Disposable Airway Pressure Manometers for Endotracheal Tube Cuff Inflation

被引:0
作者
Klonner, Moriz Ettore [1 ]
Mattaliano, Giorgio [1 ]
Casoria, Vincenzo [2 ]
Vogl, Claus [3 ]
Braun, Christina [1 ]
机构
[1] Vetmeduni Vienna, Clin Unit Anaesthesiol & Perioperat Intens Care Me, A-1210 Vienna, Austria
[2] Southern Cty Vet Specialists, Ringwood BH4 3JW, England
[3] Vetmeduni Vienna, Inst Anim Breeding & Genet, Unit Mol Genet, A-1210 Vienna, Austria
来源
ANIMALS | 2023年 / 13卷 / 03期
关键词
cuff pressure; endotracheal tube; high-volume; low-pressure; manometer; INTRACUFF PRESSURE; INTUBATION;
D O I
10.3390/ani13030475
中图分类号
S8 [畜牧、 动物医学、狩猎、蚕、蜂];
学科分类号
0905 ;
摘要
Simple Summary Evidence suggests that the cuff pressure of endotracheal tubes with high-volume, low-pressure cuffs should be maintained between 20 and 30 cmH(2)O. Blind techniques for cuff inflation have been shown to rarely result in an optimal cuff pressure, even when used by experienced anaesthesiologists. The authors therefore emphasise the necessity for direct measurement of the cuff pressure using a manometer. Despite a wide array of commercially available cuff pressure manometers, many veterinary facilities still use blind techniques or repurposed aneroid blood pressure manometers, which have been proven unreliable for this application. With cost being a major concern, especially for smaller veterinary facilities, the aim of this study was to find a cost-effective alternative to commercially available cuff manometers and test them for their usability, accuracy, precision and repeatability. This study assessed the performance of two disposable airway manometers for endotracheal tube cuff inflation in a benchtop model. Both of the tested devices present cost-effective and accurate alternatives to commercial cuff manometers. This study aimed to assess the performance, accuracy, precision and repeatability of two single-use airway pressure manometers as a cost-effective alternative for inflation of endotracheal tubes with high-volume, low-pressure cuffs. The manometers were tested in a bench top model against a U-tube manometer. Eighteen units of each device were tested. Three consecutive measurements were performed at pressures of 20, 25 and 30 cmH(2)O each. The mean +/- SD of the recorded pressures and maximum deviation from the target pressures were calculated for each device and each target pressure. For device A, the mean +/- SD pressures were 19.6 +/- 0.7, 23.6 +/- 0.8 and 28.3 +/- 0.8 cmH(2)O; for device B, the mean +/- SD pressures were 19.3 +/- 0.6, 24.3 +/- 0.9 and 29.2 +/- 0.67 cmH(2)O for target pressures of 20, 25 and 30 cmH(2)O, respectively. The bias for device A was -0.4, -1.4, and -1.7 cmH(2)O and for device B, -0.7, -0.7, and -0.8 cmH(2)O for target pressures of 20, 25, and 30 cmH(2)O, respectively. Both devices showed results comparable to those reported for commercial cuff manometers. They represent inexpensive tools that provide clinically sufficient accuracy, precision and repeatability for ETT cuff inflation between pressures of 20 and 30 cmH(2)O.
引用
收藏
页数:10
相关论文
共 37 条
  • [1] A cuff inflator for tracheal tubes
    Abdelatti, MO
    Kamath, BSK
    [J]. ANAESTHESIA, 1997, 52 (08) : 765 - 769
  • [2] Tracheal necrosis following tracheal intubation in a dog
    Alderson, B.
    Senior, J. M.
    Dugdale, A. H. A.
    [J]. JOURNAL OF SMALL ANIMAL PRACTICE, 2006, 47 (12) : 754 - 756
  • [3] Annoni Raquel, 2015, Rev. bras. ter. intensiva, V27, P228, DOI 10.5935/0103-507X.20150037
  • [4] ADJUSTMENT OF INTRACUFF PRESSURE TO PREVENT ASPIRATION
    BERNHARD, WN
    COTTRELL, JE
    SIVAKUMARAN, C
    PATEL, K
    YOST, L
    TURNDORF, H
    [J]. ANESTHESIOLOGY, 1979, 50 (04) : 363 - 366
  • [5] Aspects of in vivo endotracheal tube intracuff pressure in cats
    Bird, Alice R.
    Bird, David J.
    McMillan, Matthew W.
    [J]. VETERINARY ANAESTHESIA AND ANALGESIA, 2019, 46 (01) : 55 - 63
  • [6] Blanch Paul B, 2004, Respir Care, V49, P166
  • [7] Evaluation of the endotracheal tube cuff pressure resulting from four different methods of inflation in dogs
    Briganti, Angela
    Portela, Diego A.
    Barsotti, Giovanni
    Romano, Marta
    Breghi, Gloria
    [J]. VETERINARY ANAESTHESIA AND ANALGESIA, 2012, 39 (05) : 488 - 494
  • [8] Castilho Emanuel Celice, 2003, Rev. Bras. Anestesiol., V53, P743
  • [9] EVOLUTION OF TRACHEAL INJURY DUE TO VENTILATORY ASSISTANCE THROUGH CUFFED TUBES - A PATHOLOGIC STUDY
    COOPER, JD
    GRILLO, HC
    [J]. ANNALS OF SURGERY, 1969, 169 (03) : 334 - &
  • [10] Air leakage around endotracheal tube cuffs
    Dullenkopf, A
    Schmitz, A
    Frei, M
    Gerber, AC