Evaluating the performance of the Roche FEN2 fentanyl immunoassay and its clinical implementation: The role of LDT-based mass spectrometry testing

被引:6
作者
Menlyadiev, Marlen [1 ]
Suhandynata, Raymond T. [1 ,2 ]
Lund, Kyle [1 ]
Kelner, Michael J. [1 ]
Fitzgerald, Robert L. [1 ]
机构
[1] Univ Calif San Diego, Ctr Adv Lab Med, Dept Pathol, San Diego Hlth Syst, San Diego, CA 92093 USA
[2] Univ Calif San Diego, Skaggs Sch Pharm & Pharmaceut Sci, San Diego, CA USA
来源
JOURNAL OF MASS SPECTROMETRY AND ADVANCES IN THE CLINICAL LAB | 2023年 / 28卷
关键词
Mass spectrometry; Immunoassay; Clinical sensitivity; Fentanyl; Norfentanyl; FEN2;
D O I
10.1016/j.jmsacl.2023.02.009
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Introduction: While laboratory-developed tests (LDTs) using liquid chromatography tandem mass spectrometry (LC-MS/MS) are widely employed to support the development of FDA-cleared drug immunoassays, their sig-nificance in the clinical implementation and evaluation of such assays is often overlooked. This paper reports on the important role of LC-MS/MS LDTs in demonstrating improved performance of the Roche FEN2 fentanyl immunoassay compared with the Thermo DRI fentanyl immunoassay.Methods: The FEN2 assay was implemented according to the manufacturer's instructions and its performance was compared to the existing DRI assay using LC-MS/MS as a reference. Clinical sensitivity and specificity were determined using 250 consecutive random patient specimens. Spiking experiments were conducted to determine cross-reactivity with 31 fentanyl analogs. Select DRI false-positive samples were analyzed by the FEN2 assay via time-of-flight mass spectrometry method (LC-QTOF).Results: The FEN2 assay showed improved clinical sensitivity compared to the DRI (98% vs 61%) in 250 consecutive patient samples due to its ability to detect norfentanyl. It also showed better clinical specificity by correctly classifying select DRI false-positive results. Upon implementation in clinical practice, the FEN2 resulted in a higher screening positivity rate than the DRI (17.3% vs 13.3%) and a greater LC-MS/MS confirmation rate of immunoassay-positive samples (96.8% vs 88.8%, respectively).Conclusion: The use of LC-MS/MS LDTs demonstrated that the FEN2 assay has greater clinical sensitivity and is less prone to false-positives than the DRI assay. These findings support the use of FEN2 in routine clinical practice and emphasize the role of mass spectrometry-based LDTs in clinical toxicology testing.
引用
收藏
页码:105 / 113
页数:9
相关论文
共 24 条
[1]  
[Anonymous], 2021, FEN2 FENT 2 ENZ IMM
[2]   Fentanyl, fentanyl analogs and novel synthetic opioids: A comprehensive review [J].
Armenian, Patil ;
Vo, Kathy T. ;
Barr-Walker, Jill ;
Lynch, Kara L. .
NEUROPHARMACOLOGY, 2018, 134 :121-132
[3]  
Baselt RC, 2014, DISPOSITION TOXIC DR, V10th, P846
[4]   Management of Pain in the United States-A Brief History and Implications for the Opioid Epidemic [J].
Bernard, Stephen A. ;
Chelminski, Paul R. ;
Ives, Timothy J. ;
Ranapurwala, Shabbar I. .
HEALTH SERVICES INSIGHTS, 2018, 11
[5]   Comparison of Two Commercially Available Fentanyl Screening Immunoassays for Clinical Use [J].
Budelier, Melissa M. ;
Franks, Caroline E. ;
Logsdon, Nicole ;
Jannetto, Paul J. ;
Scott, Mitchell G. ;
Roper, Stephen M. ;
Farnsworth, Christopher W. .
JOURNAL OF APPLIED LABORATORY MEDICINE, 2020, 5 (06) :1277-1286
[6]   Liquid Chromatography High-Resolution TOF Analysis: Investigation of MSE for Broad-Spectrum Drug Screening [J].
Chindarkar, Nandkishor S. ;
Wakefield, Michael R. ;
Stone, Judith A. ;
Fitzgerald, Robert L. .
CLINICAL CHEMISTRY, 2014, 60 (08) :1115-1125
[7]  
Clinical and Laboratory Standards Institute (CLSI), 2014, C62A CLSI
[8]   The North American opioid epidemic: opportunities and challenges for clinical laboratories [J].
Delaney, Sarah R. ;
Tacker, Danyel H. ;
Snozek, Christine L. H. .
CRITICAL REVIEWS IN CLINICAL LABORATORY SCIENCES, 2022, 59 (05) :309-331
[9]   Performance of Two Fentanyl Immunoassays against a Liquid Chromatography-Tandem Mass Spectrometry Method [J].
Feng, Sheng ;
Rutledge, Theodore J. ;
Manzoni, Maureen ;
Le, Thuan ;
Gardiner, JoAnn ;
Milone, Michael ;
Shaw, Leslie ;
Wang, Ping .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2021, 45 (02) :117-123
[10]   Fentanyl Buccal Soluble Film: A Review in Breakthrough Cancer Pain [J].
Garnock-Jones, Karly P. .
CLINICAL DRUG INVESTIGATION, 2016, 36 (05) :413-419