A Safe Framework for Quantitative In Vivo Human Evaluation of Image Guidance

被引:0
|
作者
Cannon, Piper C. [1 ]
Ferguson, James M. [1 ]
Pitt, E. Bryn [1 ]
Shrand, Jason A. [1 ]
Setia, Shaan A. [2 ]
Nimmagadda, Naren [2 ,3 ]
Barth, Eric J. [1 ]
Kavoussi, Nicholas L. [2 ]
Galloway, Robert L. [1 ]
Herrell, S. Duke [2 ]
Webster, Robert J. [1 ]
机构
[1] Vanderbilt Univ, Nashville, TN 37235 USA
[2] Vanderbilt Univ, Med Ctr, Nashville, TN 37232 USA
[3] Johns Hopkins Univ, Sch Med, Baltimore, MD 21287 USA
来源
IEEE OPEN JOURNAL OF ENGINEERING IN MEDICINE AND BIOLOGY | 2024年 / 5卷
基金
美国国家卫生研究院;
关键词
Surgery; Protocols; Robots; Phantoms; Kidney; In vivo; Three-dimensional displays; Image-guided surgery; partial nephrectomy; robotic surgery; LAPAROSCOPIC PARTIAL NEPHRECTOMY; SURGERY SYSTEM; TECHNOLOGY; ACCURACY; TRACKING; DISPLAY; MODEL;
D O I
10.1109/OJEMB.2023.3271853
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Goal: We present a new framework for in vivo image guidance evaluation and provide a case study on robotic partial nephrectomy. Methods: This framework (called the "bystander protocol") involves two surgeons, one who solely performs the therapeutic process without image guidance, and another who solely periodically collects data to evaluate image guidance. This isolates the evaluation from the therapy, so that in-development image guidance systems can be tested without risk of negatively impacting the standard of care. We provide a case study applying this protocol in clinical cases during robotic partial nephrectomy surgery. Results: The bystander protocol was performed successfully in 6 patient cases. We find average lesion centroid localization error with our IGS system to be 6.5 mm in vivo compared to our prior result of 3.0 mm in phantoms. Conclusions: The bystander protocol is a safe, effective method for testing in-development image guidance systems in human subjects.
引用
收藏
页码:133 / 139
页数:7
相关论文
共 50 条
  • [31] Quantitative Imaging In Vivo of Functioning Lymphatic Vessels Around Human Melanoma and Benign Nevi
    Akhras, Victoria
    Ramakrishnan, Rathi
    Stanton, Anthony W. B.
    Levick, John R.
    Cook, Martin G.
    Chong, Heung
    Mortimer, Peter S.
    MICROCIRCULATION, 2015, 22 (06) : 454 - 463
  • [32] A Smooth Velocity Transition Framework Based on Hierarchical Proximity Sensing for Safe Human-Robot Interaction
    Wang, Ruohan
    Li, Chen
    Lv, Honghao
    Pang, Gaoyang
    Wu, Haiteng
    Yang, Geng
    IEEE ROBOTICS AND AUTOMATION LETTERS, 2024, 9 (06): : 4910 - 4917
  • [33] Towards a Framework for the Design of Quantitative Experiments: Human-Computer Interaction and Accessibility Research
    Sandnes, Frode Eika
    Eika, Evelyn
    Medola, Fausto Orsi
    UNIVERSAL ACCESS IN HUMAN-COMPUTER INTERACTION: METHODS, TECHNOLOGIES, AND USERS, UAHCI 2018, PT I, 2018, 10907 : 107 - 120
  • [34] The In Vitro, In Vivo, and PBPK Evaluation of a Novel Lung-Targeted Cardiac-Safe Hydroxychloroquine Inhalation Aerogel
    Alsmadi, Mo'tasem M.
    Jaradat, Mays M.
    Obaidat, Rana M.
    Alnaief, Mohammad
    Tayyem, Rabab
    Idkaidek, Nasir
    AAPS PHARMSCITECH, 2023, 24 (06)
  • [35] In vivo evaluation of lateral lipid chain packing in human stratum corneum
    Kikuchi, S.
    Aosaki, T.
    Bito, K.
    Naito, S.
    Katayama, Y.
    SKIN RESEARCH AND TECHNOLOGY, 2015, 21 (01) : 76 - 83
  • [36] Quantitative Ex Vivo and In Vivo Comparison of Lumen Dimensions Measured by Optical Coherence Tomography and Intravascular Ultrasound in Human Coronary Arteries
    Gonzalo, Nieves
    Serruys, Patrick W.
    Garcia-Garcia, Hector M.
    van Soest, Gijs
    Okamura, Takayuki
    Ligthart, Jurgen
    Knaapen, Michiel
    Verheye, Stefan
    Bruining, Nico
    Regar, Evelyn
    REVISTA ESPANOLA DE CARDIOLOGIA, 2009, 62 (06): : 615 - 624
  • [37] Dosimetric and geometric evaluation of a hybrid strategy of offline adaptive planning and online image guidance for prostate cancer radiotherapy
    Liu, Han
    Wu, Qiuwen
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (15) : 5045 - 5062
  • [38] Force-Based Control for Safe Robot-Assisted Retinal Interventions: In Vivo Evaluation in Animal Studies
    Patel, Niravkumar
    Urias, Muller
    Ebrahimi, Ali
    Taylor, Russell H.
    Gehlbach, Peter
    Iordachita, Iulian
    IEEE TRANSACTIONS ON MEDICAL ROBOTICS AND BIONICS, 2022, 4 (03): : 578 - 587
  • [39] Image-derived arterial input function for quantitative fluorescence imaging of receptor-drug binding in vivo
    Elliott, Jonathan T.
    Samkoe, Kimberley S.
    Davis, Scott C.
    Gunn, Jason R.
    Paulsen, Keith D.
    Roberts, David W.
    Pogue, Brian W.
    JOURNAL OF BIOPHOTONICS, 2016, 9 (03) : 282 - 295
  • [40] A quantitative evaluation of deformable image registration based on MV cone beam CT images: Impact of deformation magnitudes and image modalities
    Huang, Yuliang
    Li, Chenguang
    Wang, Haiyang
    Hu, Qiaoqiao
    Wang, Ruoxi
    Chang, Cheng
    Ma, Wenjun
    Li, Weibo
    Wu, Hao
    Zhang, Yibao
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 71 : 82 - 87