Multi-point calorimeter using distributed fiber Bragg gratings for small field dosimetry in radiotherapy

被引:2
|
作者
Lebel-Cormier, Marie-Anne [1 ,2 ,3 ]
Boilard, Tommy [3 ,4 ]
Bernier, Martin [3 ,4 ]
Beaulieu, Luc [1 ,2 ,3 ]
机构
[1] Univ Laval, CHU Quebec, Ctr Integre cancerol & Axe Oncol, CRCHU Quebec, Quebec City, PQ, Canada
[2] Univ Laval, Ctr Rech Canc, Quebec City, PQ, Canada
[3] Univ Laval, Dept phys genie Phys & opt, Quebec City, PQ, Canada
[4] Univ Laval, Ctr opt photon & lasers, Quebec City, PQ, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
FBG captor; FBG dosimeter; optical fiber dosimeter; RADIATION SENSITIVITY; OPTICAL-FIBERS;
D O I
10.1002/mp.16955
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundThe interest of using fiber Bragg gratings (FBGs) dosimeters in radiotherapy (RT) lies in their (i) microliter detection volume, (ii) customizable spatial resolution, (iii) multi-point dose measurement, (iv) real-time data acquisition and (v) insensitivity to Cherenkov light. These characteristics could prove very useful for characterizing dose distributions of small and nonstandard fields with high spatial resolution.PurposeWe developed a multi-point FBGs dosimeter customized for small field RT dosimetry with a spatial resolution of similar to 1 mm.MethodsThe 3 cm-long multi-point dosimeter is made by embedding a 80 mu m$\umu{\rm {m}}$ silica fiber containing an array of thirty (30) co-located similar to 1 mm-long fs-written FBGs inside a plastic cylinder with an UV curing optical adhesive. With its higher thermal expansion coefficient, the plastic cylinder increases the sensitivity of the dosimeter by stretching the fiber containing the FBGs when the temperature rises slightly due to radiation energy deposition. Irradiations (2000 MU at 600 MU/min) were performed with a Varian TrueBeam linear accelerator.ResultsThe dose profile of a 2 x 2 cm2 6 MV beam was measured with a mean relative difference of 1.8% (excluding the penumbra region). The measured output factors for a 6 MV beam are in general agreement with the expected values within the experimental uncertainty (except for the 2 x$\,\times $ 2 cm2 field). The detector response to different energy of photon and electron beams is within 5% of the mean response (0.068 +/- 0.002 pm/Gy). The calorimeter's post-irradiation thermal decay is in agreement with the theory.ConclusionsAn energy-independent small field calorimeter that allows dose profile and output factor measurements for RT using FBGs was developed, which, to our knowledge, has never been done before. This type of detector could prove really useful for small field dosimetry, but also potentially for MRI-LINAC since FBGs are insensitive to magnetic fields and for FLASH since FBGs have been used to measure doses up to 100 kGy.
引用
收藏
页码:3758 / 3765
页数:8
相关论文
共 2 条
  • [1] Point-by-point femtosecond laser micro-processing of independent core-specific fiber Bragg gratings in a multi-core fiber
    Donko, A.
    Beresna, M.
    Jung, Y.
    Hayes, J.
    Richardson, D. J.
    Brambilla, G.
    OPTICS EXPRESS, 2018, 26 (02): : 2039 - 2044
  • [2] POINT-BY-POINT FABRICATION OF MICRO-BRAGG GRATINGS IN PHOTOSENSITIVE FIBER USING SINGLE EXCIMER PULSE REFRACTIVE-INDEX MODIFICATION TECHNIQUES
    MALO, B
    HILL, KO
    BILODEAU, F
    JOHNSON, DC
    ALBERT, J