Development and application of absorbed dose primary standard for 60Co and high-energy photon beams using water calorimetry

被引:1
|
作者
Wang, Zhipeng [1 ]
Wang, Kun [1 ]
Jin, Sunjun [1 ]
Wang, Feifei [2 ]
Zhang, Xi [1 ]
Liu, Fubin [1 ]
Yao, Jintao [1 ,3 ]
Huang, Ji [1 ]
Qi, Yaping [1 ]
Zhang, Jian [1 ]
机构
[1] Natl Inst Metrol NIM, Div Ionizing Radiat Metrol, Beijing, Peoples R China
[2] China Inst Atom Energy CIAE, Beijing, Peoples R China
[3] Anhui Univ Sci & Technol, Huainan, Peoples R China
关键词
primary standard; absorbed-dose-to-water; water calorimeter; ionization chamber; beam quality conversion factor; IONIZATION CHAMBERS; K(Q) FACTORS; EFFECTIVE POINT; MV PHOTON; DOSIMETRY; HEAT;
D O I
10.1088/1681-7575/acc01b
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Several new glass vessels, thermistor probes, and water phantoms have been designed and built at the National Institute of Metrology (China) to upgrade and develop the existing water calorimeter. The increased plane-parallel vessels have shorter thermal stability times and shallower positioning depths (similar to 1.3 cm) than the previous cylindrical vessels, which makes them suitable for electron beams. The sensitivity of the new probes is 10% greater than the previous ones. In this study, detailed experimental and theoretical investigations of various factors affecting the new water calorimeter are performed. The system uncertainty of the water calorimeter is reduced and the robustness of the determination of the absorbed-dose-to-water D (w) is improved using a variety of geometric detector vessels and two kinds of high-purity water systems saturated with high-purity gases. This new calorimeter is employed as a primary standard for determining the D (w), has achieved a combined standard uncertainty of 0.24% for a Co-60 beam, 0.27% for 6 MV and 10 MV photon beams and 0.30% for a 25 MV photon beam. The beam quality conversion factors k(Q) of ten cylindrical and three plane-parallel ionization chambers are measured using the new calorimeter to improve the reference dosimetry accuracy of high-energy clinical photon beams.
引用
收藏
页数:17
相关论文
共 50 条
  • [21] The contribution of secondary heavy particles to the absorbed dose from high-energy photon beams
    Spurny, F
    Johansson, L
    Satherberg, A
    Bednar, J
    Turek, K
    PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (12): : 2643 - 2656
  • [23] A METHOD OF MEASURING THE PRIMARY DOSE COMPONENT IN HIGH-ENERGY PHOTON BEAMS
    NIZIN, P
    KASE, K
    MEDICAL PHYSICS, 1988, 15 (05) : 683 - 685
  • [24] A standard for absorbed dose rate to water in a 60Co field using a graphite calorimeter at the National Metrology Institute of Japan
    Morishita, Y.
    Kato, M.
    Takata, N.
    Kurosawa, T.
    Tanaka, T.
    Saito, N.
    RADIATION PROTECTION DOSIMETRY, 2013, 154 (03) : 331 - 339
  • [25] Absorbed dose to water standard of high-energy x-rays at the KRISS
    Kim, In Jung
    Kim, Yun Ho
    Yi, Chul-Young
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2022, 81 (10) : 1004 - 1012
  • [26] Absorbed dose to water standard of high-energy x-rays at the KRISS
    In Jung Kim
    Yun Ho Kim
    Chul-Young Yi
    Journal of the Korean Physical Society, 2022, 81 : 1004 - 1012
  • [27] A water calorimeter for on-site absorbed dose to water calibrations in 60Co and MV-photon beams including MRI incorporated treatment equipment
    de Prez, Leon
    de Pooter, Jacco
    Jansen, Bartel
    Aalbers, Tony
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (13): : 5051 - 5076
  • [29] Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams
    Ravichandran, Ramamoorthy
    Binukumar, John Pichy
    Al Amri, Iqbal
    Davis, Cheriyathmanjiyil Antony
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 291 - 303
  • [30] A PROTOCOL FOR THE DETERMINATION OF ABSORBED DOSE FROM HIGH-ENERGY PHOTON AND ELECTRON-BEAMS - COMMENTS
    WHITE, GA
    GIBBS, GL
    MEDICAL PHYSICS, 1985, 12 (01) : 114 - 114