Hemodynamic Wall Shear Stress Measurements Using Ultrasound Feature Tracking Algorithm

被引:0
作者
Park, Dong Chan [1 ]
Park, Dae Woo [1 ]
Chang, Jun Geun [1 ]
机构
[1] Natl Canc Ctr, Res Inst & Hosp, Diagnost & Therapeut Technol Branch, 323 Ilsan Ro, Goyang Si 10408, Gyeonggi Do, South Korea
关键词
Ultrasound; Wall shear stress; Wall shear rate; Blood flow; Ultrasound feature tracking; PARTICLE IMAGE VELOCIMETRY; SPECKLE TRACKING; IN-VITRO; FLOW; QUANTIFICATION;
D O I
10.1007/s42835-023-01489-w
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Accurate wall shear rate (WSR) measurements are important for the early diagnosis of cardiovascular disease progression and acute events, such as atherosclerotic plaque rupture and aneurysm rupture. The Kanade-Lucas-Tomasi (KLT) feature tracking is an ultrasound imaging technique and competitive candidate for WSR measurements. In this study, we hypothesized that the accuracy of WSR measurements can be improved using KLT feature tracking. Thereby, we aimed to evaluate the accuracy of KLT feature tracking for flow velocities and WSR measurements by comparing the results with those from the conventional two-dimensional ultrasound speckle tracking (2D-UST) technique. An in-vitro flow experiment was performed using a flow phantom, and blood flow velocity profiles were generated using KLT feature tracking and the 2D-UST algorithm. A quantitative error analysis was performed for the flow velocities and WSRs measured using both methods based on the analytical values. The WSRs measured using KLT feature tracking demonstrated significantly higher agreement with the analytical values than those measured using 2D-UST. The measured WSRs using KLT feature tracking were found to match the analytical values with absolute errors of 0.9% and 14.4% at the top and bottom luminal walls, respectively. By contrast, the estimated WSRs using the 2D-UST algorithm were found to be in poor agreement with the analytical values, with absolute errors of 80.7% and 67.1% at the top and bottom luminal walls, respectively. Therefore, our results demonstrate the feasibility of KLT feature tracking for clinical application of in-vivo wall shear stress measurements.
引用
收藏
页码:3111 / 3116
页数:6
相关论文
共 24 条
[1]   Adaptive Spatiotemporal SVD Clutter Filtering for Ultrafast Doppler Imaging Using Similarity of Spatial Singular Vectors [J].
Baranger, Jerome ;
Arnal, Bastien ;
Perren, Fabienne ;
Baud, Olivier ;
Tanter, Mickael ;
Demene, Charlie .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (07) :1574-1586
[2]   A pilot study to measure vascular compliance changes during fistula maturation using open-source software [J].
Belmont, Barry ;
Park, Dae Woo ;
Shih, Albert ;
Leavitt, Matthew ;
Plattner, Brett ;
Henke, Peter K. ;
Weitzel, William F. .
JOURNAL OF VASCULAR ACCESS, 2019, 20 (01) :41-45
[3]   An Open-Source Ultrasound Software for Diagnosis of Fistula Maturation [J].
Belmont, Barry ;
Park, Dae Woo ;
Weitzel, William F. ;
Shih, Albert J. .
ASAIO JOURNAL, 2018, 64 (01) :70-76
[4]   Speckle tracking for multi-dimensional flow estimation [J].
Bohs, LN ;
Geiman, BJ ;
Anderson, ME ;
Gebhart, SC ;
Trahey, GE .
ULTRASONICS, 2000, 38 (1-8) :369-375
[5]   Aneurysm Growth Occurs at Region of Low Wall Shear Stress Patient-Specific Correlation of Hemodynamics and Growth in a Longitudinal Study [J].
Boussel, Loic ;
Rayz, Vitaliy ;
McCulloch, Charles ;
Martin, Alastair ;
Acevedo-Bolton, Gabriel ;
Lawton, Michael ;
Higashida, Randall ;
Smith, Wade S. ;
Young, William L. ;
Saloner, David .
STROKE, 2008, 39 (11) :2997-3002
[6]   Quantification of wall shear stress in large blood vessels using lagrangian interpolation functions with cine phase-contrast magnetic resonance imaging [J].
Cheng, CP ;
Parker, D ;
Taylor, CA .
ANNALS OF BIOMEDICAL ENGINEERING, 2002, 30 (08) :1020-1032
[7]   Effects of Disturbed Flow on Vascular Endothelium: Pathophysiological Basis and Clinical Perspectives [J].
Chiu, Jeng-Jiann ;
Chien, Shu .
PHYSIOLOGICAL REVIEWS, 2011, 91 (01) :327-387
[8]   An automatic fascicle tracking algorithm quantifying gastrocnemius architecture during maximal effort contractions [J].
Drazan, John F. ;
Hullfish, Todd J. ;
Baxter, Josh R. .
PEERJ, 2019, 7
[9]   SHUNT FLOW EVALUATION IN CONGENITAL HEART DISEASE BASED ON TWO-DIMENSIONAL SPECKLE TRACKING [J].
Fadnes, Solveig ;
Nyrnes, Siri Ann ;
Torp, Hans ;
Lovstakken, Lasse .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2014, 40 (10) :2379-2391
[10]   Three-Dimensional Analysis of Segmental Wall Shear Stress in the Aorta by Flow-Sensitive Four-Dimensional-MRI [J].
Frydrychowicz, Alex ;
Stalder, Aurelien F. ;
Russe, Maximilian F. ;
Bock, Jelena ;
Bauer, Simon ;
Harloff, Andreas ;
Berger, Alexander ;
Langer, Mathias ;
Hennig, Juergen ;
Markl, Michael .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2009, 30 (01) :77-84