Sphygmomanometer Dynamic Pressure Measurement Using a Condenser Microphone

被引:0
作者
Tomazini, Zan [1 ]
Gersak, Gregor [1 ]
Begus, Samo [1 ]
机构
[1] Univ Ljubljana, Fac Elect Engn, Ljubljana 1000, Slovenia
关键词
RF microphone; blood pressure; low frequency; oscillometric; BP simulator; CALIBRATION; HYPERTENSION;
D O I
10.3390/s23198340
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
There is a worldwide need to improve blood pressure (BP) measurement error in order to correctly diagnose hypertension. Cardiovascular diseases cause 17.9 million deaths annually and are a substantial monetary strain on healthcare. The current measurement uncertainty of 3 mmHg should be improved upon. Dynamic pressure measurement standards are lacking or non-existing. In this study we propose a novel method of measuring air pressure inside the sphygmomanometer tubing during BP measurement using a condenser microphone. We designed, built, and tested a system that uses a radiofrequency (RF) modulation method to convert changes in capacitance of a condenser microphone into pressure signals. We tested the RF microphone with a low-frequency (LF) sound source, BP simulator and using a piezoresistive pressure sensor as a reference. Necessary tests were conducted to assess the uncertainty budget of the system. The RF microphone prototype has a working frequency range from 0.5 Hz to 280 Hz in the pressure range from 0 to 300 mmHg. The total expanded uncertainty (k = 2, p = 95.5%) of the RF microphone was 4.32 mmHg. The proposed method could establish traceability of BP measuring devices to acoustic standards described in IEC 61094-2 and could also be used in forming dynamic BP standards.
引用
收藏
页数:15
相关论文
共 40 条
  • [21] Mellert V., 2004, P JOINT C CFA DAGA 0
  • [22] Primary calibration of measurement microphones in the world: state of art
    Milhomem, T. A. B.
    Soares, Z. M. D.
    [J]. 8TH BRAZILIAN CONGRESS ON METROLOGY (METROLOGIA 2015), 2016, 733
  • [23] Nilsson M., 2016, Dynamic Calibration of Pressure Sensors
  • [24] Ori Z, 1992, Cardiol Clin, V10, P499
  • [25] Low Frequency Calibration of Measurement Microphones
    Rasmussen, Gunnar
    Nielson, Kim M.
    [J]. JOURNAL OF LOW FREQUENCY NOISE VIBRATION AND ACTIVE CONTROL, 2009, 28 (03): : 223 - 228
  • [26] Range of normal blood pressure - A statistical and clinical study of 11,383 persons
    Robinson, SC
    Brucer, M
    [J]. ARCHIVES OF INTERNAL MEDICINE, 1939, 62 (03) : 409 - 444
  • [27] Evaluating the impact of measurement uncertainty in blood pressure measurement on hypertension diagnosis
    Rodrigues Filho, Bruno A.
    Farias, Rafael F.
    dos Anjos, Wilian
    [J]. BLOOD PRESSURE MONITORING, 2018, 23 (03) : 141 - 147
  • [28] The extent and implications of sphygmomanometer calibration error in primary care
    Rouse, A
    Marshall, T
    [J]. JOURNAL OF HUMAN HYPERTENSION, 2001, 15 (09) : 587 - 591
  • [29] Heart rate variability
    Stauss, HM
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY-REGULATORY INTEGRATIVE AND COMPARATIVE PHYSIOLOGY, 2003, 285 (05) : R927 - R931
  • [30] Turner MJ, 2007, AUST FAM PHYSICIAN, V36, P834