Dynamic Heart Simulator for Ultrasound-Guided Pericardiocentesis

被引:0
作者
Yan, Kim [1 ,2 ]
Yan, Wanquan [1 ,2 ]
Cheng, Shing Shin [3 ,4 ]
机构
[1] Chinese Univ Hong Kong, Dept Mech & Automat Engn, Hong Kong, Peoples R China
[2] Chinese Univ Hong Kong, CUHK T Stone Robot Inst, Hong Kong, Peoples R China
[3] Chinese Univ Hong Kong, CUHK T Stone Robot Inst, Shun Hing Inst Adv Engn, Dept Mech & Automat Engn, Hong Kong, Peoples R China
[4] Chinese Univ Hong Kong, Multi Scale Med Robot Ctr, Hong Kong, Peoples R China
来源
2023 IEEE/RSJ INTERNATIONAL CONFERENCE ON INTELLIGENT ROBOTS AND SYSTEMS, IROS | 2023年
关键词
MOTION ESTIMATION; MODEL;
D O I
10.1109/IROS55552.2023.10342122
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Pericardiocentesis is an important surgical intervention to treat a medical condition called pericardial effusion, during which excessive fluid accumulates around the heart, potentially leading to life-threatening situation. It involves the insertion of a needle and catheter towards the heart into the pericardial space to drain the excessive fluid under ultrasound (US) guidance. The risky procedure requires surgeons to acquire sufficient training to ensure safe execution of the procedure. However, existing heart simulators lack dynamic features, do not offer realistic images under US imaging, and are not reusable. This work presents a dynamic heart simulator (DHS) with pericardial effusion to mimic the beating motion of the human heart and the realistic US imaging results. The beating heart motion is realized using a hydraulic actuation system connected to a double-layer balloon set. The clear and realistic US imaging results are obtained through a unique formula proposed for the chest tissue and the cardiac muscle. A characterization method was also developed to allow customization of important anatomical parameters in the DHS. The experimental results show that the DHS allowed highly realistic simulation of the beating heart, cardiac muscle, and pericardium under US imaging and has been demonstrated to enable successful US-guided pericardiocentesis.
引用
收藏
页码:4640 / 4647
页数:8
相关论文
共 32 条
[1]   Estimating heart mass from heart volume as measured from post-mortem computed tomography [J].
Aitken-Buck, Hamish M. ;
Moore, Matthew ;
Whalley, Gillian A. ;
Lohner, Larissa ;
Ondruschka, Benjamin ;
Coffey, Sean ;
Tse, Rexson D. ;
Lamberts, Regis R. .
FORENSIC SCIENCE MEDICINE AND PATHOLOGY, 2022, 18 (03) :333-342
[2]  
[Anonymous], 2003, 2003 EUR CONTR C ECC
[3]   Real-world heart rate norms in the Health eHeart study [J].
Avram, Robert ;
Tison, Geoffrey H. ;
Aschbacher, Kirstin ;
Kuhar, Peter ;
Vittinghoff, Eric ;
Butzner, Michael ;
Runge, Ryan ;
Wu, Nancy ;
Pletcher, Mark J. ;
Marcus, Gregory M. ;
Olgin, Jeffrey .
NPJ DIGITAL MEDICINE, 2019, 2 (1)
[4]   Model-based motion estimation of elastic surfaces for minimally invasive cardiac surgery [J].
Bader, Thomas ;
Wiedemann, Alexander ;
Roberts, Kathrin ;
Hanebeck, Uwe D. .
PROCEEDINGS OF THE 2007 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS 1-10, 2007, :2261-+
[5]   Inexpensive, High-Fidelity Model to Simulate Ultrasound-Guided Pericardiocentesis for Cardiology Resident Training [J].
Belliveau, Daniel J. ;
Moeller, Andrew ;
Ramer, Sarah .
CANADIAN JOURNAL OF CARDIOLOGY, 2019, 35 (11) :1600-1603
[6]  
Daly Robert, 2017, West J Emerg Med, V18, P114, DOI 10.5811/westjem.2016.10.31506
[7]   Polyvinyl chloride plastisol breast phantoms for ultrasound imaging [J].
de Carvalho, Isabela Miller ;
De Matheo, Lucas Lobianco ;
Silva Costa Junior, Jose Francisco ;
Borba, Cecilia de Melo ;
von Kruger, Marco Antonio ;
Catelli Infantosi, Antonio Fernando ;
de Albuquerque Pereira, Wagner Coelho .
ULTRASONICS, 2016, 70 :98-106
[8]   Needle-tissue interaction modeling using ultrasound-based motion estimation: Phantom study [J].
Dehghan, Ehsan ;
Wen, Xu ;
Zahiri-Azar, Reza ;
Marchal, Maud ;
Salcudean, Septimiu E. .
COMPUTER AIDED SURGERY, 2008, 13 (05) :265-280
[9]  
Doyle A., 2019, SCI REPORTS, V11, P11160
[10]  
Gaudette Connor, 2022, 2022 IEEE MIT Undergraduate Research Technology Conference (URTC), P1, DOI 10.1109/URTC56832.2022.10002212