Effectiveness of Derivo® Flow Diverter Device in the Endovascular Treatment of Cerebrovascular Aneurysms

被引:0
|
作者
Coban, Mustafa Kemal [1 ]
Ulku, Goktug [1 ]
Sahin, Mustafa Caglar [2 ]
Unal, Cansu [3 ]
Okay, Hilmi Onder [4 ]
机构
[1] Erzurum Reg Educ & Res Hosp, Dept Neurosurg, Erzurum, Turkiye
[2] Gazi Univ, Sch Med, Dept Neurosurg, Ankara, Turkiye
[3] Acibadem Mehmet Ali Aydinlar Univ, Sch Med, Istanbul, Turkiye
[4] Kafkas Univ, Hlth Res & Applicat Ctr, Dept Neurosurg, Kars, Turkiye
关键词
Aneurysm; Cerebrovascular; Derivo & REG; flow diverter; Endovascular; FOLLOW-UP;
D O I
10.5137/1019-5149.JTN.40190-22.3
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
AIM: To report our experience with flow diverter devices in the treatment of intracranial aneurysms, and to share the efficacy and clinical results of the Derivo (R) flow diverter device in the endovascular treatment of cerebrovascular aneurysms.MATERIAL and METHODS: This retrospective study was conducted at a regional training and research hospital between October 2015 and March 2020 after approval by the Clinical Research Ethics Committee. The radiology and file records of 21 patients with cerebrovascular aneurysms treated endovascularly using a Derivo (R) flow diverter device were examined.RESULTS: As per our findings, flow diverter devices were used to treat 27 aneurysms from 21 cases. Among the treated aneurysms, 3, 2, and 22 were in the middle cerebral artery, anterior communicating artery, and internal cerebral artery, respectively. Eight of the patients with a mean age of 56.9 years presented with subarachnoid hemorrhage. Derivo (R) flow diverters were used alone in 19 cases, whereas concurrent diverter device and coiling were used in only 3 patients. Complete closure of the aneurysms was observed in 3 (14.2%) cases; meanwhile, a 50% reduction in aneurysm size was observed in 2 (9.5%) cases. Complete aneurysm closure occurred in 20 cases (95%) at the 6-month follow-up. Mortality and morbidity each occurred in 1 (4.7%) case.CONCLUSION: Flow diverter devices provide an efficient and safe method of treatment, especially for fusiform, large, giant, widenecked intracranial aneurysms. Endovascular coil embolization treatment is deemed inappropriate for small aneurysms.
引用
收藏
页码:591 / 595
页数:5
相关论文
共 50 条
  • [1] Endovascular treatment of intracranial aneurysms using the Derivo Embolization Device: a multicenter experience
    Lourenco, Galileu Chagas
    Dantas, Francois
    Carvalho, Tiago Silva E.
    Rodrigues Firmino, Roberti Uili
    Filgueiras Filho, Marcelo Quesado
    Tosello, Renato Tavares
    Leao, Claudio Jose
    Piske, Ronie Leo
    Darwich, Rogerio Zenobio
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2023, 15 (08) : 776 - 780
  • [2] Endovascular treatment of complex intracranial aneurysms by pipeline flow-diverter embolization device: a single-center experience
    Keskin, Fatih
    Erdi, Fatih
    Kaya, Bulent
    Poyraz, Necdet
    Keskin, Suat
    Kalkan, Erdal
    Ozbek, Orhan
    Koc, Osman
    NEUROLOGICAL RESEARCH, 2015, 37 (04) : 359 - 365
  • [3] Study Protocol DEDICATE: Italian Multicenter Study on the Treatment of Visceral Aneurysms with the Derivo Peripher and Derivo 2 Flow Diverter Stent
    Discalzi, Andrea
    Nardelli, Floriana
    Ficara, Bruno
    Comelli, Simone
    de Donato, Gianmarco
    Fanelli, Fabrizio
    Quaretti, Pietro
    Sallemi, Claudio
    Semeraro, Vittorio
    Sirovich, Roberta
    Calandri, Marco
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2025, : 694 - 701
  • [4] Flow diverter device for the treatment of small middle cerebral artery aneurysms
    Briganti, Francesco
    Delehaye, Luigi
    Leone, Giuseppe
    Sicignano, Carmine
    Buono, Giuseppe
    Marseglia, Mariano
    Caranci, Ferdinando
    Tortora, Fabio
    Maiuri, Francesco
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2016, 8 (03) : 287 - 294
  • [5] The Derivo Embolization Device, a Second-Generation Flow Diverter for the Treatment of Intracranial Aneurysms, Evaluated in an Elastase-Induced Aneurysm Model
    Ley, D.
    Muehl-Benninghaus, R.
    Yilmaz, U.
    Koerner, H.
    Cattaneo, G. F. M.
    Mailaender, W.
    Kim, Y-J.
    Scheller, B.
    Reith, W.
    Simgen, A.
    CLINICAL NEURORADIOLOGY, 2017, 27 (03) : 335 - 343
  • [6] Endovascular treatment of wide-necked paraophthalmic aneurysms with flow diverter stent implantation
    Guzeldag, Hamit
    Onan, Bilen
    Piskin, Ferhat Can
    Sozutok, Sinan
    CUKUROVA MEDICAL JOURNAL, 2021, 46 (04): : 1666 - 1674
  • [7] Endovascular treatment of unruptured aneurysms of cavernous and ophthalmic segment of internal carotid artery with flow diverter device Pipeline
    Jevsek, Marko
    Mounayer, Charbel
    Seruga, Tomaz
    RADIOLOGY AND ONCOLOGY, 2016, 50 (04) : 378 - 384
  • [8] The Experience of the Derivo® Embolisation Device in Intracranial Aneurysms
    Daglioglu, Ergun
    Akmangit, Ilkay
    Acik, Vedat
    Alagoz, Fatih
    Sayin, Bige
    Uckun, Ozhan Merzuk
    Belen, Ahmet Deniz
    Arat, Anil
    TURKISH NEUROSURGERY, 2020, 30 (01) : 30 - 37
  • [9] Hemodynamic Characteristics and Clinical Outcome for Intracranial Aneurysms Treated with the Derivo Embolization Device, a Novel Second-Generation Flow Diverter
    Fujimura, Soichiro
    Brehm, Alex
    Takao, Hiroyuki
    Uchiyama, Yuya
    Karagiozov, Kostadin
    Fukudome, Koji
    Yamamoto, Makoto
    Murayama, Yuichi
    Psychogios, Marios-Nikos
    WORLD NEUROSURGERY, 2022, 159 : E252 - E259
  • [10] Flow diverter stents for endovascular treatment of aneurysms: a comparative study of efficacy and safety between FREDX and FRED
    Guimaraens, Leopoldo
    Saldana, Jesus
    Vivas, Elio
    Cifuentes, Sebastian
    Balaguer, Ernest
    Mon, Dunia
    Macias-Gomez, Adria
    Ois, Angel
    Guisado-Alonso, Daniel
    Cuadrado-Godia, Elisa
    Jimenez-Balado, Joan
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2024, : e159 - e165