Estimating blurless and noise-free Ir-192 source images from gamma camera images for high-dose-rate brachytherapy using a deep-learning approach

被引:3
|
作者
Nakanishi, Kohei [1 ]
Yamamoto, Seiichi [2 ]
Yabe, Takuya [3 ]
Yogo, Katsunori [1 ]
Noguchi, Yumiko [4 ]
Okudaira, Kuniyasu [4 ]
Kawachi, Naoki [3 ]
Kataoka, Jun [2 ]
机构
[1] Nagoya Univ, Grad Sch Med, Dept Integrated Hlth Sci, Nagoya, Japan
[2] Waseda Univ, Fac Sci & Engn, Tokyo, Japan
[3] Natl Inst Quantum Sci & Technol QST, Takasaki Inst Adv Quantum Sci, Fdn Quantum Technol Res Directorate, Chiba Chi, Japan
[4] Nagoya Univ Hosp, Dept Radiol Technol, Nagoya, Japan
关键词
HDR brachytherapy; Ir-192; source; deep learning; gamma camera; REAL-TIME VERIFICATION; RADIATION-THERAPY; EMISSION; POSITION;
D O I
10.1088/2057-1976/ad0bb2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective. Precise monitoring of the position and dwell time of iridium-192 (Ir-192) during high-dose-rate (HDR) brachytherapy is crucial to avoid serious damage to normal tissues. Source imaging using a compact gamma camera is a potential approach for monitoring. However, images from the gamma camera are affected by blurring and statistical noise, which impact the accuracy of source position monitoring. This study aimed to develop a deep-learning approach for estimating ideal source images that reduce the effect of blurring and statistical noise from experimental images captured using a compact gamma camera. Approach. A double pix2pix model was trained using the simulated gamma camera images of an Ir-192 source. The first model was responsible for denoising the Ir-192 images, whereas the second model performed super resolution. Trained models were then applied to the experimental images to estimate the ideal images. Main results. At a distance of 100 mm between the compact gamma camera and the Ir-192 source, the difference in full width at half maximum (FWHM) between the estimated and actual source sizes was approximately 0.5 mm for a measurement time of 1.5 s. This difference has been improved from approximately 2.7 mm without the use of DL. Even with a measurement time of 0.1 s, the ideal images could be estimated as accurately as in the 1.5 s measurements. This method consistently achieved accurate estimations of the source images at any position within the field of view; however, the difference increased with the distance between the Ir-192 source and the compact gamma camera. Significance. The proposed method successfully provided estimated images from the experimental images within errors smaller than 0.5 mm at 100 mm. This method is promising for reducing blurring and statistical noise from the experimental images, enabling precise real-time monitoring of Ir-192 sources during HDR brachytherapy.
引用
收藏
页数:9
相关论文
共 7 条
  • [1] Imaging of Ir-192 source using a high energy gamma camera for high-dose-rate brachytherapy
    Yamamoto, Seiichi
    Yogo, Katsunori
    Noguchi, Yumiko
    Nakaya, Takayoshi
    Okudaira, Kuniyasu
    RADIATION MEASUREMENTS, 2019, 126
  • [2] Development of a simultaneous imaging system to measure the optical and gamma ray images of Ir-192 source for high-dose-rate brachytherapy
    Nagata, J.
    Yamamoto, S.
    Noguchi, Y.
    Nakaya, T.
    Okudaira, K.
    Kamada, K.
    Yoshikawa, A.
    JOURNAL OF INSTRUMENTATION, 2021, 16 (12):
  • [3] Optimization of the energy window setting in Ir-192 source imaging for high-dose-rate brachytherapy using a YAP(Ce) gamma camera
    Nagata, Jura
    Nakanishi, Kohei
    Yamamoto, Seiichi
    Yabe, Takuya
    Yogo, Katsunori
    Noguchi, Yumiko
    Okudaira, Kuniyasu
    Kamada, Kei
    Yoshikawa, Akira
    Kataoka, Jun
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 103 : 66 - 73
  • [4] Trials of transmission imaging using clinically used Ir-192 source for high-dose-rate brachytherapy
    Nagata, J.
    Yamamoto, S.
    Nakanishi, K.
    Noguchi, Y.
    Okudaira, K.
    Kataoka, J.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (06)
  • [5] Automatic segmentation of magnetic resonance images for high-dose-rate cervical cancer brachytherapy using deep learning
    Yoganathan, S. A.
    Paul, Siji Nojin
    Paloor, Satheesh
    Torfeh, Tarraf
    Chandramouli, Suparna Halsnad
    Hammoud, Rabih
    Al-Hammadi, Noora
    MEDICAL PHYSICS, 2022, 49 (03) : 1571 - 1584
  • [6] Approaching automated applicator digitization from a new angle: Using sagittal images to improve deep learning accuracy and robustness in high-dose-rate prostate brachytherapy
    Weishaupt, Luca Leon
    Sayed, Hisham Kamal
    Mao, Ximeng
    Choo, Richard
    Stish, Bradley J.
    Enger, Shirin A.
    Deufel, Christopher
    BRACHYTHERAPY, 2022, 21 (04) : 520 - 531
  • [7] Development of a low-sensitivity high resolution YAP(Ce) scintillation camera system toward the real-time imaging of an 192Ir source during high-dose-rate brachytherapy
    Nagata, J.
    Yamamoto, S.
    Noguchi, Y.
    Nakaya, T.
    Okudaira, K.
    Kamada, K.
    Yoshikawa, A.
    JOURNAL OF INSTRUMENTATION, 2020, 15 (12):