Influence of the Use of Transepithelial Abutments vs. Titanium Base Abutments on Microgap Formation at the Dental Implant-Abutment Interface: An In Vitro Study

被引:3
|
作者
Cascos, Rocio [1 ,2 ,3 ]
Celemin-Vinuela, Alicia [1 ]
Mory-Rubinos, Nataly [1 ]
Gomez-Polo, Cristina [4 ]
Ortega, Rocio [3 ]
Agustin-Panadero, Ruben [5 ]
Gomez-Polo, Miguel [1 ]
机构
[1] Univ Complutense Madrid, Fac Dent, Dept Conservat Dent & Orofacial Prosthodont, Madrid 28040, Spain
[2] Rey Juan Carlos Univ, Fac Hlth Sci, Dept Nursing & Estomatol, Madrid 28922, Spain
[3] European Univ Madrid, Sch Dent, Dept Prosthet Dent, Madrid 28670, Spain
[4] Univ Salamanca, Fac Med, Dept Surg, Salamanca 37007, Spain
[5] Univ Valencia, Fac Med & Dent, Dept Stomatol, Prosthodont & Occlus Unit, Valencia 46010, Spain
关键词
dental implants; dental abutments; dental implant-abutment design; dental prosthesis; implant-supported; scanning electron microscopy; PERI-IMPLANT; ACCURACY; SURVIVAL; ZIRCONIA; BEHAVIOR; CAD/CAM; FIT; FABRICATION; TOMOGRAPHY; PROSTHESES;
D O I
10.3390/ma16196532
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
This in vitro study aimed to assess the presence of microgaps at the implant-abutment interface in monolithic zirconia partial implant-supported fixed prostheses on transepithelial abutments versus Ti-base abutments.<bold>Methods: </bold>Sixty conical connection dental implants were divided into two groups (n = 30). The control group consisted of three-unit bridge monolithic zirconia connected to two implants by a transepithelial abutment. The test group consisted of monolithic zirconia three-unit restoration connected to two implants directly by a titanium base (Ti-base) abutment. The sample was subjected to thermocycling (10,000 cycles at 5 degrees C to 55 degrees C, dwelling time 50 s) and chewing simulation (300,000 cycles, under 200 N at frequencies of 2 Hz, at a 30 degrees angle). The microgap was evaluated at six points (mesiobuccal, buccal, distobuccal, mesiolingual, lingual, and distolingual) of each implant-abutment interface by using a scanning electron microscope (SEM). The data were analyzed using the Mann-Whitney U tests (p > 0.05).<bold>Results: </bold>The SEM analysis showed a smaller microgap at the implant-abutment interface in the control group (0.270 mu m) than in the test group (3.902 mu m). Statistically significant differences were observed between both groups (p < 0.05).<bold>Conclusions: </bold>The use or not of transepithelial abutments affects the microgap size. The transepithelial abutments group presented lower microgap values at the interface with the implant than the Ti-base group in monolithic zirconia partial implant-supported fixed prostheses. However, both groups had microgap values within the clinically acceptable range.
引用
收藏
页数:12
相关论文
共 33 条
  • [21] Contact analysis of gap formation at dental implant-abutment interface under oblique loading: A numerical-experimental study
    He, Yiting
    Fok, Alex
    Aparicio, Conrado
    Teng, Wei
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2019, 21 (04) : 741 - 752
  • [22] Mechanical stability of posterior implant-supported monolithic zirconia cantilever on titanium-base abutments. An in vitro study
    Riera, Carme
    Karasan, Duygu
    Sailer, Irena
    Mojon, Phillipe
    Fehmer, Vincent
    Pitta, Joao
    CLINICAL ORAL IMPLANTS RESEARCH, 2024, 35 (08) : 1042 - 1053
  • [23] Gingival Response to Dental Implant: Comparison Study on the Effects of New Nanopored Laser-Treated vs. Traditional Healing Abutments
    Ghinassi, Barbara
    Di Baldassarre, Angela
    D'Addazio, Gianmaria
    Traini, Tonino
    Andrisani, Mauro
    Di Vincenzo, Giorgio
    Gaggi, Giulia
    Piattelli, Maurizio
    Caputi, Sergio
    Sinjari, Bruna
    INTERNATIONAL JOURNAL OF MOLECULAR SCIENCES, 2020, 21 (17) : 1 - 16
  • [24] In vitro synchrotron-based radiography of micro-gap formation at the implant-abutment interface of two-piece dental implants
    Rack, A.
    Rack, T.
    Stiller, M.
    Riesemeier, H.
    Zabler, S.
    Nelson, K.
    JOURNAL OF SYNCHROTRON RADIATION, 2010, 17 : 289 - 294
  • [25] Impact of cement type and abutment height on pull-off force of zirconia reinforced lithium silicate crowns on titanium implant stock abutments: an in vitro study
    Lisa Müller
    Angelika Rauch
    Daniel R. Reissmann
    Oliver Schierz
    BMC Oral Health, 21
  • [26] Impact of cement type and abutment height on pull-off force of zirconia reinforced lithium silicate crowns on titanium implant stock abutments: an in vitro study
    Mueller, Lisa
    Rauch, Angelika
    Reissmann, Daniel R.
    Schierz, Oliver
    BMC ORAL HEALTH, 2021, 21 (01)
  • [27] Temporary Implant-Supported Single Crowns Using Titanium Base Abutments: An In Vitro Study on Bonding Stability and Pull-out Forces
    Pitta, Joao
    Bijelic-Donova, Jasmina
    Burkhardt, Felix
    Fehmer, Vincent
    Narhi, Timo
    Sailer, Irena
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2020, 33 (05) : 546 - 552
  • [28] Influence of exposure of customized dental implant abutments to different cleaning procedures: an in vitro study using AI-assisted SEM/EDS analysis
    Paul Hofmann
    Andreas Kunz
    Franziska Schmidt
    Florian Beuer
    Dirk Duddeck
    International Journal of Implant Dentistry, 9
  • [29] Influence of exposure of customized dental implant abutments to different cleaning procedures: an in vitro study using AI-assisted SEM/EDS analysis
    Hofmann, Paul
    Kunz, Andreas
    Schmidt, Franziska
    Beuer, Florian
    Duddeck, Dirk
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2023, 9 (01)
  • [30] Antibacterial Activity of Dissolved Silver Fractions Released from Silver-Coated Titanium Dental Implant Abutments: A Study on Streptococcus mutans Biofilm Formation
    Salaie, Ranj Nadhim
    Hassan, Pakhshan A.
    Meran, Zhala Dara
    Hamad, Shehab Ahmed
    ANTIBIOTICS-BASEL, 2023, 12 (07):