Study of Temperature Changes Around Fibres in Super Pulse Thulium Fibre Laser in Vitro Lithotripsy

被引:4
|
作者
Ding, Tianfu [1 ]
Bo, Xiao [1 ]
Xue, Zeng [1 ]
Lei, Liang [1 ]
Ji, Chaoyue [1 ]
Li, Jianxing [1 ]
机构
[1] Tsinghua Univ, Tsinghua Changgung Hosp, Sch Clin Med, Dept Urol, Beijing 102218, Peoples R China
关键词
Super-pulse thulium fibre laser; temperature; lithotripsy; safety threshold; STONES;
D O I
10.22037/uj.v20i.7326
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose: To investigate temperature changes around the fibres of the super pulse thulium fibre laser (SP-TFL) during in vitro lithotripsy. Materials and Methods: Stones were placed in the in vitro model. The laser was continuously excited for 180 s; the probe was positioned 5 mm around the fibre; the laser power was set at 10, 15, 20, 25, and 30 W; and the irri-gation rate was set at 0, 15, 25, 35 ml/min, using a domestic SP-TFL. The temperature variations around the fibre under different power settings and different irrigation rates were compared. Results: Without irrigation, the temperature around the fibre rapidly reached the safety threshold of 43? at 24 s. At an irrigation rate 15 ml/min and laser power <15 W, the temperature around the fibre was < 43?. Once the laser power increased to > 20 W, the temperature around the fibre increased to >43? at its lowest plateau. At irrigation rate 25 ml/min and laser power < 25 W, the temperature around the fibre was < 43?. At irrigation rate 35 ml/min and laser power < 30 W, the fibre temperature was < 43?. When laser power was > 30 W, the fibre temperature was > 43?.Conclusion: In extracorporeal ureteroscope SP-TFL lithotripsy, when the laser power is < 15 W, < 25 W, and < 30 W, the irrigation rate should be maintained at > 15 ml/min, > 25 ml/min, and > 35 ml/min, respectively.
引用
收藏
页码:96 / 101
页数:6
相关论文
共 50 条
  • [31] Holmium-YAG laser: impact of pulse energy and frequency on local fluid temperature in an in-vitro obstructed kidney calyx model
    Sourial, Michael W.
    Ebel, Joshua
    Francois, Nathaly
    Box, Geoffrey N.
    Knudsen, Bodo E.
    JOURNAL OF BIOMEDICAL OPTICS, 2018, 23 (10)
  • [32] Ureteroscopic Performance of High Power Super Pulse Thulium Fiber Laser for the Treatment of Urolithiasis: Results of the First Case Series in North America
    Carrera, Raphael, V
    Randall, J. Hogan
    Garcia-Gil, Maurilio
    Knudsen, Bodo E.
    Chew, Ben H.
    Thompson, Jeffrey A.
    Humphreys, Mitchell R.
    Molina, Wilson R.
    UROLOGY, 2021, 153 : 87 - 92
  • [33] Temperature changes of renal calyx during high-power flexible ureteroscopic Moses holmium laser lithotripsy: a case analysis study
    Zhiyue Wu
    Jingchao Wei
    Chengfang Sun
    Qi Huangfu
    Bohan Wang
    Guangyi Huang
    Kai Qi
    Zeping Jiang
    Wuan Zhu
    Jiaming Wen
    International Urology and Nephrology, 2023, 55 : 1685 - 1692
  • [34] Temperature changes of renal calyx during high-power flexible ureteroscopic Moses holmium laser lithotripsy: a case analysis study
    Wu, Zhiyue
    Wei, Jingchao
    Sun, Chengfang
    Qi, Huangfu
    Wang, Bohan
    Huang, Guangyi
    Qi, Kai
    Jiang, Zeping
    Zhu, Wuan
    Wen, Jiaming
    INTERNATIONAL UROLOGY AND NEPHROLOGY, 2023, 55 (07) : 1685 - 1692
  • [35] Outcomes of holmium: YAG laser vs. Thulium fiber laser for ureteric stones during ureterorenoscopic lithotripsy - a prospective, randomized single-centre study
    Kaushik, T. G. Subash
    Sekar, Hariharasudhan
    Chandru, T.
    Krishnamoorthy, Sriram
    WORLD JOURNAL OF UROLOGY, 2025, 43 (01)
  • [36] Influence of an 810-nm Diode Laser on the Temperature Changes of the External Root Surface: An In Vitro Study
    Shehab, Njwan F.
    Al-Sabawi, Nawala
    Alkhalidi, Emad F.
    JOURNAL OF INTERNATIONAL SOCIETY OF PREVENTIVE AND COMMUNITY DENTISTRY, 2020, 10 (04) : 445 - 451
  • [37] Laser operator duty cycle effect on temperature and thermal dose: in-vitro study
    Louters, Marne M.
    Dau, Julie J.
    Hall, Timothy L.
    Ghani, Khurshid R.
    Roberts, William W.
    WORLD JOURNAL OF UROLOGY, 2022, 40 (06) : 1575 - 1580
  • [38] In vitro study of intrapulpar temperature changes when using Er:YAG laser for Class V cavity preparation
    Brugnera, A
    Zanin, F
    Marchesan, MA
    da Silva, RS
    Pécora, JD
    LASERS IN DENTISTRY VII, 2001, 4249 : 180 - 182
  • [39] Temperature profiles during ureteroscopy with thulium fiber laser and holmium:YAG laser: Findings from a pre-clinical study
    Aesoy, M. S.
    Juliebo-Jones, P.
    Beisland, C.
    Ulvik, O.
    SCANDINAVIAN JOURNAL OF UROLOGY, 2022, 56 (04) : 313 - 319
  • [40] Effect of Chilled Irrigation on Caliceal Fluid Temperature and Time to Thermal Injury Threshold During Laser Lithotripsy: In Vitro Model
    Dau, Julie J.
    Hall, Timothy L.
    Maxwell, Adam D.
    Ghani, Khurshid R.
    Roberts, William W.
    JOURNAL OF ENDOUROLOGY, 2021, 35 (05) : 700 - 705