Study of Temperature Changes Around Fibres in Super Pulse Thulium Fibre Laser in Vitro Lithotripsy

被引:4
|
作者
Ding, Tianfu [1 ]
Bo, Xiao [1 ]
Xue, Zeng [1 ]
Lei, Liang [1 ]
Ji, Chaoyue [1 ]
Li, Jianxing [1 ]
机构
[1] Tsinghua Univ, Tsinghua Changgung Hosp, Sch Clin Med, Dept Urol, Beijing 102218, Peoples R China
关键词
Super-pulse thulium fibre laser; temperature; lithotripsy; safety threshold; STONES;
D O I
10.22037/uj.v20i.7326
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose: To investigate temperature changes around the fibres of the super pulse thulium fibre laser (SP-TFL) during in vitro lithotripsy. Materials and Methods: Stones were placed in the in vitro model. The laser was continuously excited for 180 s; the probe was positioned 5 mm around the fibre; the laser power was set at 10, 15, 20, 25, and 30 W; and the irri-gation rate was set at 0, 15, 25, 35 ml/min, using a domestic SP-TFL. The temperature variations around the fibre under different power settings and different irrigation rates were compared. Results: Without irrigation, the temperature around the fibre rapidly reached the safety threshold of 43? at 24 s. At an irrigation rate 15 ml/min and laser power <15 W, the temperature around the fibre was < 43?. Once the laser power increased to > 20 W, the temperature around the fibre increased to >43? at its lowest plateau. At irrigation rate 25 ml/min and laser power < 25 W, the temperature around the fibre was < 43?. At irrigation rate 35 ml/min and laser power < 30 W, the fibre temperature was < 43?. When laser power was > 30 W, the fibre temperature was > 43?.Conclusion: In extracorporeal ureteroscope SP-TFL lithotripsy, when the laser power is < 15 W, < 25 W, and < 30 W, the irrigation rate should be maintained at > 15 ml/min, > 25 ml/min, and > 35 ml/min, respectively.
引用
收藏
页码:96 / 101
页数:6
相关论文
共 50 条
  • [21] Analysis of stone trajectory during laser lithotripsy: an in vitro study
    Yang, Bingyuan
    Zhang, Jian J.
    Ray, Aditi
    Turney, Ben
    ADVANCED PHOTONICS IN UROLOGY 2023, 2023, 12353
  • [22] Stone ablation efficacy: a comparison of a thulium fibre laser and two pulse-modulated holmium:YAG lasers
    Bingyuan Yang
    Aditi Ray
    Jian James Zhang
    Steven Peng
    Mike O’Brien
    Ben Turney
    Urolithiasis, 51
  • [23] Stone ablation efficacy: a comparison of a thulium fibre laser and two pulse-modulated holmium:YAG lasers
    Yang, Bingyuan
    Ray, Aditi
    Zhang, Jian James
    Peng, Steven
    O'Brien, Mike
    Turney, Ben
    UROLITHIASIS, 2023, 51 (01)
  • [24] How much energy do we need to ablate 1 cubic millimeter of stone during Thulium Fiber Laser lithotripsy? An in vitro study
    Panthier, Frederic
    Chicaud, Marie
    Doizi, Steeve
    Kutchukian, Stessy
    Lapouge, Pierre
    Solano, Catalina
    Candela, Luigi
    Daudon, Michel
    Berthe, Laurent
    Corrales, Mariela
    Traxer, Olivier
    WORLD JOURNAL OF UROLOGY, 2024, 42 (01)
  • [25] Impact of Laser Fiber Diameter and Irrigation Fluids on Induced Bubble Stream Dynamics with Thulium Fiber Laser: An In Vitro Study
    Panthier, Frederic
    Doizi, Steeve
    Gorny, Cyril
    Berthe, Laurent
    Traxer, Olivier
    JOURNAL OF ENDOUROLOGY, 2021, 35 (12) : 1883 - 1890
  • [26] Effect of temporal pulse shape on urinary stone phantom retropulsion rate and ablation efficiency using holmium:YAG and super-pulse thulium fibre lasers
    Ventimiglia, Eugenio
    Doizi, Steeve
    Kovalenko, Anastasiya
    Andreeva, Viktoria
    Traxer, Olivier
    BJU INTERNATIONAL, 2020, 126 (01) : 159 - 167
  • [27] Ablation Efficiency of a Novel Thulium Fiber Laser: An In Vitro Study on Laser Setting and Fiber Usage
    Liu, Min
    Peng, Yonghan
    Wang, Zeyu
    Li, Ling
    Ming, Shaoxiong
    Fang, Ziyu
    Dong, Hao
    Li, Rui
    Gao, Xiaofeng
    JOURNAL OF ENDOUROLOGY, 2021, 35 (08) : 1211 - 1216
  • [28] Does the Novel Thulium Fiber Laser Have a Higher Risk of Urothelial Thermal Injury than the Conventional Holmium Laser in an In Vitro Study?
    Belle, Joshua D.
    Chen, Ricky
    Srikureja, Nathaniel
    Amasyali, Akin S.
    Keheila, Mohammed
    Baldwin, D. Duane
    JOURNAL OF ENDOUROLOGY, 2022, 36 (09) : 1249 - 1254
  • [29] Kidney Stone Ablation Times and Peak Saline Temperatures During Holmium: YAG and Thulium Fiber Laser Lithotripsy, In Vitro, in a Ureteral Model
    Hardy, Luke A.
    Wilson, Christopher R.
    Irby, Pierce B.
    Fried, Nathaniel M.
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS XI, 2015, 9303
  • [30] Flexible Ureteroscopic Lithotripsy with the Pulsed Thulium:Yttrium Aluminum Garnet Laser Thulio: Preliminary Results from a Prospective Study
    Proietti, Silvia
    Marchioni, Michele
    Oo, Mon Mon
    Scalia, Riccardo
    Gisone, Stefano
    Monroy, Rebeca Escobar
    Schips, Luigi
    Gaboardi, Franco
    Giusti, Guido
    EUROPEAN UROLOGY OPEN SCIENCE, 2024, 67 : 77 - 83