Clinical use conditions of lead deployment and simulated lead fracture rate in left bundle branch area pacing

被引:11
作者
Zou, Jiangang [1 ]
Chen, Keping [2 ]
Liu, Xingbin [3 ]
Xu, Yuanning [3 ]
Jiang, Lingyun [3 ]
Dai, Yan [2 ]
Lin, Jinxuan [2 ]
Hou, Xiaofeng [1 ]
Qiu, Yuanhao [1 ]
Himes, Adam [4 ]
Lucas, Ryan [4 ]
Demmer, Wade [4 ]
Mara, Nicki [4 ]
Zhou, Xiaohong [4 ]
Lu, Hongyang [5 ]
机构
[1] Nanjing Med Univ, Affiliated Hosp 1, Dept Cardiol, Nanjing, Peoples R China
[2] Chinese Acad Med Sci, Fuwai Hosp, Peking Union Med Coll, Arrhythmia Ctr, Beijing, Peoples R China
[3] Sichuan Univ, West China Hosp, Dept Cardiol, Chengdu, Peoples R China
[4] Medtronic Plc, Cardiac Rhythm Management, Mounds View, MN USA
[5] Medtron Shanghai Ltd, Medtron Technol Ctr, Cardiac Rhythm Management, Shanghai, Peoples R China
关键词
3830 SelectSecure pacing lead; cardiac pacing; left bundle branch area pacing; mechanical use condition; LUMENLESS; EXPERIENCE; BIPOLAR;
D O I
10.1111/jce.15843
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionLeft bundle branch area pacing (LBBAP) is achieved by advancing the lead tip deep in the septum. Most LBBAP implants are performed using the Medtronic SelectSecure (TM) MRI SecureScan (TM) Model 3830 featuring a unique 4 Fr fixed helix lumenless design. Details of lead use conditions and long-term reliability have not been reported. This study was designed to quantify the mechanical use conditions for the 3830 lead during and after LBBAP implant, and to evaluate reliability using bench testing and simulation. MethodsFifty bradycardia patients with implantation of the 3830 lead for LBBAP were enrolled. Use conditions of lead deployment at implantation were collected and computed tomography (CT) scans were performed at 3-month follow-up. Curvature amplitude along the pacing lead was determined with CT images. Fatigue bending was performed using accelerated testing in a more severe environment than routine clinical use conditions. Conductor fracture rate in a simulated patient population was estimated based on clinical use conditions and fatigue test results. ResultsThe number of attempts to place the 3830 lead for LBBAP was 2.1 +/- 1.3 (range: 1-7) with 13 +/- 6 lead rotations at the final attempt. Extreme implant conditions were simulated in bench testing with 5 applications of 20 turns followed by up to 400 million bending cycles. Reliability modeling predicted a 10-year fracture rate of 0.02%. ConclusionsLBBAP implants require more lead rotations than standard pacing implants and result in unique lead bending. Application of simulated LBBAP use conditions to the 3830 lead in an accelerated in-vitro model does not produce excess conductor fractures.IMAGE-LBBP Study ID of ClinicalTrial.GOV: NCT04119323.
引用
收藏
页码:718 / 725
页数:8
相关论文
共 50 条
  • [41] Clinical Outcomes of Permanent Left Bundle Branch Area Pacing in Patients With Left Bundle Branch Block and Left Ventricular Ejection Fraction >35 vs. ≤35%
    Jiang, Zhixin
    Wu, Tian
    Wu, Yixian
    Chen, Zenghong
    Yang, Wen
    Chen, Chongchong
    Zhou, Xiujuan
    Shan, Qijun
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2022, 9
  • [42] Preliminary study on left bundle branch area pacing in children: Clinical observation of 12 cases
    Dai Wenlong
    Guo Baojing
    Dai Chencheng
    Dong Jianzeng
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (07) : 1558 - 1566
  • [43] Predictors of functional mitral regurgitation improvement in patients with left bundle branch block treated with left bundle branch area pacing
    Miao, Huazhong
    Chen, Zenghong
    Luo, Yushi
    Cheng, Yandi
    Cooray, P. L. R. Krisunika
    Wu, Tian
    Yang, Wen
    Zhou, Xiujuan
    Shan, Qijun
    Jiang, Zhixin
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2024, : 1412 - 1417
  • [44] Feasibility, safety, and short-term follow-up of defibrillator lead at left bundle branch area pacing location: A pilot study
    Ponnusamy, Shunmuga Sundaram
    Ganesan, Vithiya
    Vijayaraman, Pugazhendhi
    HEART RHYTHM, 2024, 21 (10) : 1900 - 1906
  • [45] Practical and systematic approach using the steerable catheter and stylet-driven lead to deliver safe and effective left bundle branch area pacing
    Foo, David
    Tan, Kenny
    Ng, Shonda
    Tan, Li W.
    Chia, Pow-Li
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2023, 34 (09) : 1944 - 1950
  • [46] Comparison of clinical and echocardiographic outcomes between left bundle branch area pacing and right ventricular pacing in older patients
    Wang, Qian
    He, Chen
    Fan, Xiaohan
    Zhu, Haojie
    Li, Xiaofei
    Liu, Zhimin
    Yao, Yan
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2024, : 1346 - 1357
  • [47] Incidence of pacing-induced cardiomyopathy: left bundle branch area pacing versus leadless pacing
    Chopra, Maya
    Hsieh, Ji-Cheng
    Mueller, William
    Braunstein, Eric D.
    Beldner, Stuart
    Mitra, Raman L.
    Epstein, Laurence M.
    Willner, Jonathan
    Gabriels, James K.
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2025,
  • [48] Clinical Outcomes in Patients With Left Bundle Branch Area Pacing vs. Right Ventricular Pacing for Atrioventricular Block
    Li, Xiaofei
    Zhang, Junmeng
    Qiu, Chunguang
    Wang, Zhao
    Li, Hui
    Pang, Kunjing
    Yao, Yan
    Liu, Zhimin
    Xie, Ruiqin
    Chen, Yangxin
    Wu, Yongquan
    Fan, Xiaohan
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2021, 8
  • [49] Long-term outcomes of left bundle branch area pacing versus biventricular pacing in patients with heart failure and complete left bundle branch block
    Juan Hua
    Yang Chen
    Jianhua Yu
    Qinmei Xiong
    Zhen Xia
    Zirong Xia
    Qianghui Huang
    Qiling Kong
    Huolong Chen
    Yichu Zhang
    Jianxin Hu
    Juxiang Li
    Jinzhu Hu
    Qi Chen
    Kui Hong
    Heart and Vessels, 2022, 37 : 1162 - 1174
  • [50] Long-term outcomes of left bundle branch area pacing versus biventricular pacing in patients with heart failure and complete left bundle branch block
    Hua, Juan
    Chen, Yang
    Yu, Jianhua
    Xiong, Qinmei
    Xia, Zhen
    Xia, Zirong
    Huang, Qianghui
    Kong, Qiling
    Chen, Huolong
    Zhang, Yichu
    Hu, Jianxin
    Li, Juxiang
    Hu, Jinzhu
    Chen, Qi
    Hong, Kui
    HEART AND VESSELS, 2022, 37 (07) : 1162 - 1174