Accuracy of Intraocular Lens Power Calculation Based on Total Keratometry in Patients With Flat and Steep Corneas

被引:9
|
作者
Qin, Y. I. N. G. Y. A. N. [1 ,2 ]
Liu, L. I. A. N. G. P. I. N. G. [1 ,2 ]
Mao, Y. A. N. [1 ,2 ]
Ding, Y. U. J. I. E. [1 ,2 ]
Ye, S. H. A. O. B., I [1 ,2 ]
Sun, A. O. [1 ,2 ]
Wu, M. I. N. G. X. I. N. G. [1 ,2 ,3 ]
机构
[1] Sun Yat sen Univ, Zhongshan Ophthalm Ctr, State Key Lab Ophthalmol, Guangdong Prov Key Lab Ophthalmol & Visual Sci, Guangzhou, Peoples R China
[2] Guangdong Prov Clin Res Ctr Ocular Dis, Guangzhou, Peoples R China
[3] Sun Yat sen Univ, Zhongshan Ophthalm Ctr, Guangzhou, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
ASTIGMATISM;
D O I
10.1016/j.ajo.2022.11.011
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
center dot PURPOSE: To analyze the accuracy of the current in-traocular lens power calculation formulas using standard keratometry (K) and total keratometry (TK) data in pa-tients with fiat and steep corneas. center dot DESIGN: Retrospective consecutive cross-sectional study. center dot METHODS: An optical biometer with swept-source op-tical coherence tomography was used in this retrospective study. The standard deviation (SD), mean absolute error (MAE), median absolute error (MedAE), and the pro-portion of eyes with prediction error (PE) within +/- 0.25 diopter (D), +/- 0.5 D, +/- 0.75 D, and +/- 1.00 D were calcu-lated to evaluate the refractive outcomes of each formula. center dot RESULTS: A total of 231 eyes from 231 patients were included. In the entire study cohort, the Emmetropia Verifying Optical (EVO) formula using TK data showed the lowest SD (0.383) and MAE (0.30) and the highest percentage of cases with a PE within +/- 0.5 D (81.4%). In the fiat keratometry group, the EVO ( P = .042), Haigis ( P = .043), Hoffer Q ( P = .038) and Holladay 1 ( P = .013) formulas using TK data had significantly lower SD than using K data. The EVO formula using TK data showed the lowest SD (0.357) and MAE (0.28). In the steep keratometry group, the Hoffer Q ( P = .036) and SRK/T ( P = .029) formulas using TK data had sig-nificantly lower SD than using K data. The BUII TK formula showed the lowest SD (0.431), MedAE (0.26), and MAE (0.32). center dot CONCLUSION: The TK data set showed a better trend of refractive outcomes, especially in the fiat and steep ker-atometry groups. EVO (TK) and BUII TK formulas were suggested for eyes with K values lower than 42 D and K values higher than 46 D, respectively. (Am J Ophthal-mol 2023;247: 103-110. (c) 2022 Elsevier Inc. All rights reserved.)
引用
收藏
页码:103 / 110
页数:8
相关论文
共 24 条
  • [21] Effect of two different preoperative calculation schemes on visual outcomes of patients after toric intraocular lens implantation
    Xuewen Yu
    Jingwen Wang
    Xueqi Lin
    Dejian Xu
    Shuang Ni
    Siting Sheng
    Wen Xu
    International Ophthalmology, 2023, 43 : 491 - 501
  • [22] Accuracy of Prediction of Steep Axis and Cylinder Power Using IOL Master versus Pentacam in Patients Undergoing Toric IOL Implantation
    Brown, Sarah Elise
    Miller, Darby
    Ostrovsky, Ann
    Wallenstein, Sylvan
    Speaker, Mark
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [23] Comparative meta-analysis of toric intraocular lens alignment accuracy in cataract patients: Image-guided system versus manual marking
    Zhou, Fengqi
    Jiang, Wu
    Lin, Zhuoling
    Li, Xiaoyan
    Li, Jing
    Lin, Haotian
    Chen, Weirong
    Wang, Qiwei
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (09) : 1340 - 1345
  • [24] Artificial intelligence-based refractive error prediction and EVO-implantable collamer lens power calculation for myopia correction
    Jiang, Yinjie
    Shen, Yang
    Chen, Xun
    Niu, Lingling
    Li, Boliang
    Cheng, Mingrui
    Lei, Yadi
    Xu, Yilin
    Wang, Chongyang
    Zhou, Xingtao
    Wang, Xiaoying
    EYE AND VISION, 2023, 10 (01)