Body composition profiling at 0.55T: Feasibility and precision

被引:3
|
作者
Nayak, Krishna S. [1 ,4 ]
Cui, Sophia X. [2 ]
Tasdelen, Bilal [1 ]
Yagiz, Ecrin [1 ]
Weston, Sarah [3 ]
Zhong, Xiaodong [2 ]
Ahlgren, Andre [3 ]
机构
[1] Univ Southern Calif, Elect & Comp Engn, Los Angeles, CA USA
[2] Siemens Med Solut USA, Los Angeles, CA USA
[3] AMRA Med, Linkoping, Sweden
[4] 3740 McClintock Ave,Suite 400C, Los Angeles, CA 90089 USA
基金
美国国家科学基金会;
关键词
0; 55; tesla; body composition MRI; low field; metabolic disease; obesity; screening; FAT; QUANTIFICATION; MR;
D O I
10.1002/mrm.29682
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeBody composition MRI captures the distribution of fat and lean tissues throughout the body, and provides valuable biomarkers of obesity, metabolic disease, and muscle disorders, as well as risk assessment. Highly reproducible protocols have been developed for 1.5T and 3T MRI. The purpose of this work was to demonstrate the feasibility and test-retest repeatability of MRI body composition profiling on a 0.55T whole-body system. MethodsHealthy adult volunteers were scanned on a whole-body 0.55T MRI system using the integrated body RF coil. Experiments were performed to refine parameter settings such as TEs, resolution, flip angle, bandwidth, acceleration, and oversampling factors. The final protocol was evaluated using a test-retest study with subject removal and replacement in 10 adult volunteers (5 M/5F, age 25-60, body mass index 20-30). ResultsCompared to 1.5T and 3T, the optimal flip angle at 0.55T was higher (15 degrees), due to the shorter T1 times, and the optimal echo spacing was larger, due to smaller chemical shift between water and fat. Overall image quality was comparable to conventional field strengths, with no significant issues with fat/water swapping or inadequate SNR. Repeatability coefficient of visceral fat, subcutaneous fat, total thigh muscle volume, muscle fat infiltration, and liver fat were 11.8 cL (2.2%), 46.9 cL (1.9%), 14.6 cL (0.5%), 0.1 pp (2%), and 0.2 pp (5%), respectively (coefficient of variation in parenthesis). ConclusionsWe demonstrate that 0.55T body composition MRI is feasible and present optimized scan parameters. The resulting images provide satisfactory quality for automated post-processing and produce repeatable results.
引用
收藏
页码:1114 / 1120
页数:7
相关论文
共 50 条
  • [31] A 20-gauge active needle design with thin-film printed circuitry for interventional MRI at 0.55T
    Yildirim, Dursun Korel
    Bruce, Christopher
    Uzun, Dogangun
    Rogers, Toby
    O'Brien, Kendall
    Ramasawmy, Rajiv
    Campbell-Washburn, Adrienne
    Herzka, Daniel A.
    Lederman, Robert J.
    Kocaturk, Ozgur
    MAGNETIC RESONANCE IN MEDICINE, 2021, 86 (03) : 1786 - 1801
  • [32] Oxygen-Enhanced Lung Imaging in Patients with LAM Using a High-Performance 0.55t MRI System
    Campbell-Washburn, A. E.
    Bhattacharya, I.
    Chen, M. Y.
    Majeed, W.
    Benkert, T.
    Balaban, R. S.
    Moss, J.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2020, 201
  • [33] T1 and T2 measurements across multiple 0.55T MRI systems using open-source vendor-neutral sequences
    Keenan, Kathryn E.
    Tasdelen, Bilal
    Javed, Ahsan
    Ramasawmy, Rajiv
    Rizzo, Rudy
    Martin, Michele N.
    Stupic, Karl F.
    Seiberlich, Nicole
    Campbell-Washburn, Adrienne E.
    Nayak, Krishna S.
    MAGNETIC RESONANCE IN MEDICINE, 2025, 93 (01) : 289 - 300
  • [34] Advanced body composition assessment: from body mass index to body composition profiling
    Borga, Magnus
    West, Janne
    Bell, Jimmy D.
    Harvey, Nicholas C.
    Romu, Thobias
    Heymsfield, Steven B.
    Leinhard, Olof Dahlqvist
    JOURNAL OF INVESTIGATIVE MEDICINE, 2018, 66 (05) : 887 - 895
  • [35] Passive artifact behavior prediction of interventional tools in high-field MRI using a 0.55T portable benchtop MR scanner
    Abadi, H.
    Krug, J.
    Illanes, A.
    Friebe, M.
    2016 38TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2016, : 1252 - 1255
  • [36] Concomitant magnetic-field compensation for 2D spiral-ring turbo spin-echo imaging at 0.55T and 1.5T
    Wang, Zhixing
    Ramasawmy, Rajiv
    Feng, Xue
    Campbell-Washburn, Adrienne E.
    Mugler III, John P.
    Meyer, Craig H.
    MAGNETIC RESONANCE IN MEDICINE, 2023, 90 (02) : 552 - 568
  • [37] BODY COMPOSITION PROFILING IN A DUTCH SARCOIDOSIS POPULATION
    Cremers, J. P.
    Drent, M.
    Elfferich, M. D.
    Nelemans, P. J.
    Wijnen, P. A.
    Witteman, B. J.
    Schols, A. M.
    SARCOIDOSIS VASCULITIS AND DIFFUSE LUNG DISEASES, 2013, 30 (04) : 289 - 299
  • [38] Single breath-hold volumetric lung imaging at 0.55T using stack-of-spiral (SoS) out-in balanced SSFP
    Tian, Ye
    Lee, Nam G.
    Zhao, Ziwei
    Wilcox, Alison G.
    Nieva, Jorge J.
    Nayak, Krishna S.
    MAGNETIC RESONANCE IN MEDICINE, 2025, 93 (05) : 1999 - 2007
  • [39] Image Quality of Lumbar Spine Imaging at 0.55T Low-Field MRI is Comparable to Conventional 1.5T MRI - Initial Observations in Healthy Volunteers
    Breit, Hanns-Christian
    Vosshenrich, Jan
    Hofmann, Verena
    Rusche, Thilo
    Kovacs, Balazs K.
    Bach, Michael
    Manneck, Sebastian
    Harder, Dorothee
    ACADEMIC RADIOLOGY, 2023, 30 (11) : 2440 - 2446
  • [40] Feasibility of MR fingerprinting using a high-performance 0.55 T MRI system
    Campbell-Washburn, Adrienne E.
    Jiang, Yun
    Koerzdoerfer, Gregor
    Nittka, Mathias
    Griswold, Mark A.
    MAGNETIC RESONANCE IMAGING, 2021, 81 : 88 - 93