ADEPT: Accurate Diffusion Echo-Planar imaging with multi-contrast shoTs

被引:3
作者
Shafieizargar, Banafshe [1 ,2 ]
Jeurissen, Ben [1 ,2 ]
Poot, Dirk H. J. [3 ]
Klein, Stefan [3 ]
Van Audekerke, Johan [2 ,4 ]
Verhoye, Marleen [2 ,4 ]
den Dekker, Arnold J. [1 ,2 ]
Sijbers, Jan [1 ,2 ]
机构
[1] Univ Antwerp, Dept Phys, Iimec Vis Lab, Univ Pl 1,DN 103, B-2610 Antwerp, Belgium
[2] Univ Antwerp, NEURO Res Ctr Excellence, Antwerp, Belgium
[3] Erasmus MC, Dept Radiol & Nucl Med, Biomed Imaging Grp Rotterdam, Rotterdam, Netherlands
[4] Univ Antwerp, Dept Biomed Sci, Bioimaging Lab, Antwerp, Belgium
基金
欧盟地平线“2020”;
关键词
diffusion MRI; model-based reconstruction; multi-shot EPI; phase correction; QMRI; MODEL-BASED RECONSTRUCTION; RICIAN DISTRIBUTION; WEIGHTED-MRI; RESOLUTION; MOTION; CONVERGENCE; ARTIFACTS; SENSE; EPI;
D O I
10.1002/mrm.29398
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To introduce a novel imaging and parameter estimation framework for accurate multi-shot diffusion MRI. Theory and Methods We propose a new framework called ADEPT (Accurate Diffusion Echo-Planar imaging with multi-contrast shoTs) that enables fast diffusion MRI by allowing diffusion contrast settings to change between shots in a multi-shot EPI acquisition (i.e., intra-scan modulation). The framework estimates diffusion parameter maps directly from the acquired intra-scan modulated k-space data, while simultaneously accounting for shot-to-shot phase inconsistencies. The performance of the estimation framework is evaluated using Monte Carlo simulation studies and in-vivo experiments and compared to that of reference methods that rely on parallel imaging for shot-to-shot phase correction. Results Simulation and real-data experiments show that ADEPT yields more accurate and more precise estimates of the diffusion metrics in multi-shot EPI data in comparison with the reference methods. Conclusion ADEPT allows fast multi-shot EPI diffusion MRI without significantly degrading the accuracy and precision of the estimated diffusion maps.
引用
收藏
页码:396 / 410
页数:15
相关论文
共 56 条
[41]   Automatic localization of EEG electrode markers within 3D MR data [J].
Sijbers, J ;
Vanrumste, B ;
Van Hoey, G ;
Boon, P ;
Verhoye, M ;
Van der Linden, A ;
Van Dyck, D .
MAGNETIC RESONANCE IMAGING, 2000, 18 (04) :485-488
[42]   Clinical multishot DW-EPI through parallel imaging with considerations of susceptibility, motion, and noise [J].
Skare, Stefan ;
Newbould, Rexford D. ;
Clayton, David B. ;
Albers, Gregory W. ;
Nagle, Scott ;
Bammer, Roland .
MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (05) :881-890
[44]  
Tamir J, 2016, ISMRM WORKSH DAT SAM, P33
[45]  
THOMAS FC, 1994, MATH PROGRAM, V67, P189
[46]   MRtrix3: A fast, flexible and open software framework for medical image processing and visualisation [J].
Tournier, J-Donald ;
Smith, Robert ;
Raffelt, David ;
Tabbara, Rami ;
Dhollander, Thijs ;
Pietsch, Maximilian ;
Christiaens, Daan ;
Jeurissen, Ben ;
Yeh, Chun-Hung ;
Connelly, Alan .
NEUROIMAGE, 2019, 202
[47]   ESPIRiT-An Eigenvalue Approach to Autocalibrating Parallel MRI: Where SENSE Meets GRAPPA [J].
Uecker, Martin ;
Lai, Peng ;
Murphy, Mark J. ;
Virtue, Patrick ;
Elad, Michael ;
Pauly, John M. ;
Vasanawala, Shreyas S. ;
Lustig, Michael .
MAGNETIC RESONANCE IN MEDICINE, 2014, 71 (03) :990-1001
[48]  
van den Bos A, 2007, PARAMETER ESTIMATION FOR SCIENTISTS AND ENGINEERS, P1
[49]   Improved Pediatric MR Imaging with Compressed Sensing [J].
Vasanawala, Shreyas S. ;
Alley, Marcus T. ;
Hargreaves, Brian A. ;
Barth, Richard A. ;
Pauly, John M. ;
Lustig, Michael .
RADIOLOGY, 2010, 256 (02) :607-616
[50]   Model-based reconstruction of undersampled diffusion tensor k-space data [J].
Welsh, Christopher L. ;
DiBella, Edward V. R. ;
Adluru, Ganesh ;
Hsu, Edward W. .
MAGNETIC RESONANCE IN MEDICINE, 2013, 70 (02) :429-440