Development and Finite Element (FE) analysis of a novel short hip stem concept

被引:0
作者
Jahnke, Alexander [1 ,3 ]
Harz, Torben [1 ]
Ulloa, Carlos Alfonso Fonseca [1 ]
Ishaque, Bernd Alexander [1 ,2 ]
Rickert, Markus [1 ,2 ]
机构
[1] Justus Liebig Univ Giessen, Lab Biomech, Klinikstr 29, D-35392 Giessen, Germany
[2] Univ Hosp Giessen & Marburg UKGM, Dept Orthopaed & Orthopaed Surg, Klin Str 33, D-35392 Giessen, Germany
[3] Univ Med Ctr Giessen, Dept Orthopaed, Lab Biomech, Klin Str 29, D-35392 Giessen, Germany
关键词
Short hip stem; THA; Finite element analysis; Bone and prosthesis elasticity; ARTHROPLASTY;
D O I
10.1016/j.jor.2023.09.016
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: In order to improve the anchorage behavior of short hip stems, this development project aims at designing a short hip stem concept that preserves the femoral neck and minimizes interference with the physiological stress distribution of the femur. The new design will be evaluated according to ISO 7206-4 which is the standard for testing Implants for surgery.Methods: Basic CAD models based on an established short stem prosthesis were created and evaluated using finite element analysis. The best design was further developed to achieve a more deformable stem while maintaining stability. The model was validated through in vitro testing. Results: The "H-Beam" short stem showed a higher degree of deformation of approximately 142-144% compared to the established short stem. The FE model had a relative error of 0.98% and 1.07% compared to the in vitro tests. An operating procedure was outlined for this new short stem design.Discussion: The FE model is deemed valid due to small differences in comparison to in vitro testing. The short-stem prosthesis is more flexible and can be easily adapted to individual anatomy during surgery. The prosthesis length is similar to conventional prostheses, but the new stem design could allow better and faster osteointegration while preserving the cancellous bone structure.
引用
收藏
页码:117 / 123
页数:7
相关论文
共 19 条
[1]  
American Joint Replacement Registry, 2022, Annual Report. The Ninth Annual Reportof the AJRR on Hip and Knee Arthroplasty
[2]   Development of periprosthetic bone mass density around the cementless Metha® short hip stem during three year follow up-a prospective radiological and clinical study [J].
Augustin, Laura ;
Boller, Sarah ;
Bobach, Cathrein ;
Jahnke, Alexander ;
Ahmed, Gafar Adam ;
Rickert, Markus ;
Ishaque, Bernd Alexander .
INTERNATIONAL ORTHOPAEDICS, 2019, 43 (09) :2031-2037
[3]   Standardized Loads Acting in Hip Implants [J].
Bergmann, Georg ;
Bender, Alwina ;
Dymke, Joern ;
Duda, Georg ;
Damm, Philipp .
PLOS ONE, 2016, 11 (05)
[4]  
Ducheyne P, 2011, Comprehensive Biomaterials, P305
[5]  
Endoprothesenregister Deutschland Jahresbericht, 2022, Mit Sicherheit Mehr Qualitat
[6]   Short Stem for Total Hip Arthroplasty (THA) - Overview, Patient Selection and Perspectives by Using the Metha® Hip Stem System [J].
Ishaque, Bernd Alexander .
ORTHOPEDIC RESEARCH AND REVIEWS, 2022, 14 :77-89
[7]  
ISO, ISO 7206-4:2010-06
[8]   Comparison of Short Stems Versus Straight Hip Stems: A Biomechanical Analysis of the Primary Torsional Stability [J].
Jahnke, Alexander ;
Ghandourah, Suleiman ;
Ulloa, Carlos A. Fonseca ;
Seeger, Jorn Bengt ;
Rickert, Markus ;
Ishaque, Bernd Alexander ;
Ahmed, Gafar Adam .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2020, 142 (12)
[9]   Outcome of short- to medium-term migration analysis of a cementless short stem total hip arthroplasty using EBRA-FCA: a radiological and clinical study [J].
Jahnke, Alexander ;
Wiesmair, Ann-Kathrin ;
Fonseca Ulloa, Carlos Alfonso ;
Ahmed, Gafar Adam ;
Rickert, Markus ;
Ishaque, Bernd Alexander .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2020, 140 (02) :247-253
[10]   Analysis of the elastic bending characteristics of cementless short hip stems considering the valgus alignment of the prosthetic stem [J].
Jahnke, Alexander ;
Ulloa, Carlos A. Fonseca ;
Seeger, Joern Bengt ;
Rickert, Markus ;
Jahnke, Gerhard Walter ;
Ahmed, Gafar Adam ;
Ishaque, Bernd Alexander .
CLINICAL BIOMECHANICS, 2018, 52 :49-56