In Situ Laser Fenestrations of Aortic Endografts for Emergent Aortic Disease

被引:13
作者
Dean, Anastasia [1 ]
Wanhainen, Anders [1 ,2 ]
Mani, Kevin [1 ]
Kuzniar, Marek [1 ]
Grima, Matthew [3 ]
Zuccon, Gianmarco [1 ]
Lindstrom, David [1 ,4 ]
机构
[1] Uppsala Univ, Dept Surg Sci, Sect Vasc Surg, Uppsala, Sweden
[2] Umea Univ, Dept Surg & Perioperat Sci, Surg, Umea, Sweden
[3] Univ Malta, Fac Med & Surg, L Imsida, Malta
[4] Uppsala Acad Hosp, Dept Surg Sci, Sect Vasc Surg, S-75185 Uppsala, Sweden
关键词
STENT-GRAFT;
D O I
10.1016/j.avsg.2023.01.005
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: In situ laser fenestration (ISLF) is a novel endovascular technique which allows customization of a standard stent graft to a patient's anatomy. While most reported cases involve revascularization of the left subclavian artery (LSA), some centers have now reported their initial experience treating branches of the visceral aorta for aortic aneurysms. The aim of this study is to examine the adoption of ISLF in emergent aortic pathology at a specialized aortic center.Methods: Between December 2020 and February 2022, all patients who underwent ISLF as part of endovascular intervention for complex aortic pathology at a university hospital were iden-tified. Cases were collected from a prospective aortic database with additional information ob-tained from a retrospective review of electronic hospital records.Results: Fifteen patients (11 men and 4 women) underwent emergency ISLF, with a median age of 76 years. Eleven presented with symptomatic or ruptured aortic aneurysms, three with acute complicated aortic dissections and 1 aortic traumatic transection. Most aortic an-eurysms were thoraco-abdominal (n = 7), with 1 arch, 1 thoracic, 1 supra-renal, and one-juxta-renal aortic aneurysm. ISLF was performed to revascularize the LSA in 8 cases, and branches of the reno-visceral aorta in 7 cases. All LSA ISLF cases had left brachial ar -te ry exposure. Femoral access was percutaneous in 14 of 15 cases. Technical success was 96.3% (26/27)). Median ischemic times were: superior mesenteric artery 7 min, renal ar-teries 22 min, and celiac trunk 43.5 min. There were 2 early aortic/fenestration related rein-terventions. There was no stroke and 1 death caused by heparin-induced thrombocytopenia within 30 days. The majority of patients did not require intensive care admission (n = 8). The median intensive care unit stay was 0 days and hospital length of stay 18 days. There was no fenestration endoleak or reintervention post discharge with a median follow-up of 168 days.Conclusions: ISLF is a promising new technique that can show excellent technical results in experienced aortic centers, even during the learning curve. While custom-made devices with reinforced fenestrations are preferred in nonemergent situations, ISLF is a feasible option for complex aortic pathology in the acute setting when open surgery is not feasible.
引用
收藏
页码:329 / 337
页数:9
相关论文
共 15 条
  • [1] Impact of gap distance between fenestration and aortic wall on target artery instability following fenestrated-branched endovascular aortic repair
    Chait, Jesse
    Tenorio, Emanuel R.
    Mendes, Bernardo C.
    Lima, Guilherme B. Barbosa
    Marcondes, Giulianna B.
    Wong, Joshua
    Macedo, Thanila A.
    De Martino, Randall R.
    Oderich, Gustavo S.
    [J]. JOURNAL OF VASCULAR SURGERY, 2022, 76 (01) : 79 - +
  • [2] In Situ Laser Fenestration Technique: Bench-Testing of Aortic Endograft to Guide Clinical Practice
    Grima, Matthew Joe
    Wanhainen, Anders
    Lindstrom, David
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2024, 31 (01) : 126 - 131
  • [3] Characterization of particulates accompanying laser ablation of pressed polytetrafluorethylene (PTFE) targets
    Heitz, J
    Dickinson, JT
    [J]. APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 1999, 68 (05): : 515 - 523
  • [4] An Experimental Study of Laser in situ Fenestration of Current Aortic Endografts
    Jayet, J.
    Heim, F.
    Coggia, M.
    Chakfe, N.
    Coscas, R.
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2018, 56 (01) : 68 - 77
  • [5] Aneurysm Rupture and Mortality During the Waiting Time for a Customised Fenestrated/Branched Stent Graft in Complex Endovascular Aortic Repair
    Katsargyris, Athanasios
    Uthayakumar, Vasuki
    de Marino, Pablo Marques
    Botos, Balazs
    Verhoeven, Eric L.
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2020, 60 (01) : 44 - 48
  • [6] In situ laser fenestration for revascularization of aortic arch during treatment for iatrogenic type A aortic dissection
    Kong, Minjian
    Qian, Jianfang
    Duan, Qunjun
    Li, Xuebiao
    Dong, Aiqiang
    [J]. JOURNAL OF CARDIOTHORACIC SURGERY, 2019, 14 (1)
  • [7] In Situ Antegrade Laser Fenestrations During Endovascular Aortic Repair
    Le Houerou, Thomas
    Fabre, Dominique
    Alonso, Carlos G.
    Brenot, Philippe
    Bourkaib, Ryiad
    Angel, Claude
    Amsallem, Myriam
    Haulon, Stephan
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2018, 56 (03) : 356 - 362
  • [8] Image Fusion Guidance for In Situ Laser Fenestration of Aortic Stent graft for Endovascular Repair of Complex Aortic Aneurysm: Feasibility, Efficacy and Overall Functional Success
    Leger, Thomas
    Tacher, Vania
    Majewski, Marek
    Touma, Joseph
    Desgranges, Pascal
    Kobeiter, Hicham
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2019, 42 (10) : 1371 - 1379
  • [9] Laser Fenestration of Aortic Stent-Grafts Followed by Noncompliant vs Cutting Balloon Dilation: A Scanning Electron Microscopy Study
    Lin, Jing
    Parikh, Niraj
    Udgiri, Naval
    Wang, Shaoxia
    Miller, Daniel F.
    Li, Chaojing
    Panneton, Jean
    Nutley, Mark
    Zhang, Ze
    Huang, Yunfan
    Lu, Jun
    Zhang, Jingyi
    Wang, Lu
    Guidoin, Robert
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2018, 25 (03) : 397 - 407
  • [10] In situ stent-graft fenestration to preserve the left subdavian artery
    McWilliams, RG
    Murphy, M
    Hartley, D
    Lawrence-Brown, MMD
    Harris, PL
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2004, 11 (02) : 170 - 174