Evaluation of the analytical and clinical performance of a new high-sensitivity cardiac troponin I assay: hs-cTnI (CLIA) assay

被引:8
|
作者
Li, Ling [1 ]
Shu, Xin [2 ]
Zhang, Litao [2 ]
Xu, Ao [1 ]
Yang, Juan [1 ]
Jing, Yisha [2 ]
Wang, Hui [2 ]
Zhang, Zhenlu [1 ]
机构
[1] Wuhan Asia Heart Hosp, Clin Lab, Wuhan 430022, Hubei, Peoples R China
[2] Wuhan Asia Gen Hosp, Clin Lab, Wuhan, Chin, Myanmar
关键词
high-sensitivity; cardiac troponin I; chemiluminescence immunoassay; myocardial infarction; analytical performance; diagnostic performance; ACUTE MYOCARDIAL-INFARCTION; STANDARDIZATION; AUTOANTIBODIES; GUIDELINES;
D O I
10.1515/cclm-2023-0529
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Cardiac troponin (cTn) is the key biomarker for diagnosis of acute coronary syndrome (ACS). We performed a complete assessment of the high-sensitivity cardiac troponin I (hs-cTnI) (CLIA) assay on the analytical performance and clinical diagnostic performance, which was compared with Abbott ARCHITECT hs-cTnI assay.Methods: Sex-specific 99th percentile upper reference limits (URLs) were determined from a healthy population of 424 males and 408 females. High-sensitivity performance was assessed by examining the imprecision at sex-specific URLs and the detectable results above LoD in a cohort of healthy population. The diagnostic performance of the hs-cTnI (CLIA) assay was validated in a population of 934 patients with suspected ACS.Results: The 99th percentile URLs were 15.3 ng/L for female, 31.3 ng/L for male and 24.2 ng/L for overall population. The total imprecision near the sex-specific 99th percentile URLs were <5 %. 76.74 % of females, 97.12 % of males and 86.69 % of overall population had cTnI values exceeding the LoD, which met the criteria of high-sensitivity troponin assay. No cross-reactivity or interference was identified. The diagnostic sensitivity, specificity, PPV, NPV, and AUC of hs-cTnI (CLIA) assay were 97.97 , 90.70, 79.02, 99.21 % and 0.9885, respectively, which were comparable to ARCHITECT hs-cTnI assay.Conclusions: hs-cTnI (CLIA) assay is a high-sensitivity troponin I method with high precision, sensitivity and specificity. The clinical diagnostic performance of hs-cTnI (CLIA) is comparable to the established ARCHITECT hs-cTnI assay. Mindray's hs-cTnI (CLIA) assay is an attractive alternative for diagnosis of myocardial infarction with a high level of accuracy and safety.
引用
收藏
页码:353 / 360
页数:8
相关论文
共 50 条
  • [41] Diagnostic performance of a point-of-care high-sensitivity cardiac troponin I assay among Chinese patients with chest pain
    Ho, Jonathan Ka Ming
    Wong, Janet Yuen Ha
    Tse, Gary
    Chong, Andy Chun Yin
    Chau, Calvin Chi Wai
    Wong, Chi Yip
    Tse, Johnson Wai Keung
    Tam, Jeremy Yan Hon
    Lam, Simon Ching
    OPEN HEART, 2024, 11 (02):
  • [42] Clinical performance of a new point-of-care cardiac troponin I assay compared to three laboratory troponin assays
    Hjortshoj, Soren
    Venge, Per
    Ravkilde, Jan
    CLINICA CHIMICA ACTA, 2011, 412 (3-4) : 370 - 375
  • [43] A very high level in a cardiac troponin I assay
    Corne, C
    Durand, M
    Vergnaud, S
    Renversez, JC
    Lebrun, C
    ANNALES DE BIOLOGIE CLINIQUE, 2005, 63 (03) : 326 - 328
  • [44] Evaluation of the analytical and clinical performance of a high-sensitivity troponin I point-of-care assay in the Mersey Acute Coronary Syndrome Rule Out Study (MACROS-2)
    Dakshi, Ahmed
    Hatherley, James
    Collinson, Paul
    Phillips, Suzannah
    Bailey, Lisa
    Miller, Guy
    Shaw, Matthew
    Khand, Aleem
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2025, 63 (02) : 422 - 432
  • [45] Troponin I levels in asymptomatic patients on haemodialysis using a high-sensitivity assay
    Kumar, Neeraj
    Michelis, Michael F.
    DeVita, Maria V.
    Panagopoulos, Georgia
    Rosenstock, Jordan L.
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2011, 26 (02) : 665 - 670
  • [46] Normal Plasma Levels of Cardiac Troponin I Measured by the High-Sensitivity Cardiac Troponin I Access Prototype Assay and the Impact on the Diagnosis of Myocardial Ischemia
    Venge, Per
    Johnston, Nina
    Lindahl, Bertil
    James, Stefan
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 54 (13) : 1165 - 1172
  • [47] The new high-sensitivity cardiac troponin T assay improves risk assessment in acute coronary syndromes
    Lindahl, Bertil
    Venge, Per
    James, Stefan
    AMERICAN HEART JOURNAL, 2010, 160 (02) : 224 - 229
  • [48] Why Not Use a High-sensitivity Troponin Assay as an End Point in Clinical Trials?
    dos Santos Ferreira, Carlos Eduardo
    Franca, Carolina Nunes
    Helfenstein Fonseca, Francisco Antonio
    AMERICAN JOURNAL OF MEDICINE, 2015, 128 (04) : E29 - E29
  • [49] Diagnostic performance of a point of care high-sensitivity cardiac troponin I assay and single measurement evaluation to rule out and rule in acute coronary syndrome
    De Iuliis, Vincenzo
    Gabriele, Anna Rita
    De Santis, Francesca
    De Rugeriis, Roberta
    Di Quinzio, Luigi
    Aloisi, Steeve
    Rosati, Anna Carla
    Benvenuto, Manuela
    Fabiani, Donatello
    Chiatamone Ranieri, Sofia
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2024, 62 (11) : 2326 - 2334
  • [50] Gender-specific reference values for high-sensitivity cardiac troponin T and I in well-phenotyped healthy individuals and validity of high-sensitivity assay designation
    Giannitsis, Evangelos
    Mueller-Hennessen, Matthias
    Zeller, Tanja
    Schuebler, Anna
    Aurich, Matthias
    Biener, Moritz
    Vafaie, Mehrshad
    Stoyanov, Kiril M.
    Ochs, Marco
    Riffel, Johannes
    Mereles, Derliz
    Blankenberg, Stefan
    Katus, Hugo A.
    CLINICAL BIOCHEMISTRY, 2020, 78 : 18 - 24