A Novel Triple-Bladder Cuff Method for Blood Pressure Estimation Based on Pulse Wave Dynamics in Brachial Artery: Theoretical and Experimental Analyses

被引:0
作者
Uemura, Kazunori [1 ]
Nishikawa, Takuya [2 ]
Hasebe, Naotaka [3 ]
Yokota, Shohei [2 ]
Kakuuchi, Midori [2 ]
Saku, Keita [2 ]
Kawada, Toru [2 ]
Morodome, Shohei [3 ]
Furukoshi, Masaki [3 ]
Sugimachi, Masaru [2 ]
机构
[1] Natl Cerebral & Cardiovasc Ctr, Dept Cardiovasc Dynam, Suita, Osaka 5648565, Japan
[2] Natl Cerebral & Cardiovasc Ctr, Dept Cardiovasc Dynam, Suita, Japan
[3] A&D Co Ltd, Tokyo, Japan
关键词
Bladder; Valves; Biomedical measurement; Estimation; Brachytherapy; Arteries; Sensors; Blood pressure measurement; triple-bladder cuff; cross-correlation; pulse wave velocity; transmural pressure; OSCILLOMETRIC MAXIMUM; TRANSIT-TIME; FLOW; RECOMMENDATIONS; ASSOCIATION; ALGORITHM; STIFFNESS; ACCURACY;
D O I
10.1109/TBME.2022.3201314
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: The objective of this study was to develop a novel triple-bladder cuff method for accurate and automated estimation of systolic (SBP) and diastolic (DBP) blood pressure and validate its reliability in animal experiments. Methods: The cuff is composed of three bladders each measured one-third the width of a conventional BP cuff, which are designed to measure oscillatory pulsation at the proximal, middle, and distal segments of the upper arm. This structure allows evaluation of the pulse wave propagation in the brachial artery under the cuff. SBP is estimated (SBPe) by detecting resumption of systolic arterial flow based on statistical similarity in oscillatory pulse traces between the proximal and distal segments. DBP is estimated (DBPe) based on the relation between pulse wave velocity and transmural pressure at diastole in the brachial artery. In 7 anesthetized goats, we compared SBPe and DBPe to reference SBP and DBP, respectively, measured by an intra-arterial catheter. BP was perturbed by infusing nitroprusside or noradrenaline. Results: SBP correlated strongly with SBPe in each animal [mean coefficient of determination (R-2) = 0.98 +/- 0.01]. Mean +/- standard deviation of errors between SBP and SBPe was 0.0 +/- 4.9 mmHg. DBP correlated strongly with DBPe in each animal (R-2 = 0.96 +/- 0.03). Mean +/- standard deviation of errors between DBP and DBPe was 0.0 +/- 6.3 mmHg. Conclusion: This method estimates SBP and DBP with acceptable accuracy. Significance: Accurate and automated BP estimation by this method may potentially optimize antihypertensive treatment in patients with hypertension.
引用
收藏
页码:703 / 714
页数:12
相关论文
共 38 条
  • [1] Excess pressure as an analogue of blood flow velocity
    Armstrong, Matthew K.
    Schultz, Martin G.
    Hughes, Alun D.
    Picone, Dean S.
    Black, J. Andrew
    Dwyer, Nathan
    Roberts-Thomson, Philip
    Sharman, James E.
    [J]. JOURNAL OF HYPERTENSION, 2021, 39 (03) : 421 - 427
  • [2] Theoretical analysis of non-invasive oscillometric maximum amplitude algorithm for estimating mean blood pressure
    Baker, PD
    Westenskow, DR
    Kuck, K
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (03) : 271 - 278
  • [3] DIRECT EFFECTS OF SMOOTH-MUSCLE RELAXATION AND CONTRACTION ON IN-VIVO HUMAN BRACHIAL-ARTERY ELASTIC PROPERTIES
    BANK, AJ
    WILSON, RF
    KUBO, SH
    HOLTE, JE
    DRESING, TJ
    WANG, HY
    [J]. CIRCULATION RESEARCH, 1995, 77 (05) : 1008 - 1016
  • [4] Agreement between methods of measurement with multiple observations per individual
    Bland, J. Martin
    Altman, Douglas G.
    [J]. JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) : 571 - 582
  • [6] Smartphone-based blood pressure monitoring via the oscillometric finger-pressing method
    Chandrasekhar, Anand
    Kim, Chang-Sei
    Naji, Mohammed
    Natarajan, Keerthana
    Hahn, Jin-Oh
    Mukkamala, Ramakrishna
    [J]. SCIENCE TRANSLATIONAL MEDICINE, 2018, 10 (431)
  • [7] THEORY OF THE OSCILLOMETRIC MAXIMUM AND THE SYSTOLIC AND DIASTOLIC DETECTION RATIOS
    DRZEWIECKI, G
    HOOD, R
    APPLE, H
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 1994, 22 (01) : 88 - 96
  • [8] Oscillometric blood pressure estimation: Past, present, and future
    School of Electrical Engineering and Computer Science, University of Ottawa, Ottawa
    K1N 6N5, Canada
    不详
    MA
    02138, United States
    [J]. IEEE Rev. Biomed. Eng., (44-63): : 44 - 63
  • [9] Coefficient-Free Blood Pressure Estimation Based on Pulse Transit Time-Cuff Pressure Dependence
    Forouzanfar, Mohamad
    Ahmad, Saif
    Batkin, Izmail
    Dajani, Hilmi R.
    Groza, Voicu Z.
    Bolic, Miodrag
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2013, 60 (07) : 1814 - 1824
  • [10] Accuracy of the pulse delay time technique with triple cuff for objective indirect blood pressure measurement
    Fujikawa, Tetsuya
    Tochikubo, Osamu
    Sugano, Teruyasu
    Umemura, Satoshi