The Effects of Stimulus Repetition Rate on Electrically Evoked Auditory Brainstem Potentials in Postlingually Deafened Adult Cochlear Implant Recipients

被引:0
作者
Dziemba, Oliver C. [1 ]
Brzoska, Tina [1 ]
Hocke, Thomas [2 ]
Ihler, Friedrich [1 ]
机构
[1] Univ Med Greifswald, Dept Otorhinolarygol Head & Neck Surg, D-17475 Greifswald, Germany
[2] Cochlear Deutschland GmbH & Co KG, D-30539 Hannover, Germany
关键词
objective measurement; cochlear implant; differential diagnostics; CHILDREN;
D O I
10.3390/jcm12227188
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: By using outcome prediction scores, it is possible to distinguish between good and poor performers with cochlear implants (CI) after CI implantation. The reasons for poor performance, despite good basic conditions, can be manifold. On the one hand, the postoperative fitting may be inadequate; on the other, neurophysiological disease processes may impair speech understanding with a CI. These disease processes are not yet fully understood. In acoustics, it is known that the auditory brainstem responses (ABR) and their latencies and amplitudes allow differential diagnosis based on reference values for normal-hearing individuals. The aim of this study was to provide reference values for electrically evoked brainstem responses (EABRs) in terms of rate-dependent latencies and amplitudes. Methods: 20 ears of 18 experienced adult CI recipients with a predicted and measured good postoperative word recognition score were recruited from the clinic's patient pool. In the same stimulation mode and intensity we measured latencies and interpeak-latencies of EABRs and electrically evoked compound action potentials (ECAPs). With a defined supra-threshold stimulation intensity above the individual ECAP threshold, we applied stimulation at several rates between 11 and 91 stimuli per second. Results: We found rate dependences for EABR latency t3 and t5 in the order of 0.19 ms and 0.37 ms, respectively, while ECAP was not affected by rate. Correspondingly, the interpeak intervals' rate dependences for t5-t1, t5-t3 and t3-t1 were of the order of 0.37 ms, 0.18 ms and 0.19 ms. Comparing the EABR amplitudes between the stimulation rates 11/s and 81/s, we found that at 81/s the amplitudes were significantly reduced down: to 73% for A3 and 81% for A5. These rate dependences of latency and amplitude in EABR have characteristics comparable to those of acoustic ABR. Conclusions: These data may serve to provide reference values for EABR and ECAP latencies, interpeak intervals and amplitudes with respect to stimulation rate. Altered response patterns of ECAPs and EABRs to normalised stimulation modes could be used in the future to describe and classify neuropathological processes in a better-differentiated way.
引用
收藏
页数:10
相关论文
共 33 条
  • [1] [Anonymous], 2005, OTOL NEUROTOL, V26, P1097
  • [2] Atcherson S.R., 2012, Auditory Electrophysiology, P27
  • [3] Neural Response Telemetry Reconsidered: II. The Influence of Neural Population on the ECAP Recovery Function and Refractoriness
    Botros, Andrew
    Psarros, Colleen
    [J]. EAR AND HEARING, 2010, 31 (03) : 380 - 391
  • [4] Unilateral Cochlear Implants for Severe, Profound, or Moderate Sloping to Profound Bilateral Sensorineural Hearing Loss A Systematic Review and Consensus Statements
    Buchman, Craig A.
    Gifford, Rene H.
    Haynes, David S.
    Lenarz, Thomas
    O'Donoghue, Gerard
    Adunka, Oliver
    Biever, Allison
    Briggs, Robert J.
    Carlson, Matthew L.
    Dai, Pu
    Driscoll, Colin L.
    Francis, Howard W.
    Gantz, Bruce J.
    Gurgel, Richard K.
    Hansen, Marlan R.
    Holcomb, Meredith
    Karltorp, Eva
    Kirtane, Milind
    Larky, Jannine
    Mylanus, Emmanuel A. M.
    Roland, J. Thomas, Jr.
    Saeed, Shakeel R.
    Skarzynski, Henryk
    Skarzynski, Piotr H.
    Syms, Mark
    Teagle, Holly
    Van de Heyning, Paul H.
    Vincent, Christophe
    Wu, Hao
    Yamasoba, Tatsuya
    Zwolan, Terry
    [J]. JAMA OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2020, 146 (10) : 942 - 953
  • [5] Campbell KB, 1981, SENSUS, V1, P21
  • [6] Deutsche Gesellschaft fur Hals-Nasen-Ohren-Heilkunde, 2020, Kopfund Hals-Chirurgie e.V S2k-Leitlinie Cochlea-Implantat Versorgung
  • [7] Speech Perception Performance in Cochlear Implant Recipients Correlates to the Number and Synchrony of Excited Auditory Nerve Fibers Derived From Electrically Evoked Compound Action Potentials
    Dong, Yu
    Briaire, Jeroen J.
    Stronks, H. Christiaan
    Frijns, Johan H. M.
    [J]. EAR AND HEARING, 2023, 44 (02) : 276 - 286
  • [8] Dziemba OC., 2022, GMS AUDIOL, DOI DOI 10.3205/ZAUD000023
  • [9] Evaluative audiometry after cochlear implant provision. German Version
    Dziemba, Oliver C.
    Merz, Stephan
    Hocke, Thomas
    [J]. HNO, 2023, 71 (10) : 669 - 677
  • [10] Excitation characteristic of a bipolar stimulus for broadband stimulation in measurements of electrically evoked auditory potentials
    Dziemba, Oliver C.
    Hocke, Thomas
    Mueller, Alexander
    Kaftan, Holger