Body composition profiling at 0.55T: Feasibility and precision

被引:3
|
作者
Nayak, Krishna S. [1 ,4 ]
Cui, Sophia X. [2 ]
Tasdelen, Bilal [1 ]
Yagiz, Ecrin [1 ]
Weston, Sarah [3 ]
Zhong, Xiaodong [2 ]
Ahlgren, Andre [3 ]
机构
[1] Univ Southern Calif, Elect & Comp Engn, Los Angeles, CA USA
[2] Siemens Med Solut USA, Los Angeles, CA USA
[3] AMRA Med, Linkoping, Sweden
[4] 3740 McClintock Ave,Suite 400C, Los Angeles, CA 90089 USA
基金
美国国家科学基金会;
关键词
0; 55; tesla; body composition MRI; low field; metabolic disease; obesity; screening; FAT; QUANTIFICATION; MR;
D O I
10.1002/mrm.29682
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeBody composition MRI captures the distribution of fat and lean tissues throughout the body, and provides valuable biomarkers of obesity, metabolic disease, and muscle disorders, as well as risk assessment. Highly reproducible protocols have been developed for 1.5T and 3T MRI. The purpose of this work was to demonstrate the feasibility and test-retest repeatability of MRI body composition profiling on a 0.55T whole-body system. MethodsHealthy adult volunteers were scanned on a whole-body 0.55T MRI system using the integrated body RF coil. Experiments were performed to refine parameter settings such as TEs, resolution, flip angle, bandwidth, acceleration, and oversampling factors. The final protocol was evaluated using a test-retest study with subject removal and replacement in 10 adult volunteers (5 M/5F, age 25-60, body mass index 20-30). ResultsCompared to 1.5T and 3T, the optimal flip angle at 0.55T was higher (15 degrees), due to the shorter T1 times, and the optimal echo spacing was larger, due to smaller chemical shift between water and fat. Overall image quality was comparable to conventional field strengths, with no significant issues with fat/water swapping or inadequate SNR. Repeatability coefficient of visceral fat, subcutaneous fat, total thigh muscle volume, muscle fat infiltration, and liver fat were 11.8 cL (2.2%), 46.9 cL (1.9%), 14.6 cL (0.5%), 0.1 pp (2%), and 0.2 pp (5%), respectively (coefficient of variation in parenthesis). ConclusionsWe demonstrate that 0.55T body composition MRI is feasible and present optimized scan parameters. The resulting images provide satisfactory quality for automated post-processing and produce repeatable results.
引用
收藏
页码:1114 / 1120
页数:7
相关论文
共 50 条
  • [41] 3D MR elastography at 0.55 T: Concomitant field effects and feasibility
    Darwish, Omar Isam
    Di Cio, Pierluigi
    Sinkus, Ralph
    Neji, Radhouene
    MAGNETIC RESONANCE IN MEDICINE, 2024,
  • [42] Low-Field-Strength Body MRI: Challenges and Opportunities at 0.55 T
    Shetty, Anup S.
    Ludwig, Daniel R.
    Ippolito, Joseph E.
    Andrews, Trevor J.
    Narra, Vamsi R.
    Fraum, Tyler J.
    RADIOGRAPHICS, 2023, 43 (12)
  • [43] Phase-Resolved Functional Lung 0.55T MRI Analysis Demonstrates Ventilation-perfusion Defects Driven by Abnormal Perfusion in Patients With Fontan Circulation
    Kato, R. M.
    Chun, S.
    Detterich, J.
    Cui, S. X.
    Iyer, N. P.
    Lee, N.
    Nayak, K.
    Doyle, E.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2024, 209
  • [44] Efficient spiral in-out and EPI balanced steady-state free precession cine imaging using a high-performance 0.55T MRI
    Restivo, Matthew C.
    Ramasawmy, Rajiv
    Bandettini, W. Patricia
    Herzka, Daniel A.
    Campbell-Washburn, Adrienne E.
    MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (05) : 2364 - 2375
  • [45] Body composition measurements with Lunar iDXA: Precision evaluation
    Toussirot, E.
    Semon, C.
    Penfornis, F.
    Wendling, D.
    JOURNAL OF BONE AND MINERAL RESEARCH, 2007, 22 : S300 - S300
  • [46] Precision evaluation of lunar idxa body composition measurements
    Toussirot, E.
    Penfornis, F.
    Wendling, D.
    CALCIFIED TISSUE INTERNATIONAL, 2007, 80 : S151 - S151
  • [47] Body Composition Profiling in the UK Biobank Imaging Study
    Linge, Jennifer
    Borga, Magnus
    West, Janne
    Tuthill, Theresa
    Miller, Melissa R.
    Dumitriu, Alexandra
    Thomas, E. Louise
    Romu, Thobias
    Tunon, Patrik
    Bell, Jimmy D.
    Leinhard, Olof Dahlqvist
    OBESITY, 2018, 26 (11) : 1785 - 1795
  • [48] Reliability and Feasibility of Low-Field-Strength Fetal MRI at 0.55 T during Pregnancy
    Verdera, Jordina Aviles
    Story, Lisa
    Hall, Megan
    Finck, Tom
    Egloff, Alexia
    Seed, Paul T.
    Malik, Shaihan J.
    Rutherford, Mary A.
    Hajnal, Joseph, V
    Tomi-Tricot, Raphael
    Hutter, Jana
    RADIOLOGY, 2023, 309 (01)
  • [49] Diagnostic Image Quality of a Low-Field (0.55T) Knee MRI Protocol Using Deep Learning Image Reconstruction Compared with a Standard (1.5T) Knee MRI Protocol
    Lopez Schmidt, Ingo
    Haag, Nina
    Shahzadi, Iram
    Frohwein, Lynn Johann
    Schneider, Claus
    Niehoff, Julius Henning
    Kroeger, Jan Robert
    Borggrefe, Jan
    Moenninghoff, Christoph
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (05)
  • [50] Technical feasibility of CRISPR-based functional profiling for solid tumor precision medicine
    Hodel, Karl
    Varkaris, Andreas
    Chevalier, Nicholas
    Meyers, Maci
    Kehlmann, Allison
    Corbel, Stephane
    Stavridi, Anastasia
    Kilfeather, Peter
    Hagmann, Leonardo
    Tran, Kathleen
    Li, Yali
    Schmedt, Christian
    Sanidas, Ioannis
    Sampath, Srinath
    Sampath, Srihari
    Juric, Dejan
    CANCER RESEARCH, 2024, 84 (06)