A large field-of-view angled panel detector for the development of a plant PET system

被引:1
|
作者
Yamamoto, Seiichi [1 ]
Watabe, Hiroshi [2 ]
Yamaguchi, Mitsutaka [3 ]
Suzui, Nobuo [3 ]
Kawachi, Naoki [3 ]
Kataoka, Jun [1 ]
机构
[1] Waseda Univ, Fac Sci & Engn, Tokyo, Japan
[2] Tohoku Univ, Cyclotron Radioisotope Ctr, Sendai, Japan
[3] Natl Inst Quantum Sci & Technol QST, Fdn Quantum Technol Res Directorate, Takasaki, Japan
关键词
Detector design and construction technologies and materials; Gamma camera; SPECT; PET PET/CT; coronary CT angiography (CTA); Inspection with gamma rays; Instruments for environmental monitoring; food control and medical use; POSITRON-EMISSION-TOMOGRAPHY; LONG-DISTANCE TRANSPORT; CADMIUM; ACCUMULATION;
D O I
10.1088/1748-0221/19/01/T01002
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In plant research, positron emission tomography (PET) is sometimes employed for physiological studies of plants. However, due to the generally high cost of PET systems, their application in plant research centers can be challenging. Furthermore, the configurations of commercial PET systems might not always align well with the requirements of plant research. To address these issues, we have taken the initial steps towards a solution by developing a large field-of-view (LFOV) angled panel detector intended for the development of a plant PET system. This panel detector is composed of two gadolinium orthosilicate (GSO) scintillator blocks, an angled light guide, and four flat panel photomultiplier tubes (FP-PMTs). The GSO pixel size used in the scintillator block measures 2.65 mm (radial)x2.85 mm (axial)x15.0 mm, arranged in a configuration of 16 (radial)x32 (axial) to form a scintillator block with overall dimensions of 44.0 mm (radial)x94.4 mm (axial)x15 mm (depth). By implementing an angled light guide, the two scintillator blocks were positioned at a 22.5-degree angle, resulting in the creation of a LFOV angled panel detector. This specific angle allowed for the arrangement of 16 scintillator blocks in a hexadecagonal shape using 8 panel detectors when arranged in a circular configuration. We successfully constructed a panel detector with an approximate field-of -view of 10 cm x 10 cm and conducted a comprehensive performance evaluation. Upon irradiating the panel detector with 511 keV gamma photons, nearly all 1024 GSO pixels were accurately resolved in the position histogram, displaying minimal distortion, and exhibiting an average peak-to-valley ratio of 5.0. Additionally, the energy resolution achieved was 14% full width at half maximum (FWHM). The resulting LFOV angled panel detector holds considerable promise for the use within a cost-effective, user-friendly as well as high performance plant PET system.
引用
收藏
页数:16
相关论文
共 26 条
  • [1] Performance evaluation of a large axial field-of-view PET scanner: SET- 2400W
    Fujiwara T.
    Watanuki S.
    Yamamoto S.
    Miyake M.
    Seo S.
    Itoh M.
    Ishii K.
    Orihara H.
    Fukuda H.
    Satoh T.
    Kitamura K.
    Tanaka K.
    Takahashi S.
    Annals of Nuclear Medicine, 1997, 11 (4) : 307 - 313
  • [2] Development and Evaluation of mini-EXPLORER: A Long Axial Field-of-View PET Scanner for Nonhuman Primate Imaging
    Berg, Eric
    Zhang, Xuezhu
    Bec, Julien
    Judenhofer, Martin S.
    Patel, Brijesh
    Peng, Qiyu
    Kapusta, Maciej
    Schmand, Matthias
    Casey, Michael E.
    Tarantal, Alice F.
    Qi, Jinyi
    Badawi, Ramsey D.
    Cherry, Simon R.
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59 (06) : 993 - 998
  • [3] Development of novel low-cost readout electronics for large field-of-view gamma camera detectors
    Radnia, Aram
    Alikhani, Amirhossein
    Teimourian, Behnoosh
    Nejad, Mahyar Yousef
    Farahani, Mohammad Hossein
    Pashaei, Fakhereh
    Rahmim, Arman
    Zaidi, Habib
    Ay, Mohammad Reza
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 121
  • [4] Progress in large field-of-view interventional planar scintigraphy and SPECT imaging
    Dietze, Martijn M. A.
    de Jong, Hugo W. A. M.
    EXPERT REVIEW OF MEDICAL DEVICES, 2022, 19 (05) : 393 - 403
  • [5] Long-axial field-of-view PET/CT improves radiomics feature reliability
    Alberts, Ian L.
    Xue, Song
    Sari, Hasan
    Cavinato, Lara
    Prenosil, George
    Afshar-Oromieh, Ali
    Mingels, Clemens
    Shi, Kuangyu
    Caobelli, Federico
    Rahmim, Arman
    Pyka, Thomas
    Rominger, Axel
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2025, 52 (03) : 1004 - 1016
  • [6] Theoretical study of the benefit of long axial field-of-view PET on region of interest quantification
    Zhang, Xuezhu
    Badawi, Ramsey D.
    Cherry, Simon R.
    Qi, Jinyi
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (13)
  • [7] Evaluation of a 25-511keV List Mode Readout System for a Large Field-of-View Gamma Camera
    Villena, J. L.
    Tapias, G.
    Lage, E.
    Kreuger, R.
    Beekman, F. J.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2168 - 2173
  • [8] Parallax error in long-axial field-of-view PET scanners-a simulation study
    Schmall, Jeffrey P.
    Karp, Joel S.
    Werner, Matt
    Surti, Suleman
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (14) : 5443 - 5455
  • [9] 18FDG-PET/CT in endocarditis. Is inclusion of lower limbs in acquisition field-of-view mandatory?
    Mikail, N.
    Benali, K.
    Ou, P.
    Slama, J.
    Hyafil, F.
    Le Guludec, D.
    Rouzet, F.
    MEDECINE NUCLEAIRE-IMAGERIE FONCTIONNELLE ET METABOLIQUE, 2015, 39 (05): : 450 - 457
  • [10] Long-axial field-of-view PET/CT: perspectives and review of a revolutionary development in nuclear medicine based on clinical experience in over 7000 patients
    Alberts, Ian
    Sari, Hasan
    Mingels, Clemens
    Afshar-Oromieh, Ali
    Pyka, Thomas
    Shi, Kuangyu
    Rominger, Axel
    CANCER IMAGING, 2023, 23 (01)