Three-Dimensional CT-Based Limb Length Evaluation Is Highly Dependent on Anatomical Landmark Selection and Pelvic Asymmetry

被引:1
作者
O'Callaghan, William B. [1 ,2 ]
Thompson, Matt [3 ]
Wuestemann, Thies [3 ]
Whitehouse, Sarah L. [1 ,4 ]
Crawford, Ross W. [1 ,4 ,5 ]
机构
[1] Queensland Hlth, Prince Charles Hosp, Dept Orthopaed, Chermside, Qld, Australia
[2] Orthopaed Res Inst Queensland ORIQL, Townsvillle, Qld, Australia
[3] Stryker Corp, Kalamazoo, MI USA
[4] Queensland Univ Technol, Kelvin Grove, Qld, Australia
[5] Queensland Univ Technol, Prince Charles Hosp, Orthopaed Res, Rode Rd, Chermside, Qld 4032, Australia
来源
ARTHROPLASTY TODAY | 2023年 / 23卷
关键词
Total hip arthroplasty; Pelvic asymmetry; Global offset; Hip length; Robotics; Digital templating; TOTAL HIP-ARTHROPLASTY; DIRECTIONAL ASYMMETRY; DISCREPANCY; BONE; PREVALENCE; DIFFERENCE; IMPACT; RISK;
D O I
10.1016/j.artd.2023.101206
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Pelvic skeletal asymmetry can result in rotational differences and morphologic bony prominence variance between the left and right hemipelvis. When selecting bony reference points for modern computed tomography-based robotic total hip arthroplasty planning, it is unclear which bony landmarks are the most reliable and accurate, especially in the presence of significant pelvic asymmetry.Methods: A retrospective study was conducted utilizing a database of computed tomography scans. Multiple bony landmarks in the pelvis and femur were selected for comparison, with the aim of measuring pelvic asymmetry. Specifically, the study measured the average difference in lateral offset between the left and right hemipelvis caused by pelvic asymmetry. Landmarks were also compared to determine the impact of pelvic asymmetry on hip length, femur length, and limb length discrepancies. Furthermore, a scenario was simulated in the software whereby a total hip replacement was inserted, potentially changing the hip length. The impact of pelvic reference point selection on the measurement of this simulated change in hip length was examined.Results: This study population showed widespread pelvic asymmetry. The anatomical landmarks of the opposite side cannot be relied upon for predicting the anatomy of the affected side. The center of rotation axis is more reliable than the inferior obturator foramen axis for hip length discrepancy due to pelvic asymmetry (P < .05).Conclusions: Current computer-assisted surgery THR software reports measurements of global offset and hip length that do not consider pelvic asymmetry. Surgeons are not given confidence ranges to represent the potential impact of asymmetry on the global offset and hip length values. Surgeons following these numbers to guide implant position may incur implant placement error should significant pelvic asymmetry be present in a given patient.(c) 2023 The Authors. Published by Elsevier Inc. on behalf of The American Association of Hip and Knee Surgeons. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/ 4.0/).
引用
收藏
页数:6
相关论文
共 32 条
  • [1] ABRAHAM WD, 1992, ORTHOP CLIN N AM, V23, P201
  • [2] Patterns of clavicular bilateral asymmetry in relation to the humerus: variation among humans
    Auerbach, Benjamin M.
    Raxter, Michelle H.
    [J]. JOURNAL OF HUMAN EVOLUTION, 2008, 54 (05) : 663 - 674
  • [3] Pelvic bone asymmetry in 323 study participants receiving abdominal CT scans
    Badii, M
    Shin, S
    Torreggiani, WC
    Jankovic, B
    Gustafson, P
    Munk, PL
    Esdaile, JM
    [J]. SPINE, 2003, 28 (12) : 1335 - 1339
  • [4] Three-dimensional study of pelvic asymmetry on anatomical specimens and its clinical perspectives
    Boulay, C
    Tardieu, C
    Bénaim, C
    Hecquet, J
    Marty, C
    Prat-Pradal, D
    Legaye, J
    Duval-Beaupère, G
    Pélissier, J
    [J]. JOURNAL OF ANATOMY, 2006, 208 (01) : 21 - 33
  • [5] Does the demand for asymmetric functional lower body postures in lateral sports relate to structural asymmetry of the pelvis?
    Bussey, Melanie D.
    [J]. JOURNAL OF SCIENCE AND MEDICINE IN SPORT, 2010, 13 (03) : 360 - 364
  • [6] Pelvic post-traumatic asymmetry: assessment and sequenced treatment
    Cano-Luis, Pedro
    Angel Giraldez-Sanchez, Miguel
    Andres-Cano, Pablo
    [J]. EFORT OPEN REVIEWS, 2018, 3 (05): : 335 - 346
  • [7] The use of a 3-dimensional computed tomography bone database to evaluate the risk of distal contact between the rasp tip and the endosteal cortical bone
    Connor, Emmalynn
    Cowie, Jonathan G.
    Wuestemann, Thies
    Howell, Jonathan R.
    Whitehouse, Sarah L.
    Crawford, Ross W.
    [J]. JOURNAL OF ORTHOPAEDIC SURGERY, 2016, 24 (03) : 312 - 316
  • [8] THE EFFECT OF IMPOSED LEG LENGTH DIFFERENCE ON PELVIC BONE SYMMETRY
    CUMMINGS, G
    SCHOLZ, JP
    BARNES, K
    [J]. SPINE, 1993, 18 (03) : 368 - 373
  • [9] A three-dimensional analysis of bilateral directional asymmetry in the human clavicle
    Fatah, Emam ElHak Abdel
    Shirley, Natalie R.
    Mahfouz, Mohamed R.
    Auerbach, Benjamin M.
    [J]. AMERICAN JOURNAL OF PHYSICAL ANTHROPOLOGY, 2012, 149 (04) : 547 - 559
  • [10] Leg length discrepancy and lower limb alignment after total hip arthroplasty in unilateral hip osteoarthritis patients
    Fujimaki, Hiroshi
    Inaba, Yutaka
    Kobayashi, Naomi
    Tezuka, Taro
    Hirata, Yasuhide
    Saito, Tomoyuki
    [J]. JOURNAL OF ORTHOPAEDIC SCIENCE, 2013, 18 (06) : 969 - 976