Triple-tuned birdcage and single-tuned dipole array for quadri-nuclear head MRI at 7 T

被引:0
作者
Paska, Jan [1 ,2 ]
Wang, Bili [1 ,2 ]
Chen, Anna M. [1 ,2 ]
Madelin, Guillaume [1 ,2 ]
Brown, Ryan [1 ,2 ,3 ]
机构
[1] NYU, Ctr Adv Imaging Innovat & Res, Grossman Sch Med, Dept Radiol, New York, NY USA
[2] NYU, Grossman Sch Med, Bernard & Irene Schwartz Ctr Biomed Imaging, Dept Radiol, New York, NY USA
[3] Ctr Biomed Imaging, Dept Radiol, 660 First Ave,4th Floor, New York, NY 10016 USA
关键词
deuterium-H-2; MRI; phosphorus-P-31; MRSI; sodium-Na-23; simultaneous multi-nuclear MRI; ultrahigh field MRI; BRAIN ENERGY-METABOLISM; MAGNETIC-RESONANCE-SPECTROSCOPY; TRAP DESIGN; SODIUM MRI; RF COIL; P-31; QUADRATURE; H-1;
D O I
10.1002/mrm.29977
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this work was to design and build a coil for quadri-nuclear MRI of the human brain at 7 T.Methods: We built a transmit/receive triple-tuned (45.6 MHz for H-2, 78.6 MHz for Na-23, and 120.3 MHz for P-31) quadrature four-rod birdcage that was geometrically interleaved with a transmit/receive four-channel dipole array (297.2 MHz for 1H). The birdcage rods contained passive, two-pole resonant circuits that emulated capacitors required for single-tuning at three frequencies. The birdcage assembly also included triple-tuned matching networks, baluns, and transmit/receive switches. We assessed the performance of the coil with quality factor (Q) and signal-to-noise ratio (SNR) measurements, and performed in vivo multinuclear MRI and MR spectroscopic imaging (MRSI).Results: Q measurements showed that the triple-tuned birdcage efficiency was within 33% of that of single-tuned baseline birdcages at all three frequencies. The quadri-tuned coil SNR was 78%, 59%, 44%, and 48% lower than that of single or dual-tuned reference coils for H-1, H-22, Na-23, and 3131 P, respectively. Quadri-nuclear MRI and MRSI was demonstrated in brain in vivo in about 30 min.Conclusion: While the SNR of the quadruple tuned coil was significantly lower than dual- and single-tuned reference coils, it represents a step toward truly simultaneous quadri-nuclear measurements.
引用
收藏
页码:2188 / 2199
页数:12
相关论文
共 58 条
[31]   Interleaved fluid-attenuated inversion recovery (FLAIR) MRI and deuterium metabolic imaging (DMI) on human brain in vivo [J].
Liu, Yanning ;
De Feyter, Henk M. ;
Fulbright, Robert K. ;
McIntyre, Scott ;
Nixon, Terence W. ;
de Graaf, Robin A. .
MAGNETIC RESONANCE IN MEDICINE, 2022, 88 (01) :28-37
[32]  
Madelin G., 2022, X NUCLEI MAGNETIC RE
[33]   Biomedical applications of sodium MRI in vivo [J].
Madelin, Guillaume ;
Regatte, Ravinder R. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2013, 38 (03) :511-529
[34]   Cellular mechanisms of brain energy metabolism and their relevance to functional brain imaging [J].
Magistretti, PJ ;
Pellerin, L .
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY B-BIOLOGICAL SCIENCES, 1999, 354 (1387) :1155-1163
[35]  
Matson GB, 1999, MAGNET RESON MED, V42, P173, DOI 10.1002/(SICI)1522-2594(199907)42:1<173::AID-MRM23>3.0.CO
[36]  
2-O
[37]   Comparison of MEMS switches and PIN diodes for switched dual tuned RF coils [J].
Maunder, Adam ;
Rao, Madhwesha ;
Robb, Fraser ;
Wild, Jim M. .
MAGNETIC RESONANCE IN MEDICINE, 2018, 80 (04) :1746-1753
[38]   Simultaneous and Interleaved Acquisition of NMR Signals from Different Nuclei with a Clinical MRI Scanner [J].
Meyerspeer, Martin ;
Magill, Arthur W. ;
Kuehne, Andre ;
Gruetter, Rolf ;
Moser, Ewald ;
Schmid, Albrecht Ingo .
MAGNETIC RESONANCE IN MEDICINE, 2016, 76 (05) :1636-1641
[39]   An Improved Trap Design for Decoupling Multinuclear RF Coils [J].
Meyerspeer, Martin ;
Roig, Eulalia Seres ;
Gruetter, Rolf ;
Magill, Arthur W. .
MAGNETIC RESONANCE IN MEDICINE, 2014, 72 (02) :584-590
[40]  
Mispelter J., 2006, NMR PROBEHEADS BIOPH, DOI DOI 10.1142/P438