Evaluation of an early-stage prototype polyurethane femoral head implant for hip arthroplasty

被引:3
作者
Siroros, Nad [1 ,2 ,3 ]
Merfort, Ricarda [1 ]
Migliorini, Filippo [1 ,4 ]
Lecouturier, Sophie [1 ]
Leven, Sophia [1 ]
Praster, Maximilian [1 ]
Hildebrand, Frank [1 ]
Eschweiler, Jorg [5 ]
机构
[1] Univ Med Ctr RWTH Aachen, Dept Orthopaed Trauma & Reconstruct Surg, Aachen, Germany
[2] Chiang Mai Univ, Biomed Engn Inst, Chiang Mai, Thailand
[3] Chiang Mai Univ, Fac Engn, Dept Mech Engn, Chiang Mai, Thailand
[4] Acad Hosp Bolzano SABES ASDAA, Teaching Hosp Paracelsus Med Univ PMU, Dept Orthopaed & Trauma Surg, I-39100 Bolzano, Italy
[5] BG Hosp Bergmannstrost, Dept Trauma & Reconstruct Surg, Halle An Der Saale, Germany
关键词
Bearings; Implant; Hip arthroplasty; Polyurethane; ALAMAR BLUE; ARTICULAR-CARTILAGE; THICKNESS;
D O I
10.1016/j.jor.2023.11.067
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction Thi study evalautes a new bone-preserving femoral head cover that mimics the articular cartilage of the femoral head.Methods A specially developed polyurethane (PU) was evaluated in biocompatibility (cytotoxicity test) and mechanical response to tensile loading. In the cytotoxicity test, steam sterilized (SS) and ethylene oxide sterilized (EtO) PU samples were incubated separately in a cell culture medium. The seeded cell line MG-63 was then added to these sample-incubated cell culture mediums. One negative control group and one positive control group were also evaluated. The cells in each group were cultured for seven days before being quantified using the alamarBlue assay. In the mechanical test, the femoral head cover implants were separated into three groups of three samples. Each group represented a different implant insertion idea: direct insertion (uc sample) and another two insertion modes (is and ss samples) representing implants with enclosure mechanisms. The test consisted of distance-controlled cyclic tensile loadings followed by a failure test.Results The cytotoxicity test results show no significant difference in fluorescence intensity between the negative control, the three SS groups, and one EtO group (P > 0.05). However, the other two EtO groups exhibit significantly lower fluorescence intensity compared with the negative control (P < 0.05). In the mechanical test, the is samples have the highest cyclic loading force at 559.50 +/- 51.41 N, while the uc samples exhibit the highest force in the failure test at 632.16 +/- 50.55 N. There are no significant differences (P > 0.05) among the uc, is, and ss groups in terms of stiffness.Conclusion The cytotoxicity test and the mechanical experiment provide initial assessments of the proposed PU femoral head cover implant. The evaluation outcomes of this study could serve as a foundation for developing more functional design and testing methods, utilizing numerical simulations, and developing animal/clinical trials in the future.
引用
收藏
页码:49 / 57
页数:9
相关论文
共 38 条
  • [1] Affatato S., 2014, Perspectives in Total Hip Arthroplasty, Advances in Biomaterials and their Tribological interactions
  • [2] Aguilar HN, 2017, RMD OPEN, V3, DOI 10.1136/rmdopen-2016-000358
  • [3] Ammarullah Muhammad Imam, 2021, IOP Conference Series: Materials Science and Engineering, V1034, DOI 10.1088/1757-899X/1034/1/012001
  • [4] Minimizing Risk of Failure from Ceramic-on-Ceramic Total Hip Prosthesis by Selecting Ceramic Materials Based on Tresca Stress
    Ammarullah, Muhammad Imam
    Santoso, Gatot
    Sugiharto, S.
    Supriyono, Toto
    Wibowo, Dwi Basuki
    Kurdi, Ojo
    Tauviqirrahman, Mohammad
    Jamari, J.
    [J]. SUSTAINABILITY, 2022, 14 (20)
  • [5] Hip resurfacing: history, current status, and future
    Amstutz, Harlan C.
    Le Duff, Michel J.
    [J]. HIP INTERNATIONAL, 2015, 25 (04) : 330 - 338
  • [6] A rapid and miniaturized system using Alamar blue to assess fungal spore viability: implications for biosecurity
    Barua, Papori
    You, Ming Pei
    Bayliss, Kirsty
    Lanoiselet, Vincent
    Barbetti, Martin J.
    [J]. EUROPEAN JOURNAL OF PLANT PATHOLOGY, 2017, 148 (01) : 139 - 150
  • [7] The Analysis of Dimple Geometry on Artificial Hip Joint to the Performance of Lubrication
    Basri, Hasan
    Syahrom, A.
    Prakoso, A. T.
    Wicaksono, D.
    Amarullah, M. I.
    Ramadhoni, T. S.
    Nugraha, R. D.
    [J]. SYMPOSIUM OF EMERGING NUCLEAR TECHNOLOGY AND ENGINEERING NOVELTY (SENTEN 2018), 2019, 1198
  • [8] HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS
    BERGMANN, G
    GRAICHEN, F
    ROHLMANN, A
    [J]. JOURNAL OF BIOMECHANICS, 1993, 26 (08) : 969 - 990
  • [9] Bundesamt Statistisches, DRG-Statistik 2021-vollstationare Patientinnen und Patienten in Krankenhausern Operationen und Prozeduren (OPS Version 2021)
  • [10] Evans JT, 2019, LANCET, V393, P647, DOI 10.1016/S0140-6736(18)31665-9