Impedance-Readout Integrated Circuits for Electrical Impedance Spectroscopy: Methodological Review

被引:5
|
作者
Cheon, Song-, I [1 ]
Choi, Haidam [1 ]
Kang, Hyoju [2 ]
Suh, Ji-Hoon [1 ]
Park, Seonghyun [1 ]
Kweon, Soon-Jae [2 ]
Je, Minkyu [1 ]
Ha, Sohmyung [3 ,4 ]
机构
[1] Korea Adv Inst Sci & Technol KAIST, Sch Elect Engn, Daejeon 34141, South Korea
[2] Catholic Univ Korea, Sch Informat Commun & Elect Engn, Bucheon 14662, South Korea
[3] New York Univ Abu Dhabi, Div Engn, Abu Dhabi 129188, U Arab Emirates
[4] NYU, Tandon Sch Engn, New York, NY 10003 USA
关键词
Integrated circuits; Impedance; Frequency measurement; Impedance measurement; Throughput; Electrodes; Demodulation; Accuracy; bio-impedance; demodulation; dynamic range; electrode; frequency range; impedance; measurement speed; noise; power consumption; spectroscopy; throughput; BIOIMPEDANCE MEASUREMENT; TOMOGRAPHY SOC; ANALYZER; BODY; CANCELLATION; ELECTRODE; ACCURACY; SENSORS; SYSTEM; CHIP;
D O I
10.1109/TBCAS.2023.3319212
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This review article provides a comprehensive overview of impedance-readout integrated circuits (ICs) for electrical impedance spectroscopy (EIS) applications. The readout IC, a crucial component of on-chip EIS systems, significantly affects key performance metrics of the entire system, such as frequency range, power consumption, accuracy, detection range, and throughput. With the growing demand for portable, wearable, and implantable EIS systems in the Internet-of-Things (IoT) era, achieving high energy efficiency while maintaining a wide frequency range, high accuracy, wide dynamic range, and high throughput has become a focus of research. Furthermore, to enhance the miniaturization and convenience of EIS systems, many emerging systems utilize two-electrode or dry electrode configurations instead of the conventional four-electrode configuration with wet electrodes for impedance measurement. In response to these trends, various technologies have been developed to ensure reliable operations even at two- or dry-electrode interfaces. This article reviews the principles, advantages, and disadvantages of techniques employed in state-of-the-art impedance-readout ICs, aiming to achieve high energy efficiency, wide frequency range, high accuracy, wide dynamic range, low noise, high throughput, and/or high input impedance. The thorough review of these advancements will provide valuable insights into the future development of impedance-readout ICs and systems for IoT and biomedical applications.
引用
收藏
页码:215 / 232
页数:18
相关论文
共 50 条
  • [21] Electrical impedance spectroscopy of single cells in hydrodynamic traps
    El Hasni, Akram
    Schmitz, Carlo
    Bui-Goebbels, Katrin
    Braeunig, Peter
    Jahnen-Dechent, Wilhelm
    Schnakenberg, Uwe
    SENSORS AND ACTUATORS B-CHEMICAL, 2017, 248 : 419 - 429
  • [22] A 0.13 μm CMOS Integrated Circuit for Electrical Impedance Spectroscopy from 1 kHz to 10 GHz
    Garcia-Ramirez, Ronny
    Chacon-Rodriguez, Alfonso
    Rimolo-Donadio, Renato
    2017 30TH IEEE INTERNATIONAL SYSTEM-ON-CHIP CONFERENCE (SOCC), 2017, : 126 - 131
  • [23] Textile-Integrated Organic Electrochemical Transistor for Selective Ion Detection via Electrical Impedance Spectroscopy
    Altana, Antonio
    Shkodra, Bajramshahe
    Ibba, Pietro
    Angeli, Martina Aurora Costa
    Ploner, Moritz
    Petrelli, Mattia
    Korek, Eva-Maria
    Lugli, Paolo
    Petti, Luisa
    IEEE SENSORS LETTERS, 2024, 8 (07)
  • [24] Electrical Impedance Sensors for Multi-Phase Flow Measurement: A Review
    Cui, Ziqiang
    Zhang, Qian
    Gao, Kai
    Xia, Zihan
    Wang, Huaxiang
    IEEE SENSORS JOURNAL, 2021, 21 (24) : 27252 - 27267
  • [25] Virtual Biopsy by Electrical Impedance Spectroscopy in Barrett's Carcinoma
    Bloesser, Sandra
    May, Andrea
    Welsch, Lukas
    Ast, Michael
    Braun, Susanne
    Velten, Thomas
    Biehl, Margit
    Tschammer, Jonas
    Roeb, Elke
    Knabe, Mate
    JOURNAL OF GASTROINTESTINAL CANCER, 2022, 53 (04) : 948 - 957
  • [26] Electrical Impedance Spectroscopy Investigation on Cucumber Dehydration
    Liu, Xing
    Fang, Qiang
    Zheng, Sicong
    Cosic, I.
    Cao, Peng
    EUROPE-ASIA SYMPOSIUM ON QUALITY MANAGEMENT IN POSTHARVEST SYSTEMS - EURASIA 2007, 2008, 804 : 637 - 644
  • [27] Electrical Impedance Spectroscopy Investigation on Banana Ripening
    Liu, Xing
    Fang, Qiang
    Zheng, Sicong
    Cosic, I.
    EUROPE-ASIA SYMPOSIUM ON QUALITY MANAGEMENT IN POSTHARVEST SYSTEMS - EURASIA 2007, 2008, 804 : 159 - 165
  • [28] Applications of Electrical Impedance Tomography (EIT): A Short Review
    Bera, Tushar Kanti
    3RD INTERNATIONAL CONFERENCE ON COMMUNICATION SYSTEMS (ICCS-2017), 2018, 331
  • [29] An experimental study of Electrical Impedance Spectroscopy analysis of conductive liquids
    Vazquez Nambo, Manuel
    Gutierrez Gnecchi, Jose Antonio
    Reyes Archundia, Enrique
    Tellez Anguiano, Adriana del Carmen
    Pintor Lazaro, Luis I.
    Rodriguez Ruiz, Julieta G.
    Lorias Espinoza, Daniel
    2018 IEEE INTERNATIONAL AUTUMN MEETING ON POWER, ELECTRONICS AND COMPUTING (ROPEC), 2018,
  • [30] Code-Division-Multiplexed Electrical Impedance Tomography Spectroscopy
    McEwan, Alistair
    Tapson, Jonathan
    van Schaik, Andre
    Holder, David S.
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2009, 3 (05) : 332 - 338