Incomplete Screw Thread Engagement of Proximal Fragment: A Possible Failure Risk After Internal Fixation for Femoral Neck Fractures

被引:1
|
作者
Tanaka, Atsuki [1 ,2 ]
Hiranaka, Takafumi [1 ,2 ]
Fujishiro, Takaaki [1 ,2 ]
Koide, Motoki [1 ,2 ]
Suda, Yoshihito [1 ,2 ]
Saito, Akira [1 ,2 ]
Arimoto, Akihiko [1 ,2 ]
Okamoto, Koji [1 ,2 ]
机构
[1] Takatsuki Gen Hosp, Dept Orthopaed Surg, Takatsuki, Japan
[2] Takatsuki Gen Hosp, Joint Surg Ctr, Takatsuki, Japan
关键词
incomplete thread engagement; fixation failure; sliding mechanism; thread length; internal fixation; femoral neck fracture; OUTCOMES;
D O I
10.7759/cureus.41349
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background For successful internal fixation for femoral neck fracture, the sliding mechanism of the screw is important because it can induce inter-fragmental compression. The thread should penetrate the fracture line and be located within the proximal fragment. If screw thread engagement is incomplete and a part of the thread remains within the distal fragment, the screw sliding can be disturbed, potentially leading to fixation failure. We hypothesized that screw thread in the fracture is a risk of fixation failure. Methods We studied 133 hips that underwent internal fixation for femoral neck fracture using dual sliding and compression screws (DSCS) with 20 mm threads. The existence of incomplete thread engagement and fixation failure (cut out, perforation, pseudoarthrosis, or femoral neck shortening) were evaluated on anteroposterior hip radiography postoperatively. The distances from the thread end to the fracture line, screw head to the femoral head cortex, and femoral head diameter were measured to analyze their relationships with any incomplete thread engagement and fixation failure. Differences in evaluation data were assessed using Fisher's exact test, Student's t -test, and receiver operating characteristic (ROC) analysis. Results Forty-six cases had at least one screw with incomplete thread engagement, and the other 87 hips had a complete engagement. The failure rate in the group of hips with incomplete thread engagement was significantly higher (7/46, 15.2%) than that in the group of hips with complete thread engagement (3/87, 3.4%) (P = 0.032). Incomplete thread engagement was found in 59 out of 266 screws (22.2%), and a femoral head <= 43.9 mm in diameter was associated with an increased risk of incomplete thread engagement. Most incomplete thread engagement screws (81.4%) had < 5 mm thread length within the distal fragment. Conclusion A partially threaded screw is a significant risk of fixation failure after internal fixation for a femoral neck fracture. The smaller femoral head diameter increases the possibility of incomplete thread engagement. Shortening the thread length by 5 mm may help to avoid incomplete thread engagement.
引用
收藏
页数:11
相关论文
共 50 条
  • [11] Subtrochanteric fracture : A rare but severe complication after screw fixation of femoral neck fractures in the elderly
    Jansen, Hendrik
    Frey, Soenke P.
    Meffert, Rainer H.
    ACTA ORTHOPAEDICA BELGICA, 2010, 76 (06): : 778 - 784
  • [12] Analysis of risk factors for modification of femoral offset after internal fixation of Garden I and II femoral neck fractures
    Druart, Thomas
    Freger, Nicolas
    Pries, Pierre
    Losson, Alexandre
    Bouget, Pierre
    Vendeuvre, Tanguy
    Severyns, Mathieu
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2024, 110 (08)
  • [13] Comparison of femoral neck shortening after femoral neck system and cannulated cancellous screw fixation for displaced femoral neck fractures in young adults
    Zheng, Shunze
    Lin, Dongze
    Chen, Peisheng
    Lin, Chaohui
    Chen, Bin
    Zheng, Ke
    Lin, Fengfei
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2024, 55 (06):
  • [14] Biomechanical Comparison of Dynamic Hip Screw, Proximal Femoral Nail, Cannulated Screw, and Monoaxial External Fixation in the Treatment of Basicervical Femoral Neck Fractures
    Imren, Y.
    Gurkan, V.
    Bilsel, K.
    Desteli, E. E.
    Tuna, M.
    Gurcan, C.
    Tuncay, I.
    Sen, C.
    ACTA CHIRURGIAE ORTHOPAEDICAE ET TRAUMATOLOGIAE CECHOSLOVACA, 2015, 82 (02) : 140 - 144
  • [15] Implant Choice of Internal Fixation for Stable Femoral Neck Fractures in Elderly: Cannulated Screw Fixation versus Helical Blade Cephalomedullary Nailing
    Roh, Y. H.
    Ahn, J. S.
    Lim, C. M.
    Nam, K. W.
    MALAYSIAN ORTHOPAEDIC JOURNAL, 2025, 19 (01) : 102 - 112
  • [16] Risk Stratification for Avascular Necrosis of the Femoral Head After Internal Fixation of Femoral Neck Fractures by Post-Operative Bone SPECT/CT
    Han S.
    Oh M.
    Yoon S.
    Kim J.
    Kim J.-W.
    Chang J.-S.
    Ryu J.-S.
    Nuclear Medicine and Molecular Imaging, 2017, 51 (1) : 49 - 57
  • [17] Can we predict failure of internal fixation when treating femoral neck fractures?
    Kemler, M. A.
    de Vries, M.
    Gerard, W. J. E.
    van Raay, J. J. A. M.
    van der Tol, A.
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2007, 17 (01) : 67 - 70
  • [18] Radiographic fracture features predicting failure of internal fixation of displaced femoral neck fractures
    N. L. Weil
    D. van Embden
    J. M. Hoogendoorn
    European Journal of Trauma and Emergency Surgery, 2015, 41 : 501 - 507
  • [19] Radiographic fracture features predicting failure of internal fixation of displaced femoral neck fractures
    Weil, N. L.
    van Embden, D.
    Hoogendoorn, J. M.
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2015, 41 (05) : 501 - 507
  • [20] Risk factors for cut-out of double lag screw fixation in proximal femoral fractures
    Buyukdogan, Kadir
    Caglar, Omur
    Isik, Samet
    Tokgozoglu, Mazhar
    Atilla, Bulent
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2017, 48 (02): : 414 - 418