    Li, Xin
    Guo, Kai
    Subei, Mutailifu
    Guo, Dudu
    INTERNATIONAL CONFERENCE ON COMPUTER VISION, APPLICATION, AND DESIGN (CVAD 2021), 2021, 12155
  • [47] High-Resolution Chest Computed Tomography Imaging of the Lungs Impact of 1024 Matrix Reconstruction and Photon-Counting Detector Computed Tomography
    Bartlett, David J.
    Koo, Chi Wan
    Bartholmai, Brian J.
    Rajendran, Kishore
    Weaver, Jayse M.
    Halaweish, Ahmed E.
    Leng, Shuai
    McCollough, Cynthia H.
    Fletcher, Joel G.
    INVESTIGATIVE RADIOLOGY, 2019, 54 (03) : 129 - 137
  • [48] A Blooming correction technique for improved vasa vasorum detection using an ultra-high-resolution photon-counting detector CT
    Marsh, Jeffrey, Jr.
    Rajendran, Kishore
    Tao, Shengzhen
    Vercnocke, Andrew
    Anderson, Jill
    Leng, Shuai
    Ritman, Erik
    McCollough, Cynthia
    MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312
  • [49] High Resolution, Full Field-of-View, Whole Body Photon-Counting Detector CT: System Assessment and Initial Experience
    Rajendran, Kishore
    Marsh, Jeffrey
    Petersilka, Martin
    Henning, Andre
    Shanblatt, Elisabeth
    Schmidt, Bernhard
    Flohr, Thomas
    Fletcher, Joel
    McCollough, Cynthia
    Leng, Shuai
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [50] First in-human quantitative plaque characterization with ultra-high resolution coronary photon-counting CT angiography
    Mergen, Victor
    Eberhard, Matthias
    Manka, Robert
    Euler, Andre
    Alkadhi, Hatem
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2022, 9