Marginal Fit Evaluation of Zirconia Substructure Computer-Aided Design and Manufacturing (CAD/CAM) by Scanning Electron Microscope

被引:0
作者
Abu-Ras, Kian [1 ]
Dolev, Eran [1 ]
Biadsee, Ameer [1 ]
Ormianer, Zeev [1 ]
机构
[1] Tel Aviv Univ, Sch Dent Med, Dept Oral Rehabil, POB 3904, Tel Aviv, Israel
来源
APPLIED SCIENCES-BASEL | 2023年 / 13卷 / 19期
关键词
implants; MD; AMD; Ti base; zirconia substructure; CAD/CAM; SEM; FIXED DENTAL PROSTHESES; TITANIUM-BASE ABUTMENTS; ALL-CERAMIC CROWNS; IMPLANT ABUTMENTS; ADAPTATION; ACCURACY; METAL; DISCREPANCY; TECHNOLOGY; SYSTEMS;
D O I
10.3390/app131910984
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Purpose. This in vitro study compared the vertical marginal discrepancy (MD) and absolute marginal discrepancy (AMD) between the zirconia substructure and titanium (Ti)-base abutment before and after cementation. Material and methods. An intraoral scanner (CEREC Omnicam scanner; Dentsply Sirona) was used to scan the dental model of missing tooth 36 and obtain a three-dimensional virtual model. It was designed and milled for 25 zirconia superstructures (inCoris ZI meso F2 L, Dentsply Sirona). The samples were cemented to Ti bases. The MD and AMD were measured for each sample by using a scanning electron microscope (SEM) before and after cementation. Results. The overall mean values of MD measurements before and after cementation were 14.12 mu m and 20.28 mu m, respectively. The overall mean values of AMD measurements prior to and after substructure cementation on the Ti base were 54.85 mu m and 116.0 mu m, respectively. ANOVA revealed significant differences in MD (p = 0.001) and AMD (p < 0.001) before and after cementation. Conclusions. Significant differences were observed in MD and AMD values before and after cementation. The MD and AMD values were significantly lower before cementation than after cementation. The mean MD and AMD values were clinically acceptable.
引用
收藏
页数:14
相关论文
共 38 条
[1]   Titanium-base abutments may have similar long-term peri-implant effects as non-bonded one-piece abutments [J].
Afrashtehfar, Kelvin I. ;
Weber, Adrian ;
Abou-Ayash, Samir .
EVIDENCE-BASED DENTISTRY, 2022, 23 (04) :134-135
[2]  
Akbar Jaber Hussain, 2006, J Prosthodont, V15, P155, DOI 10.1111/j.1532-849X.2006.00095.x
[3]   Single-tooth replacement: factors affecting different prosthetic treatment modalities [J].
Al-Quran, Firas A. ;
Al-Ghalayini, Raed F. ;
Al-Zu'bi, Bashar N. .
BMC ORAL HEALTH, 2011, 11
[4]   Titanium Base Abutments in Implant Prosthodontics: A Literature Review [J].
Al-Thobity, Ahmad M. .
EUROPEAN JOURNAL OF DENTISTRY, 2022, 16 (01) :49-55
[5]   Advancements in CAD/CAM technology: Options for practical implementation [J].
Alghazzawi, Tariq F. .
JOURNAL OF PROSTHODONTIC RESEARCH, 2016, 60 (02) :72-84
[6]   MARGINAL ADAPTATION OF THREE DIFFERENT ZIRCONIUM DIOXIDE THREE-UNIT FIXED DENTAL PROSTHESES [J].
Att, Wael ;
Komine, Futoshii ;
Gerds, Thomas ;
Strub, Joerg Rudolf .
JOURNAL OF PROSTHETIC DENTISTRY, 2009, 101 (04) :239-247
[7]   EVALUATION OF THE MARGINAL FIT OF A ZIRCONIA CERAMIC COMPUTER-AIDED MACHINED (CAM) CROWN SYSTEM [J].
Baig, Mirza Rustum ;
Tan, Keson Beng-Choon ;
Nicholls, Jack I. .
JOURNAL OF PROSTHETIC DENTISTRY, 2010, 104 (04) :216-227
[8]   Comparison of marginal and internal fit of press-on-metal and conventional ceramic systems for three- and four-unit implant-supported partial fixed dental prostheses: An in vitro study [J].
Bayramoglu, Emre ;
Ozkan, Yasemin Kulak ;
Yildiz, Coskun .
JOURNAL OF PROSTHETIC DENTISTRY, 2015, 114 (01) :52-58
[9]   Marginal accuracy of combined tooth-implant-supported fixed dental prostheses after in vitro stress simulation [J].
Boeckler, Arne F. ;
Morton, Dean ;
Kraemer, Stephan ;
Geiss-Gerstdorfer, Juergen ;
Setz, Juergen M. .
CLINICAL ORAL IMPLANTS RESEARCH, 2008, 19 (12) :1261-1269
[10]   In Vitro Marginal Fit of Three All-Ceramic Crown Systems Before and After Cementation [J].
Borges, G. A. ;
Faria, J. S. ;
Agarwal, P. ;
Spohr, A. M. ;
Correr-Sobrinho, L. ;
Miranzi, B. A. S. .
OPERATIVE DENTISTRY, 2012, 37 (06) :641-649