Sphygmomanometer Dynamic Pressure Measurement Using a Condenser Microphone

被引:0
作者
Tomazini, Zan [1 ]
Gersak, Gregor [1 ]
Begus, Samo [1 ]
机构
[1] Univ Ljubljana, Fac Elect Engn, Ljubljana 1000, Slovenia
关键词
RF microphone; blood pressure; low frequency; oscillometric; BP simulator; CALIBRATION; HYPERTENSION;
D O I
10.3390/s23198340
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
There is a worldwide need to improve blood pressure (BP) measurement error in order to correctly diagnose hypertension. Cardiovascular diseases cause 17.9 million deaths annually and are a substantial monetary strain on healthcare. The current measurement uncertainty of 3 mmHg should be improved upon. Dynamic pressure measurement standards are lacking or non-existing. In this study we propose a novel method of measuring air pressure inside the sphygmomanometer tubing during BP measurement using a condenser microphone. We designed, built, and tested a system that uses a radiofrequency (RF) modulation method to convert changes in capacitance of a condenser microphone into pressure signals. We tested the RF microphone with a low-frequency (LF) sound source, BP simulator and using a piezoresistive pressure sensor as a reference. Necessary tests were conducted to assess the uncertainty budget of the system. The RF microphone prototype has a working frequency range from 0.5 Hz to 280 Hz in the pressure range from 0 to 300 mmHg. The total expanded uncertainty (k = 2, p = 95.5%) of the RF microphone was 4.32 mmHg. The proposed method could establish traceability of BP measuring devices to acoustic standards described in IEC 61094-2 and could also be used in forming dynamic BP standards.
引用
收藏
页数:15
相关论文
共 40 条
  • [1] [Anonymous], 2022, IEC 61094-2:2009+AMD1:2022
  • [2] [Anonymous], 2004, Paper 6130
  • [3] [Anonymous], 2007, ISO 81060-1:2007
  • [4] [Anonymous], 2018, IEC 80601-2-30:2018
  • [5] DEVELOPMENT OF A PRIMARY STANDARD FOR THE MEASUREMENT OF DYNAMIC PRESSURE AND TEMPERATURE
    BEAN, VE
    BOWERS, WJ
    HURST, WS
    ROSASCO, GJ
    [J]. METROLOGIA, 1994, 30 (06) : 747 - 750
  • [6] Bruel&Kjaer Technical Documentation, 2019, Microphone Handbook, V1
  • [7] Modeling and simulation of a condenser microphone
    Chen, Jen-Yi
    Hsu, Yu-Chun
    Lee, Shu-Sheng
    Mukherjee, Tamal
    Fedder, Gary K.
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 2008, 145 : 224 - 230
  • [8] Development of Real-Time Cuffless Blood Pressure Measurement Systems with ECG Electrodes and a Microphone Using Pulse Transit Time (PTT)
    Choi, Jingyu
    Kang, Younghwan
    Park, Jaesoon
    Joung, Yeunho
    Koo, Chiwan
    [J]. SENSORS, 2023, 23 (03)
  • [9] Feng Niu, 2012, 2012 IEEE International Conference on Mechatronics and Automation (ICMA), P687, DOI 10.1109/ICMA.2012.6283225
  • [10] Frederiksen E., 2013, Int J Metrol Qual Eng, V4, P97, DOI DOI 10.1051/IJMQE/2013045