Impact of left atrial appendage occlusion device position on potential determinants of device-related thrombus: a patient-specific in silico study

被引:4
|
作者
Zhong, Zhaoyang [1 ]
Gao, Yiting [1 ]
Kovacs, Soma [1 ]
Vij, Vivian [2 ]
Nelles, Dominik [2 ]
Spano, Lukas [1 ]
Nickenig, Georg [2 ]
Sonntag, Simon [3 ]
De Backer, Ole [4 ]
Sondergaard, Lars [4 ]
Sedaghat, Alexander [2 ,5 ]
Mela, Petra [1 ]
机构
[1] Tech Univ Munich, Munich Inst Biomed Engn, Chair Med Mat & Implants, TUM Sch Engn & Design, Garching, Germany
[2] Univ Hosp Bonn, Dept Cardiol, Bonn, Germany
[3] Virtonomy GmbH, Munich, Germany
[4] Copenhagen Univ Hosp, Dept Cardiol, Rigshosp, Copenhagen, Denmark
[5] RheinAhrCardio, Praxis Kardiol, Bad Neuenahr Ahrweiler, Germany
关键词
Left atrial appendage occlusion; Device-related thrombus; Computational fluid dynamics; Patient-specific modeling; PREVENT STROKE; FLOW; FIBRILLATION; NEUTROPHILS; PREDICTORS; DEPOSITION;
D O I
10.1007/s00392-023-02228-x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Device-related thrombus (DRT) after left atrial appendage occlusion (LAAO) is potentially linked to adverse events. Although clinical reports suggest an effect of the device type and position on the DRT risk, in-depth studies of its mechanistic basis are needed. This in silico study aimed to assess the impact of the position of non-pacifier (Watchman) and pacifier (Amulet) LAAO devices on surrogate markers of DRT risk. Methods The LAAO devices were modeled with precise geometry and virtually implanted in different positions into a patient-specific left atrium. Using computational fluid dynamics, the following values were quantified: residual blood, wall shear stress (WSS) and endothelial cell activation potential (ECAP). Results In comparison to an ostium-fitted device position, deep implantation led to more residual blood, lower average WSS and higher ECAP surrounding the device, especially on the device's atrial surface and the surrounding tissue, suggesting increased risk for potential thrombus. For the non-pacifier device, an off-axis device orientation resulted in even more residual blood, higher ECAP and similar average WSS as compared to an ostium-fitted device position. Overall, the pacifier device showed less residual blood, higher average WSS and lower ECAP, compared to the non-pacifier device. Conclusions In this in silico study, both LAAO device type and implant position showed an impact on potential markers of DRT in terms of blood stasis, platelet adhesion and endothelial dysfunction. Our results present a mechanistic basis for clinically observed risk factors of DRT and the proposed in silico model may aid in the optimization of device development and procedural aspects. [GRAPHICS] .
引用
收藏
页码:1405 / 1418
页数:14
相关论文
共 50 条
  • [1] Device-related thrombus following left atrial appendage occlusion
    Simard, Trevor J.
    Hibbert, Benjamin
    Alkhouli, Mohamad Al
    Abraham, Neena S.
    Holmes, David R. Jr Jr
    EUROINTERVENTION, 2022, 18 (03) : 224 - +
  • [2] Device-Related Thrombus After Left Atrial Appendage Occlusion With the Amulet Device
    Peyrol, Michael
    Cautela, Jennifer
    Salaun, Erwan
    Miola, Charlene
    Franceschi, Frederic
    Thuny, Franck
    Pinto, Johan
    Laine, Marc
    Maximovitch, Alexandre
    Paganelli, Franck
    Armero, Sebastien
    Bonello, Laurent
    HEART LUNG AND CIRCULATION, 2019, 28 (11) : 1683 - 1688
  • [3] Challenges and advances in device-related thrombus in left atrial appendage occlusion
    Tanaka, Takamasa
    Kawai, Kenji
    Ellis, Christopher R.
    Srivastava, Mukta
    Kawakami, Rika
    Konishi, Takao
    Shiraki, Tatsuya
    Sekimoto, Teruo
    Virmani, Renu
    Finn, Aloke, V
    FUTURE CARDIOLOGY, 2024, 20 (7-8) : 343 - 358
  • [4] Left atrial appendage patency and device-related thrombus after percutaneous left atrial appendage occlusion: a computed tomography study
    Cochet, Hubert
    Iriart, Xavier
    Sridi, Soumaya
    Camaioni, Claudia
    Corneloup, Olivier
    Montaudon, Michel
    Laurent, Francois
    Selmi, Wieme
    Renou, Pauline
    Jalal, Zakaria
    Thambo, Jean-Benoit
    EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2018, 19 (12) : 1351 - 1361
  • [5] A case report of postcardioversion device-related thrombus in a patient with left atrial appendage occlusion device on apixaban
    Ahmed, Hasaan
    Ismayl, Mahmoud
    Palicherla, Anirudh
    Heppler, Miranda
    Petraskova, Terezia
    Kousa, Omar
    Vargha, Jalal
    ANNALS OF MEDICINE AND SURGERY, 2024, 86 (03): : 1729 - 1733
  • [6] Predictors of Device-Related Thrombus Following Percutaneous Left Atrial Appendage Occlusion
    Simard, Trevor
    Jung, Richard G.
    Lehenbauer, Kyle
    Piayda, Kerstin
    Pracon, Radoslaw
    Jackson, Gregory G.
    Flores-Umanzor, Eduardo
    Faroux, Laurent
    Korsholm, Kasper
    Chun, Julian K. R.
    Chen, Shaojie
    Maarse, Moniek
    Montrella, Kristi
    Chaker, Zakeih
    Spoon, Jocelyn N.
    Pastormerlo, Luigi E.
    Meincke, Felix
    Sawant, Abhishek C.
    Moldovan, Carmen M.
    Qintar, Mohammed
    Aktas, Mehmet K.
    Branca, Luca
    Radinovic, Andrea
    Ram, Pradhum
    El-Zein, Rayan S.
    Flautt, Thomas
    Ding, Wern Yew
    Sayegh, Bassel
    Benito-Gonzalez, Tomas
    Lee, Oh-Hyun
    Badejoko, Solomon O.
    Paitazoglou, Christina
    Karim, Nabeela
    Zaghloul, Ahmed M.
    Agrawal, Himanshu
    Kaplan, Rachel M.
    Alli, Oluseun
    Ahmed, Aamir
    Suradi, Hussam S.
    Knight, Bradley P.
    Alla, Venkata M.
    Panaich, Sidakpal S.
    Wong, Tom
    Bergmann, Martin W.
    Chothia, Rashaad
    Kim, Jung-Sun
    Perez de Prado, Armando
    Bazaz, Raveen
    Gupta, Dhiraj
    Valderrabano, Miguel
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2021, 78 (04) : 297 - 313
  • [7] Device-Related Thrombus After Left Atrial Appendage Occlusion The Villain of the Piece
    Friedman, Daniel J.
    Piccini, Jonathan P.
    JACC-CLINICAL ELECTROPHYSIOLOGY, 2023, 9 (01) : 108 - 110
  • [8] Left Atrial Appendage Occlusion Device-Related Thrombus Certainties and Uncertainties
    Valderrabano, Miguel
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2018, 71 (14) : 1537 - 1539
  • [9] Left atrial appendage occlusion device-related thrombus: A case report and review of the literature
    Mohammed Andaleeb Chowdhury
    Wael Elabbassi
    Arif Al Nooryani
    Journal of Echocardiography, 2012, 10 (2) : 65 - 66
  • [10] Left atrial appendage occlusion device-related thrombus: a case report and review of the literature
    Chowdhury, Mohammed Andaleeb
    Elabbassi, Wael
    Al Nooryani, Arif
    JOURNAL OF ECHOCARDIOGRAPHY, 2012, 10 (02) : 65 - 66