Design and Experimental Validation of a Noninvasive Glucose Monitoring System Using RF Antenna-Based Biosensor

被引:18
作者
Mahnashi, Yaqub [1 ,2 ]
Qureshi, Khurram Karim [1 ,2 ]
Al-Shehri, Abdullah A. [1 ,2 ]
Attia, Hussein [1 ,2 ]
机构
[1] King Fahd Univ Petr & Minerals KFUPM, Elect Engn Dept, Dhahran 31261, Saudi Arabia
[2] King Fahd Univ Petr & Minerals KFUPM, Ctr Commun Syst & Sensing, Dhahran 31261, Saudi Arabia
关键词
Glucose; Sensors; Antenna radiation patterns; Microstrip antennas; Dielectrics; Monitoring; Blood; Biomedical applications; biosensors; glucose detection using RF sensors; noninvasive glucose monitoring; DIELECTRIC-PROPERTIES; BIOLOGICAL TISSUES; SENSOR; BAND;
D O I
10.1109/JSEN.2022.3227382
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This article presents an end-to-end microwave-based system to detect the glucose level in aqueous solutions through a noninvasive scheme. A microwave signal is generated and transmitted through the sample under test (i.e., glucose-water solution). The received signal is then conditioned using a low-noise amplifier (LNA), a bandpass filter (BPF), and an RF detector. The change in the dc output voltage on the receiver side is used as a new way to detect the glucose level. The proposed glucose sensor is implemented using two RF microstrip patch antennas that resonate at 5.7 GHz and are fabricated using an FR-4 substrate. The design specifications of the sensing antennas are thoroughly studied and presented. The system is verified experimentally using glucose-water testing samples. The experimental results confirm the correlation between the glucose concentration and the dc output voltage for a concentration range of 0-5000 mg/dL. The effect of the transmitted power level on the system performance is also investigated. Finally, the proposed system is compared with the state-of-the-art systems reported in the literature.
引用
收藏
页码:2856 / 2864
页数:9
相关论文
共 41 条
[1]  
[Anonymous], 2017, CONTINUOUS GLUCOSE M
[2]   Multicomponent blood analysis by near-infrared Raman spectroscopy [J].
Berger, AJ ;
Koo, TW ;
Itzkan, I ;
Horowitz, G ;
Feld, MS .
APPLIED OPTICS, 1999, 38 (13) :2916-2926
[3]   Glucose Sensing for Diabetes Monitoring: Recent Developments [J].
Bruen, Danielle ;
Delaney, Colm ;
Florea, Larisa ;
Diamond, Dermot .
SENSORS, 2017, 17 (08)
[4]   Wideband and high-gain EBG resonator antenna based on dual layer PRS [J].
Chaabane, Abdelhalim ;
Djahli, Farid ;
Attia, Hussein ;
Abdelghani, Lamine Mohamed ;
Denidni, Tayeb Ahmed .
MICROWAVE AND OPTICAL TECHNOLOGY LETTERS, 2017, 59 (01) :98-101
[5]   Dispersion and absorption in dielectrics I. Alternating current characteristics [J].
Cole, KS ;
Cole, RH .
JOURNAL OF CHEMICAL PHYSICS, 1941, 9 (04) :341-351
[6]  
Ebrahimi Amir, 2015, 2015 IEEE 15th Mediterranean Microwave Symposium (MMS), DOI 10.1109/MMS.2015.7375441
[7]   The dielectric properties of biological tissues .1. Literature survey [J].
Gabriel, C ;
Gabriel, S ;
Corthout, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2231-2249
[8]   The dielectric properties of biological tissues .2. Measurements in the frequency range 10 Hz to 20 GHz [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2251-2269
[9]   The dielectric properties of biological tissues .3. Parametric models for the dielectric spectrum of tissues [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2271-2293
[10]   Towards Accurate and Wideband In Vivo Measurement of Skin Dielectric Properties [J].
Gao, Yuan ;
Ghasr, Mohammad Tayeb ;
Nacy, Michael ;
Zoughi, Reza .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2019, 68 (02) :512-524