Comparative experimental study of the biomechanical properties of retrograde tibial nailing and distal tibia plate in distal tibia fracture

被引:0
|
作者
Lin, Xuping [1 ,2 ]
Zhang, Cong [2 ]
Yang, Yanfang [3 ]
Yang, Wencheng [1 ]
Wang, Xiaomeng [1 ]
Lu, Haichuan [1 ]
Liu, Qingjun [2 ]
机构
[1] Fujian Med Univ, Dept Spinal Surg, Longyan Affiliated Hosp 1, Longyan, Peoples R China
[2] Xiamen Univ, Dept Orthoped Surg, Affiliated Dongnan Hosp, Zhangzhou, Peoples R China
[3] Fujian Med Univ, Shengli Clin Med Coll, Fuzhou, Peoples R China
关键词
distal tibia fracture; internal fixation; retrograde tibial nailing; distal tibia plate; biomechanics; FIXATION; OSTEOSYNTHESIS;
D O I
10.3389/fbioe.2024.1322043
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Objective: A biomechanical comparative analysis was conducted to evaluate the retrograde tibial nailing (RTN) and distal tibia plate techniques for the treatment of distal tibia fractures.Methods: Fourteen fresh adult tibia specimens were selected, consisting of seven males and seven females aged 34-55 years. The specimens were randomly divided into two groups (Group A and Group B) using a numerical table method, with seven specimens in each group. Group A underwent internal fixation of distal tibial fractures using RTN, while Group B received internal fixation using a plate. The axial compression properties of the specimens were tested in the neutral positions under pressures of 100, 200, 300, 400, and 500 N. Additionally, the torsional resistance of the two groups was assessed by subjecting the specimens to torques of 1.0, 2.0, 3.0, 4.0, and 5.0 N m.Results: At pressures of 400 and 500 N, the axial compression displacement in Group A (1.11 +/- 0.06, 1.24 +/- 0.05) mm was significantly smaller than that in Group B (1.21 +/- 0.08, 1.37 +/- 0.11) mm (p = 0.023, 0.019). Moreover, at a pressure of 500 N, the axial compression stiffness in Group A (389.24 +/- 17.79) N/mm was significantly higher than that of the control group (362.37 +/- 14.44) N/mm (p = 0.010). When subjected to torques of 4 and 5 N m, the torsion angle in Group A (2.97 degrees +/- 0.23 degrees, 3.41 degrees +/- 0.17 degrees) was significantly smaller compared to Group B (3.31 degrees +/- 0.28 degrees, 3.76 degrees +/- 0.20 degrees) (p = 0.035, 0.004). Furthermore, at a torque of 5 N m, the torsional stiffness in Group A (1.48 +/- 0.07) N m/degrees was significantly higher than that in Group B (1.36 +/- 0.06) N center dot m/degrees (p = 0.003).Conclusion: The results obtained from the study demonstrate that the biomechanical performance of RTN outperforms that of the distal tibial plate, providing valuable biomechanical data to support the clinical implementation of RTN.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Retrograde tibial nailing of far distal tibia fractures: a biomechanical evaluation of double- versus triple-distal interlocking
    Greenfield, Julia
    Appelmann, Philipp
    Wunderlich, Felix
    Mehler, Dorothea
    Rommens, Pol Maria
    Kuhn, Sebastian
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2022, 48 (05) : 3693 - 3700
  • [2] Biomechanical Comparative Analyses Between the Anterolateral and Medial Distal Tibia Locking Plates in Treating Complex Distal Tibial Fracture: A Finite Element Study
    Aizat, Raja Mohd
    Kadir, Mohammed Rafiq Abdul
    Ab Rahman, Shaifuzain
    Shihabudin, Tengku Muzaffar Tengku Md
    Robson, Noorzurani
    Kamarul, Tunku
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2013, 3 (04) : 532 - 537
  • [3] Biomechanical comparison of two locking plate systems for the distal tibia
    Hoegel, F.
    Hoffmann, S.
    Weninger, P.
    Buehren, V.
    Augat, P.
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2012, 38 (01) : 53 - 58
  • [4] Intramedullary nailing of the distal tibia illustrated with the Expert™ tibia nail
    Attal, R.
    Hansen, M.
    Rosenberger, R.
    Smekal, V.
    Rommens, P. M.
    Blauth, M.
    OPERATIVE ORTHOPADIE UND TRAUMATOLOGIE, 2011, 23 (05): : 397 - 410
  • [5] Comparative Study of the Management of Distal Tibia Fractures by Nailing Versus Plating
    Nath, Rohit
    Saxena, Saurabh
    Singh, Chetan
    Kumar, Chandan
    Singh, Santosh Kumar
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (11)
  • [6] The Retrograde Tibial Nail: Presentation and biomechanical evaluation of a new concept in the treatment of distal tibia fractures
    Kuhn, Sebastian
    Appelmann, Philipp
    Pairon, Philip
    Mehler, Dorothea
    Rommens, Pol M.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2014, 45 : S81 - S86
  • [7] Retrograde Tibial Nailing: a minimally invasive and biomechanically superior alternative to angle-stable plate osteosynthesis in distal tibia fractures
    Kuhn, Sebastian
    Appelmann, Philipp
    Mehler, Dorothea
    Pairon, Philip
    Rommens, Pol M.
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2014, 9
  • [8] Intramedullary nailing at different distal tibial fracture levels: A biomechanical study
    Guran, Ortac
    Ozmanevra, Ramadan
    Husemoglu, Resit Bugra
    Havitcioglu, Hasan
    Altinoz, Ozenc
    MEDICINE, 2024, 103 (22) : E38353
  • [9] Biomechanical comparison of two locking plate systems for the distal tibia
    F. Högel
    S. Hoffmann
    P. Weninger
    V. Bühren
    P. Augat
    European Journal of Trauma and Emergency Surgery, 2012, 38 : 53 - 58
  • [10] A clinical comparative study of intramedullary nailing and minimally invasive plate osteosynthesis for extra-articular distal tibia fractures
    Wang, Chaofeng
    Huang, Qiang
    Lu, Dongxing
    Wang, Qian
    Ma, Teng
    Zhang, Kun
    Li, Zhong
    AMERICAN JOURNAL OF TRANSLATIONAL RESEARCH, 2023, 15 (03): : 1996 - 2005