Quantifying lumen diameter in coronary artery stents with high-resolution photon counting detector CT and convolutional neural network denoising

被引:9
|
作者
Koons, Emily K. K. [1 ,2 ]
Thorne, Jamison E. E. [1 ]
Huber, Nathan R. R. [1 ]
Chang, Shaojie [1 ]
Rajendran, Kishore [1 ]
McCollough, Cynthia H. H. [1 ]
Leng, Shuai [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
[2] Mayo Clin, Dept Biomed Engn & Physiol, Rochester, MN 55905 USA
基金
美国国家卫生研究院;
关键词
computed tomography; convolutional neural network; coronary artery stents; denoising; photon counting detector; RESTENOSIS;
D O I
10.1002/mp.16415
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundSmall coronary arteries containing stents pose a challenge in CT imaging due to metal-induced blooming artifact. High spatial resolution imaging capability is as the presence of highly attenuating materials limits noninvasive assessment of luminal patency. PurposeThe purpose of this study was to quantify the effective lumen diameter within coronary stents using a clinical photon-counting-detector (PCD) CT in concert with a convolutional neural network (CNN) denoising algorithm, compared to an energy-integrating-detector (EID) CT system. MethodsSeven coronary stents of different materials and inner diameters between 3.43 and 4.72 mm were placed in plastic tubes of diameters 3.96-4.87 mm containing 20 mg/mL of iodine solution, mimicking stented contrast-enhanced coronary arteries. Tubes were placed parallel with or perpendicular to the scanner's z-axis in an anthropomorphic phantom emulating an average-sized patient and scanned with a clinical EID-CT and PCD-CT. EID scans were performed using our standard coronary computed tomography angiography (cCTA) protocol (120 kV, 180 quality reference mAs). PCD scans were performed using the ultra-high-resolution (UHR) mode (120 x 0.2 mm collimation) at 120 kV with tube current adjusted so that CTDIvol was matched to that of EID scans. EID images were reconstructed per our routine clinical protocol (Br40, 0.6 mm thickness), and with the sharpest available kernel (Br69). PCD images were reconstructed at a thickness of 0.6 mm and a dedicated sharp kernel (Br89) which is only possible with the PCD UHR mode. To address increased image noise introduced by the Br89 kernel, an image-based CNN denoising algorithm was applied to the PCD images of stents scanned parallel to the scanner's z-axis. Stents were segmented based on full-width half maximum thresholding and morphological operations, from which effective lumen diameter was calculated and compared to reference sizes measured with a caliper. ResultsSubstantial blooming artifacts were observed on EID Br40 images, resulting in larger stent struts and reduced lumen diameter (effective diameter underestimated by 41% and 47% for parallel and perpendicular orientations, respectively). Blooming artifacts were observed on EID Br69 images with 19% and 31% underestimation of lumen diameter compared to the caliper for parallel and perpendicular scans, respectively. Overall image quality was substantially improved on PCD, with higher spatial resolution and reduced blooming artifacts, resulting in the clearer delineation of stent struts. Effective lumen diameters were underestimated by 9% and 19% relative to the reference for parallel and perpendicular scans, respectively. CNN reduced image noise by about 50% on PCD images without impacting lumen quantification (<0.3% difference). ConclusionThe PCD UHR mode improved in-stent lumen quantification for all seven stents as compared to EID images due to decreased blooming artifacts. Implementation of CNN denoising algorithms to PCD data substantially improved image quality.
引用
收藏
页码:4173 / 4181
页数:9
相关论文
共 50 条
  • [21] Comparison of Radiation Dose and Image Quality of Pediatric High-Resolution Chest CT Between Photon-Counting Detector CT and Energy-Integrated Detector CT: A Matched Study
    Siegel, Marilyn J.
    Bugenhagen, Scott M.
    Sanchez, Adrian
    Kim, Stacy
    Abadia, Andres
    Ramirez-Giraldo, Juan C.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2023, 221 (03) : 363 - 371
  • [22] Intra-individual comparison of coronary artery stenosis measurements between energy-integrating detector CT and photon-counting detector CT
    Wolf, Elias V.
    Gnasso, Chiara
    Schoepf, U. Joseph
    Halfmann, Moritz C.
    O'Doherty, Jim
    Zsarnoczay, Emese
    Varga-Szemes, Akos
    Emrich, Tilman
    Fink, Nicola
    IMAGING, 2023, 15 (02): : 61 - 68
  • [23] Simultaneous denoising and spatial resolution enhancement using convolutional neural network-based linear model in diagnostic CT images
    Yim, Dobin
    Kim, Burnyoung
    Lee, Seungwan
    MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312
  • [24] Ultra-high-resolution spectral silicon-based photon-counting detector CT for coronary CT angiography: Initial results in a dynamic phantom
    Holmes, Thomas Wesley
    Yin, Zhye
    Stevens, Grant M.
    Slavic, Scott
    Okerlund, Darin R.
    Maltz, Jonathan S.
    Pourmorteza, Amir
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2023, 17 (05) : 341 - 344
  • [25] Coronary artery stenosis quantification in patients with dense calcifications using ultra-high-resolution photon-counting-detector computed tomography
    Koons, Emily K.
    Rajiah, Prabhakar Shantha
    Thorne, Jamison E.
    Weber, Nikkole M.
    Kasten, Holly J.
    Shanblatt, Elisabeth R.
    Mccollough, Cynthia H.
    Leng, Shuai
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2024, 18 (01) : 56 - 61
  • [26] Coronary artery disease detection using deep learning and ultrahigh-resolution photon-counting coronary CT angiography
    Brendel, Jan M.
    Walterspiel, Jonathan
    Hagen, Florian
    Kuebler, Jens
    Brendlin, Andreas S.
    Afat, Saif
    Paul, Jean-Francois
    Kuestner, Thomas
    Nikolaou, Konstantin
    Gawaz, Meinrad
    Greulich, Simon
    Krumm, Patrick
    Winkelmann, Moritz T.
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2025, 106 (02) : 68 - 75
  • [27] Ultra-High Spatial Resolution, Multi-Energy CT using Photon Counting Detector Technology
    Leng, S.
    Gutjahr, R.
    Ferrero, A.
    Kappler, S.
    Henning, A.
    Halaweish, A.
    Zhou, W.
    Montoya, J.
    McCollough, C.
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [28] Low-dose high-resolution chest CT in adults with cystic fibrosis: intraindividual comparison between photon-counting and energy-integrating detector CT
    Frings, Marko
    Welsner, Matthias
    Mousa, Christin
    Zensen, Sebastian
    Salhoefer, Luca
    Meetschen, Mathias
    Beck, Nikolas
    Bos, Denise
    Westhoelter, Dirk
    Wienker, Johannes
    Taube, Christian
    Umutlu, Lale
    Schaarschmidt, Benedikt M.
    Forsting, Michael
    Haubold, Johannes
    Sutharsan, Sivagurunathan
    Opitz, Marcel
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2024, 8 (01)
  • [29] A new era of high-resolution CT diagnostics of the lung: improved image quality, detailed morphology, and reduced radiation dose with high-resolution photon-counting CT of the lungs compared to high-resolution energy-integrated CT
    Aurumskjold, Marie-Louise
    Sjunnesson, Lotta
    Pistea, Adrian
    Asbjornsson, Gylfi
    Wellman, Fredrik
    Bozovic, Gracijela
    ACTA RADIOLOGICA, 2024, 65 (10) : 1211 - 1221
  • [30] A simulation study of high-resolution x-ray computed tomography imaging using irregular sampling with a photon-counting detector
    Lee, Seungwan
    Choi, Yu-Na
    Kim, Hee-Joung
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 726 : 175 - 180