Automated blood pressure measuring devices: how are they clinically validated for accuracy?

被引:5
|
作者
Ringrose, Jennifer S. [1 ,2 ]
Padwal, Raj [1 ,2 ]
机构
[1] Univ Alberta, Dept Med, Edmonton, AB, Canada
[2] Women & Childrens Hlth Res Inst, Edmonton, AB, Canada
关键词
HYPERTENSION-INTERNATIONAL-PROTOCOL; MEDICAL INSTRUMENTATION/EUROPEAN SOCIETY; EUROPEAN-SOCIETY; HYPERTENSION/INTERNATIONAL ORGANIZATION; UNIVERSAL STANDARD; ASSOCIATION; ADVANCEMENT; PREVENTION; MONITORS; ADULTS;
D O I
10.1038/s41371-022-00761-2
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Accurate blood pressure (BP) assessment is essential for the optimal diagnosis and management of hypertension. Contemporary clinical practice guidelines strongly endorse use of automated cuff blood pressure measuring devices (BPMD) as the preferred means of measuring and monitoring BP in the office, at home and with ambulatory blood pressure monitoring. To ensure that they are accurate, automated BPMDs should undergo clinical validation testing, performed using an established clinical validation standard. Unfortunately, most BPMDs sold on the global market have not been clinically validated. Furthermore, in the last thirty years, several different clinical validation protocols have been published, with major differences apparent between these standards, causing controversy with respect to which standard is considered acceptable for clinical validation. Complexly worded standards, multiple revisions, and firewalled access also contribute to a lack of understanding and use of clinical validation standards and the number of expert centers performing clinical validations is small. Recently, joint society collaborations have led to creation of the AAMI/ESH/ISO universal standard for the clinical validation of automated cuff BPMDs. Not only is this unified standard a necessary step, but oversight from regulators and influential stakeholders to ensure that only clinically validated BPMDs can be marketed is additionally needed.
引用
收藏
页码:101 / 107
页数:7
相关论文
共 50 条
  • [21] Calibration accuracy of hospital based automated blood pressure devices.
    Hoffmann, A.
    Lorde, I
    Shennan, A. H.
    HYPERTENSION IN PREGNANCY, 2006, 25 : 164 - 164
  • [22] Accuracy of automated community pharmacy-based blood pressure devices
    Gonzalvo, Jasmine
    Zillich, Alan
    JOURNAL OF THE AMERICAN PHARMACISTS ASSOCIATION, 2011, 51 (03) : 408 - 411
  • [23] Rapid estimation of the accuracy of automatic blood pressure measuring devices (READ)
    JE Naschitz
    L Gaitini
    L Lowenstein
    D Keren
    A Tamir
    D Yeshurun
    Journal of Human Hypertension, 1999, 13 : 443 - 447
  • [24] Rapid estimation of the accuracy of automatic blood pressure measuring devices (READ)
    Naschitz, JE
    Gaitini, L
    Lowenstein, L
    Keren, D
    Tamir, A
    Yeshurun, D
    JOURNAL OF HUMAN HYPERTENSION, 1999, 13 (07) : 443 - 447
  • [25] How to check whether a blood pressure monitor has been properly validated for accuracy
    Picone, Dean S.
    Padwal, Raj
    Campbell, Norm R. C.
    Boutouyrie, Pierre
    Brady, Tammy M.
    Olsen, Michael Hecht
    Delles, Christian
    Lombardi, Cintia
    Mahmud, Azra
    Meng, Yaxing
    Mokwatsi, Gontse G.
    Ordunez, Pedro
    Phan, Hoang T.
    Pucci, Giacomo
    Schutte, Aletta E.
    Sung, Ki-Chul
    Zhang, Xin-Hua
    Sharman, James E.
    JOURNAL OF CLINICAL HYPERTENSION, 2020, 22 (12): : 2167 - 2174
  • [26] HOW TO CHECK WHETHER A BLOOD PRESSURE DEVICE HAS BEEN PROPERLY VALIDATED FOR ACCURACY
    Sharman, James E.
    JOURNAL OF HYPERTENSION, 2023, 41 : E84 - E84
  • [27] The hypertension Canada blood pressure device recommendation listing: Empowering use of clinically validated devices in Canada
    Padwal, Raj
    Berg, Angelique
    Gelfer, Mark
    Tran, Karen
    Ringrose, Jennifer
    Ruzicka, Marcel
    Hiremath, Swapnil
    JOURNAL OF CLINICAL HYPERTENSION, 2020, 22 (05): : 933 - 936
  • [28] HOME BLOOD PRESSURE-MEASURING DEVICES - A COMPARATIVE-STUDY OF ACCURACY
    EVANS, CE
    HAYNES, RB
    GOLDSMITH, CH
    HEWSON, SA
    JOURNAL OF HYPERTENSION, 1989, 7 (02) : 133 - 142
  • [29] HEARTS in the Americas: a global example of using clinically validated automated blood pressure devices in cardiovascular disease prevention and management in primary health care settings
    Pedro Ordunez
    Cintia Lombardi
    Dean S. Picone
    Tammy M. Brady
    Norm R. C. Campbell
    Andrew E. Moran
    Raj Padwal
    Andres Rosende
    Paul K. Whelton
    James E. Sharman
    Journal of Human Hypertension, 2023, 37 : 126 - 129
  • [30] HEARTS in the Americas: a global example of using clinically validated automated blood pressure devices in cardiovascular disease prevention and management in primary health care settings
    Ordunez, Pedro
    Lombardi, Cintia
    Picone, Dean S.
    Brady, Tammy M.
    Campbell, Norm R. C.
    Moran, Andrew E.
    Padwal, Raj
    Rosende, Andres
    Whelton, Paul K.
    Sharman, James E.
    JOURNAL OF HUMAN HYPERTENSION, 2023, 37 (02) : 126 - 129